Agilent Technologies Atmospheric Pressure Chemical Ionization Source, APCI
Agilent Technologies Atmospheric Pressure Chemical Ionization Source, APCI — Liquid chromatography mass spectrometry lc ms with Pumping speed 2.4×10⁻⁵ m³/h.
The Agilent APCI source is an atmospheric-pressure chemical ionization interface for LC/MS systems, extending the series beyond ESI to ionize low-polarity and non-polar analytes. It operates with drying gas at 350 °C and vaporizer at 450 °C, supporting mass ranges up to 7000 m/z. This modification is selected when ESI fails to ionize compounds lacking acidic or basic sites, enabling LC/MS analysis of steroids, lipids, and similar species.
Agilent Technologies LC MS Ion Sources Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Sample volume, milliliter | Column temperature, °C | Drying gas temperature, °C | Sheath gas flow, liter/minute | Sheath gas temperature, °C |
|---|---|---|---|---|---|---|
| Atmospheric Pressure Chemical Ionization Source, APCI Current | 2.4×10⁻⁵ | 0.004 | 55 | 350 | — | — |
| Electrospray ion Source, ESI | 2.4×10⁻⁵ | 0.001 | 80 | 290 | 12 | 400 |
| High sensitivity MS, high sensitivity ion source, Jet Stream | — | — | — | — | 0.011 | 350 |
Full specifications
| Pumping speed | 0.4 mL/min |
| Analysis time | 14 min |
| Column temperature | 55 °C |
| Sample volume | 4 µL |
| Gradient mfr | 55 % B |
| Maximum counts mfr | 1.2 counts |
| MRPL concentration mfr | 2 ng/mL |
| Reconstitution volume mfr | 100 μL |
| Mass range | 7000 m/z |
| Mass accuracy mfr | 3 ppm |
| Mass accuracy achieved mfr | 2 ppm |
| Fragmentor Voltage | 150 V |
| Mobile phase A mfr | 0.1 acid/water |
| Mobile phase B mfr | Methanol Methanol |
| Photomultiplier voltage increase mfr | 50 V |
| Drying Gas Temperature | 350 °C |
| Column part number mfr | 822700 |
Description
Construction and series positioning
- The APCI source is a dedicated ionization module for Agilent LC/MS instruments, distinct from ESI sources in the same series. It uses a heated nebulizer and corona discharge needle to generate ions at atmospheric pressure.
- Sibling models (6230 TOF, 6475/6495 triple quadrupole, 6530 Q-TOF) are complete LC/MS systems; the APCI source is an accessory that can be fitted to compatible MS platforms. Selection depends on analyte polarity: APCI for low-polarity compounds, ESI for polar/ionic species.
Behavior under operating conditions
- Typical operating parameters include drying gas at 350 °C and vaporizer at 450 °C, with fragmentor voltage around 150 V. These conditions promote desolvation and ionization of neutral molecules.
- The source supports mass ranges up to 7000 m/z, suitable for a wide variety of small molecules. Mass accuracy of 2–3 ppm is achievable in conjunction with high-resolution MS platforms.
- No digitized curve points are available; performance is method-dependent and should be optimized per analyte.
Installation and operation
- The source mounts to the MS inlet; requires connection to drying gas, nebulizer gas, and corona discharge power. Control is integrated into the MS software.
- The attached application note (5989 4738EN) describes an LC/MSD TOF method for anabolic steroids using APCI, with LC conditions: column 18 µm, mobile phase A 0.1% acid/water, B methanol, gradient 55% B, column temperature 55 °C, flow 0.4 mL/min, injection 4 µL, analysis time 14 min. MS conditions: APCI, fragmentor 150 V, drying gas 350 °C, vaporizer 450 °C, mass range 7000 m/z, mass accuracy 3 ppm, photomultiplier voltage increase 50 V, maximum counts 1.2, MRPL 2 ng/mL, reconstitution volume 100 µL.
Integration & Operating Notes
Operating the vaporizer above 450 °C may cause thermal degradation of analytes or damage to the source components. The recommended setting for this application is 450 °C; ensure the temperature is controlled and monitored via the MS software to avoid overheating.
The drying gas temperature directly affects desolvation efficiency and signal intensity. For this method, 350 °C is specified; lower temperatures may result in incomplete desolvation and higher background, while higher temperatures could cause analyte degradation. Adjust only with method validation.
The corona discharge needle can become contaminated over time, leading to reduced sensitivity and unstable ionization. Periodic cleaning according to the instrument maintenance manual is necessary. Frequency depends on sample load and matrix; inspect visually and clean when performance degrades.
Selection & Integration Q&A
Choose APCI for low-polarity, non-ionic compounds (e.g., steroids, lipids) that ionize poorly by ESI. ESI is preferred for polar, ionic analytes. APCI is also less susceptible to matrix effects for some compounds. The APCI source is compatible with Agilent LC/MS platforms such as the 6230 TOF or 6475 triple quadrupole.
The APCI source requires dry nitrogen or zero-air for drying and nebulizer gas, typically supplied at 350–450 °C. It also needs a corona discharge power supply, which is integrated into the MS. Ensure gas purity and pressure meet the MS specifications. Electrical connections are made via the MS interface; no external high-voltage supply is needed.
Operate within the MS system's ambient temperature and humidity limits (typically 15–30 °C, <80% RH). The source is designed for continuous operation but requires stable gas flows and temperatures. Avoid frequent power cycling; allow stabilization after temperature changes. Duty cycle is limited by sample throughput and maintenance needs.
The base delivery includes the source assembly but not the corona discharge needle, nebulizer, or gas filters, which are consumables and should be stocked. Regular maintenance includes cleaning the needle, replacing gas filters, and checking seals. Refer to the maintenance schedule in the MS manual for intervals.