Agilent Technologies Ion Sources for Gas Chromatography/Mass Spectrometry
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Ion Sources for Gas Chromatography/Mass Spectrometry

Agilent Technologies Ion Sources for Gas Chromatography/Mass Spectrometry — Gas chromatography mass spectrometry gc ms.

The Agilent JetClean ion source system (G7077) is a hydrogen‑plasma cleaning accessory for Agilent GC/MS ion sources, operating in Acquire & Clean or Clean Only mode. It supports source temperatures of 150–350 °C, emission currents of 10–35 µA, and hydrogen flows of 0.13–0.49 mL/min (Acquire & Clean) or 0.13–3.52 mL/min (Clean Only). It is used to remove contamination in situ, extending source maintenance intervals and preserving attogram‑level detection limits.

Agilent Technologies GC MS Ion Sources Inlets Series — Model Comparison & Selection

Model Detection limitClean duration lower setpoint, minClean duration upper setpoint, minEmission current lower setpoint, microampereEmission current upper setpoint, microampereHydrogen flow acquire clean mode lower setpoint, mL/min
Ion Sources for Gas Chromatography/Mass Spectrometry Current attogram112010350.13
Maintenance free, JetClean ion source for GC/MS instrument —112010350.13

Full specifications

Core Performance & Functional Specifications
Detection limit attogram
Clean Duration, Lower Setpoint 1 min
Clean Duration, Upper Setpoint 120 min
Hydrogen Flow (acquire & Clean Mode), Lower Setpoint 0.13 mL/min
Hydrogen Flow (acquire & Clean Mode), Upper Setpoint 0.49 mL/min
Hydrogen Flow (clean Only Mode), Lower Setpoint 0.13 mL/min
Hydrogen Flow (clean Only Mode), Upper Setpoint 3.52 mL/min
Ionization source mfr low energy capable electron ionization ionizatio
Hydrogen flow step size mfr 0.069 sccm
Dimensions, Weight & Mechanical Design
Mass analyzer type mfr single quadrupole quadrupole
Electrical, Power & Drive Specifications
Emission Current, Lower Setpoint 10 µA
Emission Current, Upper Setpoint 35 µA
Operating Conditions, Environment & Protection
Source Temperature, Lower Setpoint 150 °C
Source Temperature, Upper Setpoint 350 °C

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection, control and operation

Integration & Operating Notes

Process Limits
Hydrogen flow must not exceed mode‑specific maximum

In Acquire & Clean mode the hydrogen flow must stay within 0.13–0.49 mL/min; in Clean Only mode it may be raised to 0.13–3.52 mL/min. Exceeding the Acquire & Clean limit while the MS is acquiring can perturb the ion source and affect data quality. The flow step size is 0.069 sccm, so setpoints should be chosen as multiples of this step.

Operating Manual JetClean G7077-90052, Configuration and Parameter Setpoints
Electrical
Emission current limited to 10–35 µA during cleaning

The emission current setpoint range is 10–35 µA. Operating outside this range may not sustain a stable plasma or may damage the source. The current must be set in conjunction with the source temperature (150–350 °C) to achieve effective cleaning without overheating the ion source.

Operating Manual JetClean G7077-90052, Parameter Setpoints
Maintenance
Hydrogen supply plumbing must be leak‑checked

The JetClean system uses hydrogen, which is flammable. The operating manual covers hydrogen supply plumbing and troubleshooting. Before initial use and after any plumbing change, verify all connections for leaks. Use only hydrogen‑compatible tubing and fittings. The manual does not specify a maintenance interval for the JetClean itself, but regular checks of the hydrogen supply are implied.

Operating Manual JetClean G7077-90052, Hydrogen Supply Plumbing

Selection & Integration Q&A

When should I choose the JetClean system instead of a sibling model like the 5977C GC/MSD or 7000E triple quadrupole?

The JetClean is an accessory for in‑situ source cleaning, not a standalone GC/MS. Choose it when you already have a compatible Agilent MS (5975/5977, 7000E/7010C) and want to extend source maintenance intervals. The 5977C, 7000E, 7010D, and 7250 are complete instruments; select those when you need a full GC/MS system, not just cleaning capability.

What are the requirements for the hydrogen supply and how does the JetClean connect to the MS?

The JetClean requires a hydrogen supply with a flow step size of 0.069 sccm and must be plumbed to the ion source region. The operating manual (G7077‑90052) covers hydrogen supply plumbing and configuration. The system is controlled through the host MS software (MassHunter). Ensure the hydrogen supply is regulated and leak‑tested before connection.

What maintenance is required and what consumables or accessories are not included in the base delivery?

The base delivery includes the JetClean module and documentation, but hydrogen supply plumbing, tubing, and fittings are not included and must be sourced separately. The manual covers troubleshooting and hydrogen supply plumbing but does not specify a regular maintenance schedule for the JetClean itself. Periodic leak checks of the hydrogen supply are recommended.

Dimensional Drawings & Technical Documentation

Technical documentation