Agilent Technologies 1,2, or 3 Laser Benchtop Flow Cytometer, NovoCyte Advanteon
Agilent Technologies 1,2, or 3 Laser Benchtop Flow Cytometer, NovoCyte Advanteon — Research flow cytometry.
The NovoCyte Advanteon is a benchtop flow cytometer configured with one to three solid-state lasers, each with onboard thermal-electric cooling, and a fixed 170 µm flow cell. It acquires up to 100,000 events/s at sample flow rates of 5–120 µL/min, with volumetric absolute counting at 5% precision. It serves as the entry-to-mid configuration of the NovoCyte series for routine immunophenotyping and volumetric cell analysis.
Agilent Technologies Flow Cytometer Systems Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Sample flow rate, µL/min | Sheath flow rate, mL/min | Volumetric absolute count precision, % | Laser power, W | Print technology |
|---|---|---|---|---|---|---|
| 1,2, or 3 Laser Benchtop Flow Cytometer, NovoCyte Advanteon Current | — | 5...120 | 6.5 | 5 | 0.405 | Solid state laser with o |
| 1-3 laser benchtop flow cytometer, NovoCyte | 3×10⁻⁶ | 5...120 | 6.5 | 5 | — | — |
| 4 Laser Flow Cytometer, NovoCyte Quanteon | — | 5...120 | 6.5 | 5 | 0.405 | Solid state laser with o |
| 5 Laser Flow Cytometer, NovoCyte Penteon | — | 5...120 | 6.5 | 5 | 0.405 | Solid state laser with o |
Full specifications
| Laser power | 405 mW |
| Print technology | Solid state laser with onboard thermal-electric cooling |
| Sample Flow Rate | 5...120 μL/min |
| Sheath Flow Rate | 6.5 mL/min |
| Ssc Sensitivity | 0.1 μm |
| Volumetric Absolute Count Precision | 5 % |
| Laser beam profile mfr | 10 μm |
| Fluorescence detection mfr | Silicon photomultiplier with individual photodet |
| FSC sensitivity mfr | 0.4 μm |
| Fluorescence threshold sensitivity mfr | 40 MESF |
| Fluorescence resolution mfr | 3 % CV |
| Flow cell mfr | 170 μm |
| Sample acquisition rate mfr | 100000 events/second |
| Sample acquisition volume mfr | 5 mL |
| Manual sample format mfr | 12 tube |
Description
Construction and series positioning
The Advanteon is the 1–3 laser member of the NovoCyte family; the Quanteon carries 4 lasers and the Penteon 5. Fluorescence is detected by silicon photomultipliers with an individual photodetector per channel, and the optical bench uses solid-state lasers with onboard thermal-electric cooling. The 170 µm flow cell and 10 µm laser beam profile define the core optical geometry. Choose the Advanteon when the panel fits within three laser lines; step to the Quanteon or Penteon only when additional excitation lines are required.
Analytical behavior
Sensitivity is specified at 0.4 µm for FSC and 0.1 µm for SSC, with a fluorescence threshold of 45 MESF and fluorescence resolution of 3% CV. Sample acquisition rate reaches 100,000 events/s, while the sample flow rate is adjustable from 5 to 120 µL/min and the sheath flow rate is 6.5 mL/min. Volumetric absolute counting is performed directly with 5% precision, so no reference beads are needed for concentration readout. The 5 mL sample acquisition volume and 12-tube manual format set the per-run workload.
Installation and operation
The datasheet digest covers the NovoCyte Penteon, Quanteon and Advanteon together and provides no installation, connection or control details specific to this model; those must be taken from the instrument user manual. No certificates or application notes were supplied with the verified facts.
Integration & Operating Notes
Samples concentrated enough to exceed 100,000 events/s will be acquired at the instrument ceiling, and any coincident-event rejection at that point biases the population statistics. Dilute high-concentration samples or reduce the sample flow rate before acquisition so the event rate stays below the stated maximum.
The sample flow rate is adjustable from 5 to 120 µL/min against a fixed sheath flow rate of 6.5 mL/min. Volumetric absolute counts are derived from the sample flow setting, so the chosen rate must remain stable throughout the acquisition for the 5% volumetric precision to hold.
The supplied datasheet digest contains no maintenance intervals, error codes or service procedures for the Advanteon. The fluidic system, flow cell and thermal-electric laser cooling require scheduled service defined in the instrument user manual; obtain that manual before commissioning.
No supply voltage, current or power figures for the Advanteon were included in the verified specifications. Site power planning must rely on the instrument user manual or nameplate rather than on this card.
Selection & Integration Q&A
Pick the Advanteon when the panel needs one to three laser lines; it is the 1–3 laser configuration of the NovoCyte series. Move to the Quanteon (4 lasers) or Penteon (5 lasers) only when additional excitation lines are required for the panel.
The manual sample format is 12 tubes with a 5 mL sample acquisition volume. Upstream preparation must deliver single-cell suspensions in that tube format; the datasheet digest gives no plate-handler or automation interface details for this model.
The verified facts contain no duty-cycle, ambient temperature or humidity limits for this model. The datasheet digest covers Penteon, Quanteon and Advanteon jointly without environment data, so these limits must be read from the instrument user manual before siting.
No maintenance schedule, sheath fluid, cleaning reagents, QC beads or cables are listed in the verified facts. Service intervals and consumable part numbers are defined in the user manual, which was not supplied; confirm them with the vendor before commissioning.