No photo available yet
Agilent Technologies FRG-700/702
Agilent Technologies FRG-700/702 — Vacuum technologies with Max pressure 1 megapascal.
The FRG-700/702 is a combined Pirani and inverted-magnetron cold-cathode gauge covering 5×10⁻⁹ to 1000 mbar with a single logarithmic 0.6 V/decade output. The FRG-700 uses FPM seals for elastomer-sealed KF/CF installations; the FRG-702 is the all-metal, 150 °C bakeable variant. It serves as the wide-range pressure sensor in rough-to-UHV pumping systems.
Agilent Technologies Active Gauges Series — Model Comparison & Selection
| Model | Measurement range | Operating temperature, °C | Storage temperature, °C | Internal volume, microliter | Operation | Response time, ms |
|---|---|---|---|---|---|---|
| FRG-700/702 Current | 1.82...8.6 | 5 | 40 | 20 | 29 | 10 |
| PCG 752 | 5 | 10 | 20 | 4.7 | 10 | 30 |
Full specifications
| Max pressure | 10 bar |
| Measurement range | 1.82...8.6 V |
| Diffusion Pump Pumping Speed | 50 L/s |
| Electrical connection | 28 |
| Internal Volume | 20 cm³ |
| Operation | 29 |
| Response Time | 10 ms |
| Vacuum connection | stainless steel |
| Vacuum gauge measurement range mfr | 6 pressure |
| Dry scroll pump typical size mfr | 3...30 h |
| Turbomolecular pump pumping speed mfr | 70...3 L/s |
| Measuring range (air, N2) mfr | -9...1000 mbar |
| Accuracy (N2) mfr | 30 |
| Reproducibility mfr | 5 |
| Error signal mfr | 0.5 supply |
| Output impedance mfr | 2 |
| Minimum loaded impedance mfr | 10 proof |
| Gauge identification mfr | 85 common |
| Adjustment <HV> potentiometer mfr | 10 mbar |
| Adjustment <ATM> potentiometer mfr | at atmospheric pressure pressure |
| Sensor cable conductors mfr | 8 shielding |
| Grounding concept: supply common-signal common mfr | 10 |
| Materials exposed to vacuum: Vacuum connection mfr | stainless steel steel |
| Materials exposed to vacuum: Measuring chamber mfr | stainless steel steel |
| Materials exposed to vacuum: Feedthrough mfr | ceramic ceramic |
| Materials exposed to vacuum: Internal seals mfr | Ag, Cu, soft solder solder |
| Materials exposed to vacuum: Anode mfr | Mo Mo |
| Materials exposed to vacuum: Ignition aid mfr | stainless steel steel |
| Materials exposed to vacuum: Pirani measuring tube mfr | Ni, Au Au |
| FRG700KF25 mfr | 25 KF |
| FRG702KF25 mfr | 25 KF |
| FRG700CF35 mfr | 40 C |
| FRG702KF40 mfr | 40 KF |
| FRG702CF35 mfr | 40 C |
| p > 10-6 mbar mfr | 10 ms |
| p = 10-8 mbar mfr | 1000 ms |
| p > 10-2 mbar mfr | Pirani-only mode mode |
| p < 10-2 mbar mfr | Cold cathode ignited ignited |
| Lamp mfr | 6 |
| Fuse1) mfr | 1 AT |
| Adjustment mfr | 10 mbar |
| Sensor cable mfr | 68 poles |
| Vacuum connection-signal common mfr | see paragraoh Electrical Connection Connection |
| Supply common-signal common mfr | 10 |
| Measuring chamber mfr | stainless steel steel |
| Feedthrough mfr | ceramic ceramic |
| Internal seals mfr | 75 |
| Anode mfr | Ag, Cu, soft solder solder |
| Ignition aid mfr | Mo Mo |
| Pirani measuring tube mfr | stainless steel steel |
| Product Description mfr | 16 |
| Technical Specifications mfr | 17 |
| FRG-700 and FRG-702 Outline mfr | 20 |
