No photo available yet
Agilent Technologies PCG 752
Agilent Technologies PCG 752 — Vacuum technologies.
The PCG 752 is a combination Pirani/capacitance diaphragm gauge with a nickel-plated filament, covering 5×10⁻⁵ to 1500 mbar via two constantly active sensors. It is the nickel-filament variant of the PCG-750 series, sharing the NW 16KF flange and logarithmic 0.61–10.23 V output. It serves as a single wide-range vacuum sensor for rough-to-fine processes, with relay setpoints and optional Profibus.
Agilent Technologies Active Gauges Series — Model Comparison & Selection
| Model | Measurement range | Operating temperature, °C | Storage temperature, °C | Internal volume, microliter | Operation | Response time, ms |
|---|---|---|---|---|---|---|
| PCG 752 Current | 5 | 10 | 20 | 4.7 | 10 | 30 |
| FRG-700/702 | 1.82...8.6 | 5 | 40 | 20 | 29 | 10 |
Full specifications
| Measurement range | 5 mbar |
| Data Rate | 12 Mbaud |
| Diffusion Pump Pumping Speed | 50 L/s |
| Internal Volume | 4.7 microliter |
| Operation | 10 °C |
| Response Time | 30 ms |
| Switching Time | 30 ms |
| Vacuum connection | 1.4435 millimeter |
| Vacuum gauge measurement range mfr | 6 pressure |
| Dry scroll pump typical size mfr | 3...30 h |
| Turbomolecular pump pumping speed mfr | 70...3 L/s |
| Measurement principle mfr | thermal conductance acc. to Pirani Pirani |
| Accuracy (N2) mfr | 2.5 reading |
| Repeatability (N2) mfr | 1 reading |
| Output impedance mfr | 2 proof |
| Load impedance mfr | 10 |
| Gauge identification mfr | 71.5 |
| HV adjustment mfr | 10 mbar |
| ATM adjustment mfr | 100 mbar |
| Switching functions setting range mfr | 5.0 mbar |
| Hysteresis mfr | 10 threshold |
| Switching characteristics mfr | Low Trip Point Point |
| Switching function type mfr | 1 function |
| Contact rating closed mfr | 30 resistive |
| Contact rating open mfr | LED lit solid solid |
| Ripple mfr | 1 Vpp |
| Sensor cable mfr | 0.14 conductor |
| Grounding concept vacuum connection to signal common mfr | 10 nF |
| RS232C data format mfr | 8 handshake |
| Materials exposed to vacuum - feedthrough mfr | glass glass |
| Materials exposed to vacuum - orifice mfr | stainless steel steel |
| Materials exposed to vacuum - diaphragm mfr | ceramic ceramic |
| Materials exposed to vacuum - further materials mfr | 1.4301 SnAg |
| Permissible pressure (absolute) mfr | 5 bar |
| Bursting pressure (absolute) mfr | 10 bar |
| Use mfr | 2000 NN |
| Grounding concept mfr | 10 nF |
| Materials exposed to vacuum - Feedthrough mfr | glass glass |
| Materials exposed to vacuum - Orifice mfr | stainless steel steel |
| Materials exposed to vacuum - Diaphragm mfr | ceramic ceramic |
| Materials exposed to vacuum - Further materials mfr | 1.4301 SnAg |
| AGPS45INT mfr | Power supply, PVG-PCG, international plug. plug. |
| PCG750KF16 mfr | -750 KF. |
| PCG750KF16RS mfr | -750 KF. |
| PCG750KF16SD1 mfr | -750 mbar. |
| PCG750KF16SD2 mfr | -750 Torr. |
| PCG750KF16SD3 mfr | -750 Pascal. |
| PCG750KF16SP mfr | -750 mbar. |
| PCG752KF16 mfr | -752 KF. |
| PCG752KF16SD1 mfr | -752 mbar. |
| PCG752KF16SD2 mfr | -752 Torr. |
| PCG752KF16SD3 mfr | -752 Pascal. |
| PCG752KF16SP mfr | -752 mbar. |
| Default setting mfr | 0.1 Chex |
| PCG-752 mfr | Ni Ni |
| Feedthrough mfr | glass glass |
| Orifice mfr | stainless steel steel |
| Diaphragm mfr | ceramic ceramic |
| Further materials mfr | 1.4301 SnAg |
| Year's mean mfr | 65 |
| During 60 days mfr | 85 |
| Grounding concept Vacuum connection to signal common mfr | 10 nF |
| Data format mfr | 8 handshake |
| About this manual mfr | 6 |
