Agilent Technologies TPS mini Pumping Station
Agilent Technologies TPS mini Pumping Station — Liquid chromatography with Motor power 0.22 kW, Noise level 56.4 dB.
The TPS mini Pumping Station is an integrated high-vacuum unit combining a turbomolecular pump, solid-state frequency-converter controller and oil-free four-head diaphragm forepump in one 7.9 kg package. It delivers 37 l/s N2 pumping speed down to 5 mbar base pressure, with a 70,000 RPM rotor and 2-minute run-up. It serves as a compact, maintenance-free vacuum source for analytical instruments and laboratory chambers.
Full specifications
| Motor power | 220 VA |
| Noise level | 56.4 dB |
| Base Pressure | 5 mbar |
| Maximum operating pressure | 2 pressure |
| Pumping Speed H2 | 32 l/s |
| Pumping Speed He | 44 l/s |
| Pumping Speed N2 | 48 l/s |
| Rotational speed | 70000 RPM |
| Run-up time | 2 minutes |
| Operating position mfr | any any |
| Lubricant mfr | permanent lubrication lubrication |
| Compliance mfr | 55011 |
| Installation category mfr | II II |
| Power cable mfr | 2 long |
| Indoor use mfr | Indoor Use Only Only |
| N2 mfr | 37 l/s |
| He mfr | 29 l/s |
| H2 mfr | 23 l/s |
| Compliance with: mfr | 55011 |
| DIY Full Range Gauge Kit / 74 FS KF 40 mfr | 9699190 |
| DIY Full Range Gauge Kit / 74 FS CFF 4.5” mfr | 9699193 |
| FRG-700 3 m cable mfr | 9699960 |
| FRG-700 5 m cable mfr | 9699961 |
| J1 IN-OUT mfr | Input-output connector to remote control the TPS |
| STATUS mfr | Blue LED to show the system operating condition |
| P1 VENT mfr | Connector to control the optional vent valve val |
| J5 GAUGE mfr | Connector to control external EyeSys Gauge Gauge |
| Minimum free space around system mfr | 20 cm |
| Minimum power cable wire section mfr | 0.75 |
| Shelf life mfr | 12 months |
| Weight | 7.9 kg |
| Input voltage mfr | 100...240 Vac |
| Input frequency mfr | 50...60 Hz |
| Stand-by power mfr | 10 W |
| Protection fuse mfr | 2.5 A |
| Input voltage: mfr | 100...240 Vac |
| Input frequency: mfr | 50...60 Hz |
| Stand-by power: mfr | 10 W |
| Inlet flange | 4.5 NW |
| Serial communication (kit) mfr | 232 software |
| Mains cable NEMA plug [3m] mfr | 9699958 |
| Mains cable EU plug [3m] mfr | 9699957 |
| Serial cable and T-Plus Software mfr | 9699883 |
| Inlet screen DN40 mfr | 9699309 |
| Inlet screen DN63 mfr | 9699300 |
| J2 SERIAL mfr | 232 connection |
| Counter-flange torque capability mfr | 50 Nm |
| Bakeout Temperature | 120 °C |
| Noise level | 56.4 dB |
| Storage temperature | 20 °C |
| Max altitude mfr | 2000 m |
| Pollution degree mfr | 2 |
| DIY Full Range Gauge Kit / 74 FS ISO 63 mfr | 9699192 |
| ISO 63 mfr | 10 clamps |
Description
Construction and series positioning
- The TPS mini integrates three functional blocks: a turbomolecular pump, a solid-state frequency converter (controller) and a four-head oil-free diaphragm forepump. The motor is fed at 54 Vac three-phase at 1167 Hz; the controller converts single-phase 100–240 Vac, 50–60 Hz mains into this medium-frequency output.
- Inlet flange variants are CF 4.5" O.D., ISO 63 and KF 40 NW. The unit can be installed in any position but cannot be fixed by its base; the inlet flange must connect to a counter-flange able to withstand 50 Nm around its axis.
- With no sibling models listed, this card describes the TPS mini as a single integrated configuration; flange choice is the principal variant axis.
