Agilent Technologies 5900-icp-oes
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5900-icp-oes

Agilent Technologies 5900-icp-oes — Atomic spectroscopy.

The Agilent 5900 ICP-OES is a dual-view (DV) atomic emission spectrometer built around a freeform echelle polychromator with a 50% reduced optical volume and 167–785 nm coverage. Its 1.8 mm injector and 35% TDS ceiling define the standard aqueous/acidic torch configuration, while the 40% detection-limit improvement over the preceding generation positions it for trace-element work in complex matrices.

Agilent Technologies ICP Oes Instruments Series — Model Comparison & Selection

Model Wavelength rangeDetection limit improvement, %Injector internal diameter id, mmMaximum total dissolved solids tds, %Polychromator volume reduction, %
5900-icp-oes Current 167...785401.83550
5800-icp-oes 167...785401.83550

Full specifications

Core Performance & Functional Specifications
Wavelength range 167...785 nm
Detection Limit Improvement 40 %
Injector Internal Diameter (id) 1.8 mm
Maximum Total Dissolved Solids (tds) 35 %
Polychromator Volume Reduction 50 %
Ba 614.171 detection limit mfr 32
Most aqueous or acidic samples mfr Standard quartz quartz
Volatile and semi-volatile organic samples mfr 'Organic' high purity quartz quartz
As 188.980 mfr 6.5
Mo 202.032 mfr 7
Zn 213.857 mfr 7.5
Pb 220.353 mfr 7.5
Cr 267.716 mfr 9.5
Cu 327.395 mfr 13
Ba 614.171 mfr 32
pixel count mfr 70000
DLA count mfr 70
full spectrum measurement time mfr less than half a second second
Dimensions, Weight & Mechanical Design
injector inner diameter mfr 2.4 mm
optical focal length mfr 253 mm
Electrical, Power & Drive Specifications
High-TDS matrices rich in alkali metals (including Li battery materials), hydrofluoric (HF) acid digests or fusions mfr Ceramic outer outer
Operating Conditions, Environment & Protection
Cooling time before torch removal mfr at least five minutes minutes
cooling time before torch removal mfr cinque minuti minuti

Description

Construction and series positioning

The 5900 is the higher-sensitivity member of the 5800/5900 ICP-OES pair; the sibling 5800 shares the freeform-optics architecture but the 5900 is specified with a 40% detection-limit improvement and a polychromator volume reduced by 50%. The instrument is configured for Dual View (DV) operation, which requires the slotted, longer DV outer tube-set rather than the shorter, slot-free Radial View (RV) tube-set. Torch hardware follows the Easy-fit family: one-piece, semi-demountable, and fully demountable variants are all supported, and the fully demountable torch can be switched between RV and DV by changing only the outer tube-set.

Torch and sample-path behavior

The 35% TDS ceiling is tied specifically to the 2.4 mm alumina injector; it is not a blanket limit for every injector diameter.

Installation, control and operation

The fully demountable torch ships without the injector installed. The injector is inserted until the bulge contacts the retaining clamp and clicks into position, then the locking nut is fitted by hand only. During reassembly the groove on the torch body must align with the 'T' mark on the outer tube-set, and the tube-set is pressed fully into the body. The ceramic outer tube-set carries a fixed polymer adapter on its quartz intermediate tube that must not be removed; the supplied tube-set tool is used for assembly and disassembly of the ceramic/quartz joint. Torch installation and removal must follow a straight, parallel path to avoid breaking quartz or ceramic components. After plasma switch-off, the torch and compartment must cool for at least five minutes before the torch is removed. Detailed maintenance and cleaning procedures are delegated to the ICP Expert Help and Learning Center and publication 5994-8864EN.

Integration & Operating Notes

Process Limits
Do not exceed 35% TDS with the 2.4 mm alumina injector

The 35% total-dissolved-solids ceiling applies to the G8020-47004 2.4 mm id inert alumina injector used with the ceramic outer tube-set. Higher TDS loads risk salt deposition and devitrification-related failure; for matrices above this level, dilution or a different sample-introduction strategy is required before the torch is installed.

5100 ICP OES User Manual, page 7
Mounting
Insert and remove the torch with a straight parallel motion

Any angular motion while sliding the torch or its quartz/ceramic components into the compartment can crack the outer tube-set or injector and create a hand-injury hazard from broken glass. The torch must be guided straight in and out; the same rule applies during reassembly of the fully demountable torch.

5100 ICP OES User Manual, pages 3–4
Maintenance
Hand-tighten the injector locking nut only

The locking nut (G8020-60810) must be tightened by hand. Using tools or over-torquing can damage both the injector and the torch body, and a damaged body is not field-repairable without replacement parts. The same hand-tight rule applies at first injector installation and at every reassembly.

5100 ICP OES User Manual, pages 3, 5
Maintenance
Do not detach the polymer adapter from the ceramic tube-set

The quartz intermediate tube of the ceramic outer tube-set carries a fixed polymer adapter. It is not a serviceable joint and must not be removed during cleaning or maintenance; detaching it compromises the tube-set and requires replacement of the assembly.

5100 ICP OES User Manual, page 5

Selection & Integration Q&A

When should I choose the 5900 over the 5800, and when is the 5800 sufficient?

The 5900 is specified with a 40% detection-limit improvement and a 50% smaller polychromator volume than the preceding design; the 5800 shares the freeform-optics platform. Choose the 5900 when trace-level detection limits in complex matrices drive the method; the 5800 covers routine work where those limits are not the constraint.

Which torch and injector configuration do I need for HF digests or high-TDS samples, and what must be ordered alongside?

Use the ceramic outer tube-set with the 2.4 mm id inert alumina injector G8020-47004, rated to 35% TDS. The ceramic tube-set kit does not include an injector, so a compatible quartz or alumina injector must be ordered separately; the supplied tube-set tool is required for assembly.

What spare parts and consumables are not included in the base delivery and should be stocked?

The fully demountable torch ships without an injector installed. Stock replacement injectors (quartz or inert alumina, 0.8/1.4/1.8/2.4 mm id), outer tube-sets for RV or DV, top seal pack G8014-60023, and injector locking nut G8020-60810. Cleaning procedures are in publication 5994-8864EN.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Easy-fit Torch ICP-OES User Guide
Manual ·65 pp.
View PDF
PDF
Innovative Freeform Optical Design
Brochure ·731,1 КБ ·6 pp.
View PDF