Agilent Technologies 4300 Handheld FTIR
Agilent Technologies 4300 Handheld FTIR — Molecular spectroscopy.
The Agilent 4300 Handheld FTIR is a portable, weather-resistant spectrometer with a DTGS detector covering 4500–650 cm⁻¹, designed for field analysis. It weighs 2 kg and operates on battery for up to 4 hours or via a 15 V DC external supply. Interchangeable sampling interfaces—45° and grazing-angle reflectance, diffuse reflectance, diamond ATR, and germanium ATR—enable solids, liquids, and surface films to be measured outside the laboratory.
Full specifications
| Wavelength range | 4500...650 |
| Laser | Low-powered solid state |
| Spectral resolution | 4 dimensionless |
| External reflectance angle of incidence mfr | 45 degrees |
| Grazing angle of incidence mfr | 82 degrees |
| Energy (centerburst) — Optimal mfr | 29000 |
| Energy (centerburst) — Marginal mfr | 29000 |
| Energy (centerburst) — Critical mfr | 31000 |
| Laser signal — Optimal mfr | 15000 |
| Laser signal — Marginal mfr | 15000 |
| Laser signal — Critical mfr | 17000 |
| Performance test resolution mfr | 4 |
| Performance test collection time mfr | one-minute minute |
| Performance test duration mfr | 2 test |
| Recommended reflectance absorbance range mfr | 0.1 AU |
| Energy (centerburst) diagnostic — Optimal mfr | 29000 |
| Energy (centerburst) diagnostic — Marginal mfr | 29000 |
| Energy (centerburst) diagnostic — Critical mfr | 31000 |
| Laser signal diagnostic — Optimal mfr | 15000 |
| Laser signal diagnostic — Marginal mfr | 15000 |
| Laser signal diagnostic — Critical mfr | 17000 |
| Warm-up time before performance validation mfr | 30 minutes |
| External reflectance absorbance range mfr | 0.1 AU |
| Grazing angle of incidence mfr | 82 degrees |
| Interferometer geometry mfr | High-throughput Michelson interferometer with fi |
| Source mfr | Wire-wound element element |
| DTGS Spectral range mfr | 4500...650 |
| MCT Spectral range mfr | 5000...1100 |
| Detector mfr | 1.3 MCT |
| Unpacking mfr | 9 |
| Components mfr | 11 |
| Connecting power mfr | 12 |
| Changing the desiccant mfr | 16 |
| Turning on the instrument mfr | 17 |
| Starting the software mfr | 18 |
| Communication to external PC mfr | 19 |
| Performance verification mfr | 19 |
| G8180-67560 mfr | Coarse Gold Reference Cap Cap |
| G8180-67562 mfr | SS Mirror Reference Cap Cap |
| G8180-67563 mfr | Coarse Silver Diff Reference Cap Cap |
| G8180-67564 mfr | Gold Mirror Reference Cap Cap |
| G8180-67300 mfr | 4300 Sample |
| G8180-67147 mfr | 4300 Battery |
| G8180-67000 mfr | 4300 Charger |
| G8180-67135 mfr | 4300 Supply |
| G8180-67610 mfr | 4300 Reflector |
| G8181-67611 mfr | 4300 Reflector |
| X3202-60106 mfr | O-Ring Removal Tool Tool |
| G8184-67000 mfr | 4300 Kit |
| G8184-67001 mfr | O-Ring Replacement Kit Kit |
| G8184-67002 mfr | O-Ring Replacement Kit Kit |
| G8180-67514 mfr | 4300 Cartridge |
| G8180-68001 mfr | 4300 ATR |
| G8180-68002 mfr | 4300 ATR |
| G8180-68003 mfr | 4300 Ref |
| G8180-68004 mfr | 4300 Ref |
| G8180-68005 mfr | 4300 Ref |
| G8180-20029 mfr | 4300 strap |
| Diffuse reflectance mfr | • Suitable for analysis of samples that reflect |
| Ingress protection mfr | 54 |
| Depth | 19.05 cm |
| Diamond Atr Depth Of Penetration | 2 micron |
| Height | 34.5 cm |
| Weight | 2 kg |
| Width | 10.16 cm |
| Ge ATR analysis depth mfr | 0.5...2 micrometers |
| Germanium ATR analysis depth mfr | 0.5...2 micrometers |
| Weight without batteries mfr | 1.88 kg |
| Weight with batteries mfr | 2.22 kg |
| Battery Life Under Continuous Use | 4 hours |
| Battery Operating Time | 4 hours |
| Power supply | 100...240 Hz |
| External power supply output mfr | 15 DC |
| Battery life — Optimal mfr | 30 min |
| Battery life — Marginal mfr | 5 min |
| Battery life — Critical mfr | 5 min |
| Source current — Optimal mfr | 1.9 A |
| Source current — Marginal mfr | 2.2 A |
| Source current — Critical mfr | 2.5 A |
| AC power input mfr | 110 AC |
| Battery life diagnostic — Optimal mfr | 30 min |
| Battery life diagnostic — Marginal mfr | 5 min |
| Battery life diagnostic — Critical mfr | 5 min |
| Source diagnostic — Optimal mfr | 1.9 A |
| Source diagnostic — Marginal mfr | 2.2 A |
| Source diagnostic — Critical mfr | 2.5 A |
| Battery voltage mfr | 10.8 volt |
| Power supply output mfr | 15 DC |
| Sample interface mfr | 45 Reflectance |
| Diamond ATR sample pH range mfr | 1...14 pH |
| Diamond ATR pH range mfr | 1...14 |
| Operating temperature | 32...122 °F |
| Operating humidity mfr | 95 |
| Standard beamsplitter mfr | Zinc selenide selenide |
| IEC/EN 60529 mfr | 54 |
Description
Construction and positioning
- The 4300 Handheld FTIR is a self-contained spectrometer with an integrated CPU and touch screen, a handle-mounted trigger for spectral collection, and a sealed optical housing. It is not a laboratory benchtop instrument; its weather-resistant enclosure and 2 kg weight are intended for handheld field use.