| Validity mfr | 6 |
| Definitions and terminology mfr | 7 |
| Definition of Caution, Warning and Note mfr | 7 |
| Warning Symbols mfr | 8 |
| Safety mfr | 9 |
| Proper use mfr | 9 |
| Improper use mfr | 10 |
| Safety guideline for PCG-750, PCG-752 mfr | 11 |
| Disposal mfr | 12 |
| Service mfr | 13 |
| Cable length | 100 m |
| RS232C operation cable length mfr | 30 m |
| RS232C transmission rate mfr | 57600 baud |
| Mounting orientation mfr | any any |
| Cable length, system wiring mfr | according to Profibus specifications specificati |
| Cable length FCC 68 mfr | 30 m |
| Transmission rate mfr | 57600 baud |
| RS232C cable length mfr | 30 m |
| Voltage vs. pressure mfr | 1.286 logarithmic |
| Supply voltage at the gauge mfr | 15 VDC |
| Fuse to be connected mfr | 1 AT |
| Fuse to be connected 3) mfr | 1 AT |
| Profibus Data Rate | 12 Mbaud |
| Turbomolecular pump controller protection class mfr | 54 |
| Output signal measurement range mfr | 0.61 V |
| Diagnostic port mfr | 2.5 pin |
| Power consumption without fieldbus mfr | 2.5 W |
| Power consumption with fieldbus mfr | 3 W |
| Profibus interface physical mfr | 485 |
| Profibus node address default setting mfr | 0.1 Chex |
| Profibus connection cable mfr | 9 cable |
| Profibus cable length, system wiring mfr | according to Profibus specifications specificati |
| Weight w/o fieldbus interface mfr | 115 g |
| Weight with fieldbus interface mfr | 230 g |
| Profibus connection mfr | 9 |
| w/o fieldbus interface mfr | 115 g |
| with fieldbus interface mfr | 230 g |
| Specification, data format, communication protocol mfr | 485 |
| Weight without fieldbus interface mfr | 115 g |
| Materials exposed to vacuum - filament PCG-752 mfr | Ni Ni |
| Materials exposed to vacuum - Filament PCG-752 mfr | Ni Ni |
| Filament mfr | 160 °C |
| Operating temperature | 10 °C |
| Protection rating | 40 |
| Storage temperature | 20 °C |
| Relative humidity year's mean mfr | 65 |
| Relative humidity during 60 days mfr | 85 |
| Relative humidity - Year's mean mfr | 65 |
| Relative humidity - During 60 days mfr | 85 |
Description
Construction and series positioning
- The PCG 752 combines a Pirani thermal-conductance sensor and a diaphragm capacitive sensor in one NW 16KF-flanged body; both sensors are constantly active, with the Pirani signal used below ~1 mbar, the capacitive signal above ~10–100 mbar (depending on gas), and a proportional blend in between.
- It differs from the PCG-750 only in filament material (nickel-plated vs tungsten); both share the same measurement range, accuracy, output, and electrical interface. The PCG-752 is therefore chosen where the nickel filament's compatibility or lifetime is preferred, not for different performance.
- Within the Active Gauges series, it is the wide-range combination gauge; sibling FRG-700/702 and FRG-705/707 are full-range Bayard-Alpert Pirani gauges for UHV, while CDG-500 is a high-precision capacitance diaphragm gauge for single-range use.
Behavior across the measuring range
- Accuracy is pressure-dependent: ±50% of reading at 5×10⁻⁴…1×10⁻³ mbar, ±15% at 1×10⁻³…100 mbar, ±5% at 100…950 mbar, and ±2.5% at 950…1050 mbar; repeatability is ±2% over 1×10⁻³…1100 mbar.
- The output is logarithmic: U = c + 1.286 log₁₀ p, with a measurement-range voltage of +0.61…+10.23 V; error signal defaults to 0 V. Response and switching times are both <30 ms.
- In the Pirani range below 1 mbar the reading is gas-type dependent and must be corrected with calibration factors (e.g., Ar 1.7, He 0.8, H₂ 0.5); in the capacitive range the measurement is gas-independent.
