Agilent Technologies 5800-icp-oes
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5800-icp-oes

Agilent Technologies 5800-icp-oes — Atomic spectroscopy.

The Agilent 5800 ICP-OES is a dual-view atomic emission spectrometer built around a freeform echelle polychromator covering 167–785 nm. Its 1.8 mm injector and 35% TDS ceiling define the standard aqueous/acidic configuration, while the fully demountable Easy-fit torch allows injector and outer-tube exchange for organic, HF, or high-solids matrices. It serves as the mid-series workhorse below the 5900 for routine multi-element determinations.

Agilent Technologies ICP Oes Instruments Series — Model Comparison & Selection

Model Wavelength rangeDetection limit improvement, %Injector internal diameter id, mmMaximum total dissolved solids tds, %Polychromator volume reduction, %
5800-icp-oes Current 167...785401.83550
5900-icp-oes 167...785401.83550

Full specifications

Core Performance & Functional Specifications
Wavelength range 167...785 nm
Detection Limit Improvement 40 %
Injector Internal Diameter (id) 1.8 mm
Maximum Total Dissolved Solids (tds) 35 %
Polychromator Volume Reduction 50 %
Ba 614.171 detection limit mfr 32
As 188.980 mfr 6.5
Mo 202.032 mfr 7
Zn 213.857 mfr 7.5
Pb 220.353 mfr 7.5
Cr 267.716 mfr 9.5
Cu 327.395 mfr 13
Ba 614.171 mfr 32
pixel count mfr 70000
DLA count mfr 70
full spectrum measurement time mfr less than half a second second
Dimensions, Weight & Mechanical Design
optical focal length mfr 253 mm
Operating Conditions, Environment & Protection
Cooling time before torch removal mfr at least five minutes minutes

Description

Construction and series positioning

The 5800 ICP-OES is the base dual-view instrument of the 5800/5900 pair; the 5900 is the higher-tier sibling. The optical train is a freeform echelle polychromator, quoted with a 50% reduction in polychromator volume and a 40% detection-limit improvement relative to the preceding generation. Wavelength coverage spans 167–785 nm, reaching the vacuum-UV range needed for halogens and other far-UV lines. The sample-introduction end is built around the Easy-fit torch family: one-piece, semi-demountable, and fully demountable torches share the same base and injector assembly, but the Radial View outer tube-set is shorter and lacks the top slot of the Dual View tube-set, so RV and DV tube-sets are not interchangeable.

Torch configuration and matrix behavior

The fully demountable torch is the configuration that adapts the 5800 to different matrices. Standard quartz with a 1.8 mm injector covers most aqueous and acidic samples and is the most economical choice. Organic high-purity quartz outer tube-sets with 0.8 mm (volatile) or 1.4 mm (semi-volatile) injectors resist solarization. For high-TDS matrices, HF digests, fusions, and Li-battery materials, the ceramic outer tube-set with the 2.4 mm inert alumina injector (G8020-47004) gives the longest service life and resistance to devitrification; that injector is rated to a maximum of 35% total dissolved solids. The fully demountable torch works in either RV or DV mode provided the matching outer tube-set is fitted.

Installation, control and operation

The torch ships without the injector installed: the shipping plug is removed, the injector pushed in until it clicks against the retaining clamp, and the locking nut fitted by hand only. During reassembly the groove on the torch body must align with the 'T' mark on the outer tube-set, and the tube-set pressed fully home. The ceramic outer tube-set carries a fixed polymer adapter on its quartz intermediate tube that must not be removed; the supplied tube-set tool is used to separate ceramic and quartz parts. Torch insertion and removal must follow a straight, parallel path to avoid breaking quartz or ceramic. After plasma switch-off, the torch and compartment require at least five minutes of cooling before the torch is touched. Final installation and control follow the ICP Expert Help and Learning Center; maintenance and cleaning follow publication 5994-8864EN.

Integration & Operating Notes

Process Limits
Do not exceed 35% TDS with the 2.4 mm alumina injector

The G8020-47004 inert alumina injector is the high-solids option, but its rating stops at 35% total dissolved solids. Samples above that level require dilution or a different sample-introduction strategy; running beyond the limit risks injector blockage and drift rather than a clean measurement.

5100 ICP OES User Manual, page 7
Maintenance
Hand-tighten the injector locking nut only

The locking nut (G8020-60810) must be tightened by hand. Using tools or over-torquing can crack the injector and distort the torch body, which then leaks or misaligns the plasma. Treat the click of the injector seating as the setup confirmation, not nut torque.

5100 ICP OES User Manual, pages 3 and 5
Maintenance
Wait at least five minutes after plasma switch-off

The torch and torch compartment remain hot after the plasma is extinguished. Removal before a five-minute cool-down risks burns and thermal shock to the quartz or ceramic components. Build the wait into shutdown procedures before any torch exchange.

5100 ICP OES User Manual, page 4
Mounting
Match RV or DV outer tube-set to the observation mode

The RV outer tube-set is shorter and has no top slot; the DV tube-set does. Fitting the wrong one prevents correct plasma observation and seating. Confirm the observation mode before ordering or installing a tube-set, since the fully demountable torch itself accepts either when paired correctly.

5100 ICP OES User Manual, pages 1–2
Maintenance
Keep the ceramic tube-set polymer adapter fixed

The quartz intermediate tube of the ceramic outer tube-set carries a fixed polymer adapter that must not be detached during cleaning. Use the supplied tube-set tool to separate ceramic from quartz; prying the adapter off damages the assembly and compromises the seal.

5100 ICP OES User Manual, page 5
Maintenance
Ceramic tube-set kits ship without an injector

The fully demountable kit with ceramic outer tube-set does not include an injector; order a compatible quartz or inert alumina injector separately. Standard quartz kits with the 1.8 mm injector are complete for aqueous/acidic work, so specify the injector explicitly on ceramic orders.

5100 ICP OES User Manual, page 2

Selection & Integration Q&A

When should I choose the 5800 over the 5900, and when is the fully demountable torch the right configuration?

The 5800 is the base dual-view unit; the 5900 is the higher-tier sibling in the same series. Choose the 5800 when the 167–785 nm freeform-optics coverage and standard aqueous/acidic performance meet the method. Choose the fully demountable torch when matrices vary: organic (0.8/1.4 mm), HF or high-TDS up to 35% (2.4 mm alumina).

What sample-introduction parts must be ordered alongside the instrument to cover aqueous, organic and HF work?

The fully demountable torch ships without an injector, and ceramic outer tube-set kits exclude it. Order quartz injectors in 0.8, 1.4, 1.8 and 2.4 mm id and inert alumina equivalents as needed; the 2.4 mm alumina G8020-47004 covers HF and TDS up to 35%. Also stock top seals (G8014-60023, pack of 3) and the locking nut G8020-60810.

What mounting and handling limits apply when installing or swapping the torch?

Insert and remove the torch with a straight, parallel motion to avoid breaking quartz or ceramic. Align the torch-body groove with the 'T' mark and press the tube-set fully home. Wait at least five minutes after plasma switch-off before removal, and hand-tighten the locking nut only.

What maintenance consumables and procedures are outside the base delivery?

Base delivery does not include spare injectors, outer tube-sets, top seals or locking nuts. Cleaning and maintenance follow publication 5994-8864EN and the ICP Expert Help and Learning Center. The ceramic tube-set's fixed polymer adapter must not be removed; use the supplied tube-set tool for disassembly.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Innovative Freeform Optical Design
Brochure ·731,1 КБ ·6 pp.
View PDF