| Installation mfr | 24 |
| Removing the Magnet Unit (only for Gauges with CF Flanges) mfr | 27 |
| Use With Another Control Device mfr | 28 |
| Adjusting the Gauge mfr | 32 |
| Deinstallation mfr | 34 |
| Cleaning FRG-700 mfr | 35 |
| Replacing the Pirani element mfr | 36 |
| Cleaning FRG-700, Replacing Parts mfr | 36 |
| Disassembling FRG-700 mfr | 37 |
| Reassembling FRG-700 mfr | 38 |
| Cleaning FRG-702 mfr | 39 |
| Cleaning FRG-702, Replacing Parts mfr | 40 |
| Disassembling FRG-702 mfr | 41 |
| Reassembling FRG-702 mfr | 42 |
| Troubleshooting mfr | 44 |
| Spare Parts mfr | 45 |
| p < 1x10-2 mbar mfr | Cold cathode ignited ignited |
| About this manual mfr | 6 |
| Validity mfr | 6 |
| Definitions and terminology mfr | 7 |
| Definition of Caution, Warning and Note mfr | 7 |
| Warning Symbols mfr | 8 |
| Safety mfr | 10 |
| Proper use mfr | 10 |
| Improper use mfr | 11 |
| Safety guideline for FRG-700 and FRG-702 mfr | 12 |
| Disposal mfr | 13 |
| Service mfr | 14 |
| Response time p = 10-8 mbar mfr | 1000 ms |
| Weight | 700 g |
| Sensor cable line length mfr | 50 m |
| Weight FRG702KF25 mfr | 730 g |
| Weight FRG700CF35 mfr | 980 g |
| Weight FRG702KF40 mfr | 750 g |
| Weight FRG702CF35 mfr | 1010 g |
| Line length mfr | 8 m |
| Mounting orientation mfr | any any |
| Power consumption | 2 W |
| Voltage vs. pressure mfr | 0.6 decade |
| Supply voltage at the gauge mfr | 15.0...30.0 VDC |
| Fuse mfr | 1 AT |
| Voltage at the supply unit with maximum line length mfr | 16.0...30.0 VDC |
| Operating voltage mfr | 3.3 kV |
| Operating current mfr | 500 μA |
| Grounding concept voltage differential with respect to safety mfr | 50 V |
| Grounding concept voltage differential with respect to accuracy mfr | 10 V |
| Turbomolecular pump controller protection class mfr | 54 |
| Output signal measuring range mfr | 1.82...8.6 V |
| Sensor cable connector mfr | 68 poles |
| Use With an Agilent Controller mfr | 28 |
| Gas type dependence mfr | 21 |
| Pirani filament mfr | Ni, Au Au |
| Bakeout Temperature | 150 °C |
| Operating temperature | 5 °C |
| Protection rating | 40 IP |
| Storage temperature | 40 °C |
| Relative humidity mfr | 80 °C |
Description
Construction and series positioning
- Two Pirani/cold-cathode sensor heads in one body: Pirani always active, cold cathode ignited below 1×10⁻² mbar; combined output is logarithmic, 0.6 V/decade, 1.82–8.6 V measuring band.
- FRG-700 (FPM-sealed) is limited to +5…+55 °C operating and is cleaned/rebuilt with elastomer seals; FRG-702 (all-metal, copper-sealed Pirani element) extends operating temperature to +5…+150 °C and is the choice for bakeout duty.
- Flange variants: FRG700KF25 (DN 25 ISO-KF, ~700 g), FRG700CF35 (DN 40 CF-C, ~980 g), FRG702KF25 (~730 g), FRG702KF40, FRG702CF35 (~1010 g). Siblings FRG-720/730 (Bayard-Alpert Pirani) and FRG-705/707 cover other ranges; CDG-500 is the high-precision diaphragm alternative.
Behavior across the range
- Accuracy ≈±30 % (N₂), reproducibility ≈±5 % between 1×10⁻⁸ and 100 mbar; below 1×10⁻² mbar the cold cathode takes over and ignition delay grows from ≈1 s at 10⁻⁵ mbar to ≈2 min at 5×10⁻⁹ mbar.
