Agilent Technologies 4 Laser Flow Cytometer, NovoCyte Quanteon
Agilent Technologies 4 Laser Flow Cytometer, NovoCyte Quanteon — Research flow cytometry.
The NovoCyte Quanteon is a 4-laser solid-state flow cytometer with onboard thermal-electric laser cooling, silicon photomultiplier detection per fluorescence channel, and volumetric absolute counting at 5% precision. It acquires up to 100,000 events/second across a 5–120 µL/min sample flow range, positioning it between the 1–3 laser Advanteon and the 5-laser Penteon in the NovoCyte series.
Agilent Technologies Flow Cytometer Systems Series — Model Comparison & Selection
| Model | Pumping speed, m³/h | Sample flow rate, µL/min | Sheath flow rate, mL/min | Volumetric absolute count precision, % | Laser power, W | Print technology |
|---|---|---|---|---|---|---|
| 4 Laser Flow Cytometer, NovoCyte Quanteon Current | — | 5...120 | 6.5 | 5 | 0.405 | Solid state laser with o |
| 1,2, or 3 Laser Benchtop Flow Cytometer, NovoCyte Advanteon | — | 5...120 | 6.5 | 5 | 0.405 | Solid state laser with o |
| 1-3 laser benchtop flow cytometer, NovoCyte | 3×10⁻⁶ | 5...120 | 6.5 | 5 | — | — |
| 5 Laser Flow Cytometer, NovoCyte Penteon | — | 5...120 | 6.5 | 5 | 0.405 | Solid state laser with o |
Full specifications
| Laser power | 405 mW |
| Print technology | Solid state laser with onboard thermal-electric cooling |
| Sample Flow Rate | 5...120 μL/min |
| Sheath Flow Rate | 6.5 mL/min |
| Ssc Sensitivity | 0.1 μm |
| Volumetric Absolute Count Precision | 5 % |
| Laser beam profile mfr | 10 μm |
| Fluorescence detection mfr | Silicon photomultiplier with individual photodet |
| FSC sensitivity mfr | 0.4 μm |
| Fluorescence threshold sensitivity mfr | 40 MESF |
| Fluorescence resolution mfr | 3 % CV |
| Flow cell mfr | 170 μm |
| Sample acquisition rate mfr | 100000 events/second |
| Sample acquisition volume mfr | 5 mL |
| Manual sample format mfr | 12 tube |
Description
Construction and series positioning
- The Quanteon is the 4-laser configuration of the NovoCyte platform; the Advanteon covers 1–3 lasers and the Penteon 5 lasers, so the Quanteon is the mid-tier optical package.
- Lasers are solid-state with onboard thermal-electric cooling; the beam profile is specified at 10 µm.
- Fluorescence detection uses a silicon photomultiplier with an individual photodetector for each channel.
- The flow cell is specified at 170 µm.
Measurement behavior
- Scatter sensitivity: FSC 0.4 µm, SSC 0.1 µm.
- Fluorescence threshold sensitivity 45 MESF; fluorescence resolution 3 % CV.
- Sheath flow rate 6.5 mL/min; sample flow rate adjustable 5–120 µL/min; sample acquisition volume 5 mL.
- Sample acquisition rate up to 100,000 events/second; volumetric absolute count precision 5 %.
- Manual sample format: 12 tubes.
Installation and operation
- The attached datasheet digest covers the NovoCyte Penteon, Quanteon and Advanteon together and contains no section-level installation, connection or control facts; no fluidic, electrical supply, interface or environmental figures are verified for this model, so they are not stated here.
Integration & Operating Notes
Sample acquisition rate is specified up to 100,000 events/second; samples concentrated enough to exceed this event rate at the chosen sample flow will exceed the stated acquisition capability, so dilution or a lower sample flow rate must be planned before running high-concentration samples.
The verified manual sample format is 12 tubes and the sample acquisition volume is 5 mL. The attached datasheet digest contains no loader, automation, cleaning or service-interval facts, so automation accessories and maintenance schedules cannot be specified from the verified material.
Selection & Integration Q&A
Pick the Quanteon when the panel needs four laser lines; the Advanteon covers 1–3 lasers and the Penteon 5. All three share the NovoCyte platform, so the choice is optical package size, not a different detection principle.
Volumetric absolute count precision is 5% with sample flow adjustable 5–120 µL/min and a 5 mL acquisition volume; the sample must be delivered within this flow window, and the 12-tube manual format defines the sample presentation.
The verified datasheet digest contains no environmental, bench-load or clearance figures for the Quanteon, so site requirements cannot be quoted from the supplied facts; they must be taken from the full instrument manual before installation planning.
No service intervals, cleaning routines, cables or consumables are stated in the verified datasheet digest; only the 12-tube manual sample format and 5 mL acquisition volume are confirmed, so accessories and service plans must be confirmed separately.