Agilent Technologies icp-ms-automation-systems
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies icp-ms-automation-systems

Agilent Technologies icp-ms-automation-systems — Atomic spectroscopy.

The Agilent ICP Automation System is a two-syringe autodilutor integrated with a four-rack autosampler, delivering 360-sample capacity and up to 400-fold dilution from a single stock solution. It doubles throughput versus the SPS 4 and is controlled by ICP Expert 7.7+ or MassHunter 5.3+. It serves automated sample introduction and dilution for 5800/5900 ICP-OES and 7850/7900/8900 ICP-MS.

Agilent Technologies Automation Systems Series — Model Comparison & Selection

Model Sample capacityWidth, cm
icp-ms-automation-systems Current 36015.8
icp-oes-automation-systems 36015.8

Full specifications

Core Performance & Functional Specifications
Sample Capacity 360 samples
Maximum dilution factor mfr 400 fold
Dilution range of single stock solution mfr 1 x
Autosampler rack capacity mfr four sample rack rack
Sample capacity increase vs SPS 4 mfr 50
Sample throughput mfr doubles doubles
Compatible ICP-OES instruments mfr 5800
Compatible ICP-MS instruments mfr 7850
ICP-OES software version mfr 7.7 above
ICP-MS software version mfr 5.3 above
Syringe configuration mfr two-syringe syringe
Maximum dilution factor (single stock solution) mfr 1 x
Sample rack count mfr six sample rack rack
Sample throughput increase mfr doubles doubles
maximum dilution factor mfr 400 fold
SPS 6 sample capacity mfr 540 samples
Dimensions, Weight & Mechanical Design
Width 15.8 cm

Description

Construction and series positioning

Behavior and media

Control and integration

Integration & Operating Notes

Control & I/O
Software version floor gates the autodilutor

The ADS 2 dilutor is driven by the instrument software, not a standalone controller: ICP Expert must be v7.7 or higher on 5800/5900 ICP-OES and MassHunter v5.3 or higher on 7850/7900/8900 ICP-MS. Upgrading the instrument software is therefore a prerequisite before the hardware can be commissioned.

te ads 2 icp 5994 7211en agilent, scope and variants
Process Limits
Dilution limited to 400-fold from one stock

The two-syringe design prepares all dilutions from a single stock solution with a 1x range and a maximum factor of 400. Samples requiring higher dilution must be pre-diluted offline or run through a second preparation step; planning the calibration range around this ceiling avoids failed runs.

Factory Specification Sheet, extra specs
Maintenance
Syringe and valve consumables are not in base delivery

The two-syringe dilutor and the AVS switching valve are separate modules in the workflow; syringe seals, tubing and valve rotor are wear items to be stocked and replaced on a service schedule. Confirm which of these are included with the base system versus ordered as accessories.

brochure icp automation 5994 6941ENus agilent, scope and variants

Selection & Integration Q&A

When should I choose this ICP automation system instead of the ICP-OES automation sibling?

Pick this system when the workflow spans both ICP-OES (5800/5900) and ICP-MS (7850/7900/8900) and needs on-line dilution up to 400-fold with 360-sample capacity. The ICP-OES sibling covers the OES-only automation path; choose it when no ICP-MS instrument is in the lab.

What adjacent modules and software must be in place for the autodilutor to run?

The ADS 2 dilutor works with the Advanced Valve System (AVS 7 or AVS MS) as the switching stage and requires ICP Expert v7.7+ (ICP-OES) or MassHunter v5.3+ (ICP-MS). The instrument software version must be verified before commissioning.

Which consumables and accessories are not part of the base delivery?

The two-syringe dilutor and AVS switching valve are separate workflow modules; syringe seals, tubing and valve rotor are wear items to be stocked. Confirm rack types and valve configuration against the order, as the brochure lists ADS 2, AVS 7/AVS MS and SPS 4/SPS 6 as distinct items.

Dimensional Drawings & Technical Documentation

Dimensional drawings

icp-ms-automation-systems — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Agilent ICP Workflow Automation System
Brochure ·9,3 МБ ·18 pp.
View PDF