- Do not start the gauge below 3×10⁻⁹ mbar — ignition cannot be verified there. Status output (pin 6) stays Low (0 V) in Pirani-only mode and goes High (15–30 VDC) after cold-cathode ignition.
- Continuous operation in 10⁻⁴…10⁻² mbar causes severe contamination; below 1×10⁻⁶ mbar the gauge can run over a year without cleaning.
Installation, control and operation
- Any mounting orientation; horizontal-to-upright preferred to keep condensates and particles out. KF connections need a conductive metallic clamping ring for protective grounding per EN 61010; CF connections satisfy this by design.
- Supply 15.0–30.0 VDC (ripple ≤1 Vpp), ≤2 W, fuse ≤1 AT; at maximum 50 m cable length the supply unit must deliver 16.0–30.0 VDC. Sensor cable: 8×0.14 mm² shielded, FCC68 8-pole female; third-party controllers need a cable wired to the diagram.
- Pirani circuit is adjustable via <HV> and <ATM> potentiometers; the cold-cathode circuit is factory-calibrated and not adjustable. Allow ≈10 min stabilization and leave the gauge powered permanently to shorten ignition delay.
Integration & Operating Notes
The gauge draws ≤2 W but the cold-cathode circuit runs at up to 3.3 kV / 500 µA internally, so the feed must be a SELV source with ≤1 Vpp ripple and a ≤1 AT fuse. At the full 50 m cable length, voltage drop means the supply unit must be set to 16.0–30.0 VDC, not the 15.0 V minimum at the gauge.
With KF connections, protective grounding per EN 61010 is only achieved through a conductive metallic clamping ring; elastomer-sealed KF joints also cannot withstand more than 2.5 bar. CF flange variants meet the grounding requirement inherently and are the safer choice where overpressure or fault currents are possible.
Continuous running in this decade contaminates the cold-cathode electrode and shortens uptime; below 1×10⁻⁶ mbar the gauge can operate more than a year without cleaning. Mount at a low partial pressure of pollutants and use geometric protection (screens, elbows) if the process crosses this band regularly.
At pressures below 3×10⁻⁹ mbar the electronics cannot distinguish whether the cold-cathode discharge has ignited, so the reading and status output are unreliable. Sequence the start-up so pressure is slightly above this threshold, or leave the gauge permanently powered so ignition delay drops below 1 s.
Helium entering the gauge during a tightness test can cause electric arcing that destroys the electronics. Put the product out of operation and remove the electronics unit before any helium leak test; after sensor reassembly, verify anode-pin centering (≤0.5 mm eccentricity) and leak-test to <10⁻⁹ mbar l/s.
After sensor exchange, extreme temperature exposure, aging or contamination, only the Pirani side can be readjusted via the <HV> and <ATM> potentiometers; the cold-cathode circuit is factory-calibrated. Adjustment has negligible effect between roughly 10⁻² and 10² mbar, so calibrate at the appropriate set points.
Selection & Integration Q&A
Pick FRG-702 (all-metal, copper-sealed Pirani element) when the flange must be baked to 150 °C or operating temperature exceeds +55 °C; FRG-700 with FPM seals is limited to +5…+55 °C. Choose FRG-720/730 if you need Bayard-Alpert Pirani behavior at the high-vacuum end; CDG-500 for high-precision diaphragm measurement.
Provide a SELV 15.0–30.0 VDC supply (16.0–30.0 VDC at the supply unit for 50 m cable), ≤1 Vpp ripple, ≤1 AT fuse. Use an 8×0.14 mm² shielded cable with FCC68 8-pole female connector; for third-party controllers, build the sensor cable to the manual wiring diagram and read the signal between pins 3 and 5.
Plan cleaning when operating in 10⁻⁴…10⁻² mbar; below 10⁻⁶ mbar over a year is possible without cleaning. Order kits by number: FRG700MAINT/FRG700REPR/FRG700IGN for FRG-700, FRG702MAINT/FRG702REPR for FRG-702, plus FRG700MEAS25/40 or FRG702MEAS25/40 measuring systems. Sensor cable, controller and clamping rings are not included.