Installation, control and operation
- The gauge requires a SELV supply of +15…+30 VDC with ≤1 Vpp ripple, fused at 1 AT; power consumption is ≤2.5 W without fieldbus and ≤3 W with fieldbus.
- Two floating relay setpoints (SP1/SP2) are factory set to the lower range limit and do not switch until configured; hysteresis is 10% of threshold, contact rating <30 VAC/DC at ≤1 A resistive.
- A 2.5 mm 3-pin RS232C diagnostic port (57600 baud, binary, 8 data bits, 1 stop bit, no parity) allows reading and parameter entry; Profibus variants use a 9-pin D-Sub at up to 12 Mbaud and require the GSD file and a unique node address.
- Mounting orientation is free but vibration-free; the ceramic sensor is fragile and the vacuum connection must not exceed 80 °C. Allow ≥10 minutes stabilization after power-up.
Integration & Operating Notes
The gauge is designed for a grounded protective extra-low voltage supply; connecting to a non-SELV or unfused source risks damage and violates the manual's electrical safety requirements. Agilent controllers meet this, but third-party supplies must be checked for 15–30 VDC output, ≤1 Vpp ripple, and a 1 AT fuse in the feed line.
Permissible absolute pressure is ≤5 bar and bursting pressure is 10 bar. Exceeding 5 bar risks sensor damage and unsafe operation; use overpressure-rated clamps and O-rings with outer centering ring for the KF connection, and never use the gauge with flammable or combustible gas mixtures in the explosion range.
SP1 and SP2 are factory set to the lower measurement range limit and therefore do not switch by default. Any interlock or valve control wired to the relay contacts will remain inactive until the thresholds are set via the diagnostic port or display. Also avoid pressing the <SP> button during process control, as it suppresses the measurement signal and outputs the threshold value instead.
Any orientation is permitted, but the gauge must not be subjected to vibration, so avoid direct mounting on roughing pumps or vibrating lines. The vacuum connection temperature must stay ≤80 °C; if bakeout or hot process lines are involved, provide thermal isolation or remote mounting to keep the flange within this limit.
On Profibus variants, the RS232C diagnostic port must not be operated while the Profibus interface is active. The GSD file must be installed on the bus master before commissioning, and the node address (0–125 dec, default 5Chex) is set via hexadecimal switches and polled only at power-on, so address changes require a power cycle.
The gauge is factory calibrated, but zero drift can occur with long operation or contamination. ATM adjustment should be done at atmospheric pressure after at least 10 minutes; HV adjustment after evacuating to p << 10⁻⁵ mbar and waiting at least 2 minutes. For severe contamination or malfunction, the sensor is replaceable (PCG752KF16RS); always quote the nameplate data when ordering spares.
In the Pirani range below 1 mbar, readings for gases other than air must be multiplied by calibration factors (e.g., Ar 1.7, CO₂ 0.9, H₂ 0.5). Above ~10 mbar the capacitive sensor is gas-independent, so no correction is needed there. This matters when the gauge is used for process gases rather than air/N₂.
Selection & Integration Q&A
Choose the PCG 752 when a nickel-plated filament is preferred over the PCG-750's tungsten filament; performance and range are identical. For UHV below 5×10⁻⁵ mbar, use an FRG-700/702 or FRG-705/707 Bayard-Alpert Pirani gauge instead, as the PCG series stops at 5×10⁻⁵ mbar.
A SELV supply of +15…+30 VDC with ≤1 Vpp ripple, fused at 1 AT. Use shielded sensor cable with 0.14 mm² conductors, max 100 m; the RS232C cable is limited to 30 m. Connect the shield to chassis ground on one side to avoid ground loops, and keep potential difference between supply common and housing ≤18 V.
No. The gauge must be mounted so that no vibrations occur, so avoid direct mounting on pumps or vibrating lines; use vibration isolation. It is rated IP 40 for indoor use only, altitude ≤2000 m, humidity ≤65% annual mean (≤85% for 60 days) without condensation, and ambient +10…+50 °C.
The base delivery does not include a power supply (AGPS45INT is an accessory), nor cables or a display. Plan for periodic ATM/HV adjustment via the <ADJ> button and replacement of the sensor (PCG752KF16RS) if contaminated. The RS232C diagnostic port requires a 2.5 mm 3-pin cable; Profibus variants need the GSD file and a 9-pin D-Sub cable.