- The DTGS detector variant covers 4500–650 cm⁻¹ and is the only detector option described in the manual; the MCT variant (5000–1050 cm⁻¹) is noted as no longer available. No sibling models are listed, so this card describes the single handheld configuration.
- Sampling interfaces are user-interchangeable via a locking mechanism and a keyed guiding pin with an RFID chip. The RFID chip stores gain settings per interface, which the firmware reads automatically. Interfaces must only be changed when the instrument is off or the Home screen is displayed.
Behavior under load and in different media
- The instrument is designed for ambient field conditions: operating temperature 0–50 °C (32–122 °F) and humidity up to 95%. The DTGS detector does not require cryogenic cooling, unlike the discontinued MCT option.
- For external reflectance, the 45° interface suits films thicker than 1 µm, while the grazing-angle (82°) interface targets ultra-thin films below 1 µm and nanometer-level contamination. Absorbance should be kept between 0.1 and 2 AU; the real-time data viewer aids in achieving proper sample contact.
- Diamond ATR analyzes only the top 2–3 µm and is suitable for solids, liquids, pastes, and gels with pH 1–14. Germanium ATR measures the top 0.5–2 µm and is intended for highly absorbing samples; the Ge crystal is fragile and must not be scratched or subjected to excessive pressure.
Installation, control, and operation
- Power is supplied by four hot-swappable Li-ion batteries or an external 15 V DC power supply (110/220 V AC input). Batteries are charged only in the separate battery charger, not in the instrument. A UL-approved surge protector is required on the AC line.
- The MicroLab Mobile software runs on the integrated CPU; connection to an external PC is via USB for method development and detailed spectral viewing. The instrument is not intrinsically safe and must not be used in flammable atmospheres.
- Daily or weekly diagnostic checks (energy, battery life, source current, laser signal, detector temperature, CPU temperature) are displayed as green/yellow/red indicators. Performance validation tests should be run quarterly or bi-annually after at least 30 minutes warm-up, using the appropriate Retro-Reflector interface.
Integration & Operating Notes
The 4300 requires the included 15 V DC external power supply (part G8180-67135). Substituting another supply may damage the instrument. Always connect through a UL-approved surge protection strip to guard against power surges or faulty outlets, especially in field locations.
For valid spectra with external reflectance interfaces, the absorbance must fall within 0.1–2 AU. Use the real-time data viewer to adjust sample contact or reference cap selection. Values outside this range indicate poor sample contact or inappropriate reference reflectivity.
The removable desiccant cartridge is located behind the back access panel. Replace with new or recharged desiccant as needed to prevent moisture condensation inside the optical housing, which can degrade spectral performance. The manual does not specify a fixed interval; monitor diagnostic values for signs of moisture.
Selection & Integration Q&A
Choose the DTGS 4300 for portable field use where 4500–650 cm⁻¹ coverage and ambient-temperature operation (0–50 °C) are sufficient. The MCT variant (5000–1050 cm⁻¹) is no longer available. A lab FTIR offers higher stability but lacks the 2 kg handheld form factor and battery operation.
Use the included 15 V DC power supply (G8180-67135) with a UL-approved surge protector. Connect to the PC via the separate USB port. The instrument can also run standalone on its integrated CPU; external PC connection is for method development and detailed spectral viewing.
Operate within 0–50 °C and up to 95% humidity. Battery provides up to 4 hours; carry spare charged batteries for extended sessions. Do not immerse in water; clean with a damp cloth only. The instrument is not intrinsically safe—avoid flammable atmospheres.
The base delivery includes four Li-ion batteries, battery charger, power supply, USB cable, Retro-Reflector Reference Interface, polystyrene reference sample, and MicroLab Software UFD. Spare desiccant cartridges, replacement O-rings, additional sampling interfaces (diamond ATR, Ge ATR, grazing-angle, diffuse reflectance), and reference caps are available as spare parts and must be ordered separately.