Agilent Technologies 7900-icp-ms
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 7900-icp-ms

Agilent Technologies 7900-icp-ms — Atomic spectroscopy.

The Agilent 7900 ICP-MS is a single-quadrupole ICP-MS with a helium collision/reaction cell, offering an 11-order linear dynamic range and matrix tolerance to tens of %TDS. It achieves sub-ppt detection limits for 238U and 111Cd, and is suited for trace-element and nanoparticle analysis in complex matrices.

Agilent Technologies ICP MS Instruments Series — Model Comparison & Selection

Model SensitivityMinimum dwell time, msAnalysis time, minData repeatability, %Instrument depth, mmInstrument height, mm
7900-icp-ms Current 109—————
8900-triple-quadrupole-icp-ms 7000.1————
9500-triple-quadrupole-icp-ms 100000.0535.54600530

Full specifications

Core Performance & Functional Specifications
Sensitivity 109 cps/ppm
Linear dynamic range mfr 11 magnitude
Matrix tolerance mfr tens of %TDS range %TDS %TDS
Collision/reaction cell mode mfr helium collision mode mode
238U detection limit mfr 0.001253 ppt
238U background equivalent concentration (BEC) mfr 0.0004822 ppt
111Cd detection limit mfr 0.2915 ppt
111Cd background equivalent concentration (BEC) mfr 0.07899 ppt
23Na detection limit mfr 0.008635 ppm
23Na background equivalent concentration (BEC) mfr 0.04974 ppm
Nanoparticle dwell time mfr 0.1 ms
Sample volume for small-volume analysis mfr 0.5 mL
Collision/reaction cell mfr octopole reaction system , helium collision mode
Detection limit, 238U mfr 0.001253 ppt
BEC, 238U mfr 0.0004822 ppt
Detection limit, 111Cd mfr 0.2915 ppt
BEC, 111Cd mfr 0.07899 ppt
Detection limit, 23Na mfr 0.008635 ppm
BEC, 23Na mfr 0.04974 ppm
Dwell time, single nanoparticle TRA mfr 0.1 ms
Sample volume, integrated sample introduction system mfr 0.5 mL
Autosampler mfr 4
Switching valve mfr seven-port port
Cones mfr Standard Ni, or optional Pt-tipped tipped
Design mfr Compact benchtop benchtop
Long-term stability mfr 3.5 h
Matrix tolerance (TDS) mfr 25 % TDS
Dynamic range mfr 11 magnitude
CeO ratio mfr 2 % CeO
Cell gas switching time mfr 3 seconds
Detector range mfr 0.1 Gcps
SPS 4 capacity mfr 360 samples
SPS 6 capacity mfr 540 samples
I-AS sample volume mfr 0.5 mL
I-AS vial capacity mfr 89 vials
Electrical, Power & Drive Specifications
Plasma RF frequency mfr 27 MHz
Standards, Safety & General Data
Top calibration standard, 23Na mfr 10000 ppm
CeO/Ce ratio mfr 1

Description

Construction and series positioning

The 7900 ICP-MS is the single-quadrupole member of Agilent's ICP-MS series, positioned between the 7850 and the triple-quadrupole 8900/9500 models. It uses a helium collision mode for interference suppression, distinguishing it from the 8900/9500, which add a second quadrupole for MS/MS capability. The 7900 is intended for routine multi-element analysis where high matrix tolerance and wide dynamic range are required without the added complexity of triple-quadrupole operation.

Performance characteristics

Installation and operation

The 7900 is typically configured with standard nickel or platinum-tipped cones. It supports optional automated sampling systems such as the SPS 4 autosampler and a seven-port switching valve for high-throughput or small-volume workflows. The helium collision mode requires a dedicated helium supply and optimized cell parameters. Installation requires a suitable laboratory environment with exhaust ventilation, inert gas supplies, and a stable power source; specific electrical and environmental requirements are detailed in the site preparation guide.

Integration & Operating Notes

Process Limits
Matrix tolerance limited to tens of %TDS

Samples with total dissolved solids above tens of percent may cause signal suppression, cone clogging, and drift. Dilute high-matrix samples or use an appropriate internal standard to compensate for matrix effects; exceeding this limit risks invalid results and increased maintenance.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the 7900 ICP-MS over the 8900 triple-quadrupole model?

Choose the 7900 for routine trace-element analysis with high matrix tolerance and wide dynamic range using helium collision mode. The 8900 triple-quadrupole offers superior interference removal for ultra-trace applications in complex matrices, but at higher cost and complexity. Select the 7900 when MS/MS capability is not required.

What are the requirements for connecting an autosampler and switching valve to the 7900?

The 7900 supports the SPS 4 autosampler and a seven-port switching valve for automated sample introduction. These are optional accessories and must be ordered separately. Ensure the autosampler is compatible with the sample tube sizes and that the switching valve is configured for the desired sample loop volume; the 0.5 mL minimum sample volume is achievable with the valve.

What maintenance items and consumables are not included in the base delivery?

The base system includes standard nickel cones; platinum-tipped cones are optional. Autosamplers, switching valves, and additional sample introduction kits are not included. Routine consumables such as peristaltic pump tubing, nebulizers, and torches must be ordered separately. A maintenance schedule should include cone cleaning and tubing replacement based on sample load.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent 7900 ICP-MS Brochure
Brochure ·8,5 МБ ·7 pp.