Agilent Technologies 1290 Infinity II Bio LC System

Agilent Technologies 1290 Infinity II Bio LC System — Liquid chromatography with Pumping speed 3×10⁻⁴ m³/h.

The 1290 Infinity II Bio LC System is a biocompatible UHPLC platform built from stackable modules — Bio pump, Bio/Hybrid Multisampler, Multicolumn Thermostat with Bio heat exchanger, and Bio flow-cell detectors. It operates to 1300 bar at analytical flow rates up to 5 mL/min, with column thermostating from 20 °C below ambient to 110 °C across up to eight columns. It serves high-resolution bio-separations where inert flow-path parts and high-pressure performance are both required.

Pumping speed
3×10⁻⁴ m³/h

Full specifications

Core Performance & Functional Specifications
Pumping speed 5 mL/min
Analytical Flow Rate 5 mL/min
Baseline Drift 1 AU/h
Data Rate 240 Hz
Heat Exchanger Internal Volume 1.0 µL
Internal Volume 1.0 µL
Maximum Individual Bottle Volume 2 L
Maximum pressure 1300 bar
Maximum Sampling Rate 120 Hz
Maximum System Pressure 1300 bar
Sample Capacity 6144 samples
Solvent Cabinet Maximum Volume 8 L
Valve Pressure Rating 1300 bar
Maximum number of columns mfr 8 columns
Number of columns per MCT mfr 8 columns
Double mfr 7116 A/B
Single mfr 7114 A/B
Operating Conditions, Environment & Protection
Detector noise mfr 1.5 μAU
Column thermostat cooling mfr 20 ambient
Column thermostat heating mfr 110 °C

Description

Architecture and positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Keep columns, capillaries and valves rated for 1300 bar

The system and Multisampler inject at up to 1300 bar, so every flow-path component must carry that rating. Note the G4235A bio-inert selector valve head is limited to 210 bar and must not be used in high-pressure methods; use the G5641A 2-position/10-port bio valve for 1300 bar service.

System Manual, Features page 7; Valve Kits page 63
Cooling & Media
Avoid DMF mobile phases with the PVDF leak sensor

Dimethylformamide corrodes the PVDF leak sensor and must not be used as mobile phase. Inspect the sensor regularly for corrosion and keep metal parts off the leak panel and sensor, since a triggered sensor stops the running system immediately.

System Manual, Leak Sensor page 37
Mounting
Fit drain connectors on every module except the bottom one

Install drain connectors on all modules of the stack except the lowest; the bottom module's drain outlet must be routed via waste tubing directly to a suitable waste container. Drain connectors can be retrofitted without powering off the stack and do not affect a running analysis.

System Manual, Drain Connectors Installation pages 32-34
Control & I/O
Set rear DIP switches before network connection

All switches DOWN gives default IP 192.168.254.11; switch 4 UP selects DHCP; switch 5 UP selects the STORED address. MAC-address use is only possible in DHCP configuration. Configure the addressing mode before connecting the module to the controller network.

System Manual, Instrument Configuration page 40
Maintenance
Run the LC Performance Check after stack installation

After complete installation of the module stack, run the InfinityLab LC Performance Check to confirm that each module performs and is connected correctly. This verifies the assembled system before routine use rather than relying on individual module self-tests.

System Manual, InfinityLab LC Performance Check page 58
Cooling & Media
Route waste tubing straight without bends or restrictions

Waste must drain directly into the waste containers with no restriction at bends or joints. The recommended container is 5043-1221 (6 L waste can with Stay Safe cap GL45, 4 ports); with a custom waste solution, tube ends must not immerse in the liquid to prevent back-siphon.

System Manual, Waste Concept and Waste Guidance pages 36-37

Selection & Integration Q&A

When should I choose the 1290 Infinity II Bio LC System rather than a non-bio configuration of the same platform?

Choose the Bio configuration when the flow path must be biocompatible: use bio/biocompatible parts only and never mix them with bio-inert parts. It retains 1300 bar and 5 mL/min UHPLC performance with Bio flow cells and Bio heat exchangers, so bio-separations do not sacrifice pressure capability.

Which detector and heat exchanger should I pair with the system for fast, high-resolution separations?

For fast separations select the DAD G7117B or VWD G7114B, both supporting up to 240 Hz; the DAD FS G7117A is limited to 120 Hz. Pair with the lowest-volume Bio heat exchanger compatible with the flow — 1.0 µL (G7116-60091), 1.6 µL (G7116-60071) or 3.0 µL (G7116-60081) — to minimize dispersion.

What must be supplied or scheduled beyond the base system delivery?

Plan for the recommended waste can 5043-1221 (6 L, Stay Safe cap GL45, 4 ports), the correct Bio capillaries per configuration, and the selected Bio heat exchanger and valve kit, which are listed as additional parts. Schedule the InfinityLab LC Performance Check after installation and before routine use.