Behavior under load and media
- Pumping speeds are 37 l/s for N2, 29 l/s for He and 23 l/s for H2, with a base pressure of 5 mbar. The rotor runs at 70,000 RPM and reaches speed in 2 minutes; a SOFT START input can extend ramp-up to 20 minutes for gentle acceleration.
- A thermistor near the upper bearing prevents overheating. Residual vibration after balancing is below 0.01 µm. The diaphragm forepump is oil-free; the turbopump is permanently lubricated.
- An optional inlet screen (mod. 9699300 / 9699309) protects blades from dust >0.7 mm but reduces pumping speed by about 10%.
Installation, control and operation
- Electrical supply requires a three-wire cable (Ph+N+Earth) with minimum 0.75 mm² (AWG18) cross-section; a 2.5 A protection fuse is specified. Minimum free space around the system is 20 cm for air circulation.
- The control panel provides J1 IN-OUT (15-pin D for remote START/STOP, INTERLOCK, PWM speed setting, SOFT START, +24 Vdc output, relay outputs, programmable set point and analog signal), J2 SERIAL (RS 232/RS 485), STATUS LED, P1 VENT (24 Vdc for optional vent valve), J5 GAUGE (for Agilent FRG-700 full-range gauge) and an ON-OFF switch.
- Serial format is 8 data bits, no parity, 1 stop bit, baud rate 600–9600 programmable; RS 485 address is 0x80 + device number (0–31). Use shielded cable of maximum 30 m for serial and I/O connections.
- Bakeout limits at the inlet flange are 120 °C for CF and 80 °C for ISO variants. Maximum operating pressure is 2 bar above atmospheric. Operating ambient is +5 °C to +35 °C; storage is −20 °C to +70 °C.
Integration & Operating Notes
The TPS mini can be installed in any position but cannot be fixed by its base. The inlet flange must be connected to a counter-flange able to withstand 50 Nm around its axis, so the chamber or system mounting must be designed for this rotational load in case of rotor seizure.
The power supply cable must be three wires (Ph+N+Earth) with minimum cross-section 0.75 mm² (AWG18). Ensure proper earth connection and adequate conductor size for the 220 VA input; a 2.5 A protection fuse is specified.
Do not exceed the bakeout temperature at the inlet flange: maximum 120 °C for CF flange variants and maximum 80 °C for ISO flange variants. Exceeding these limits during chamber bakeout can damage the pump.
Maximum operating pressure is 2 bar above atmospheric pressure. Do not exceed this pressure during operation or venting; the integrated diaphragm forepump and turbopump are designed for vacuum service only.
The SOFT START input can extend ramp-up time up to 20 minutes. Use it for gentle acceleration, e.g. after long storage or for delicate processes; the default run-up is 2 minutes.
Configure the host for 8 data bits, no parity, 1 stop bit, and baud rate 600/1200/2400/4800/9600 programmable. Use shielded cable of maximum 30 m for serial and I/O port connections; longer or unshielded cables may cause communication errors or EMI issues.
Shelf life of the turbomolecular pump is 12 months from shipping date. Pumps stored longer than 12 months must be returned to the factory before use. If stored below 5 °C, wait until the TPS mini reaches operating temperature before use.
Minimum free space around the system is 20 cm, required for proper air circulation and cooling. Operating ambient temperature must remain within +5 °C to +35 °C.
Selection & Integration Q&A
Choose the TPS mini when you need a compact, maintenance-free vacuum source with integrated controller and oil-free diaphragm forepump in one 7.9 kg unit. It suits analytical instruments and laboratory chambers where space is limited and a base pressure of 5 mbar with 37 l/s N2 is sufficient. Separate components are preferable only if higher pumping speeds or different flange sizes are required.
The counter-flange must withstand 50 Nm around its axis. For ISO 63, use 4 M10 clamps at 22 Nm or 4 M8 claws at 11 Nm. For ConFlat, use Agilent hardware, apply Fel-pro C-100 high-temperature lubricant to threads and between nuts and flange, and tighten sequentially; the 2.75" CF flange has six blind holes.
The TPS mini is maintenance-free; all service must be by authorized personnel. Before service, disconnect mains, vent, wait for rotor stop and surface temperature below 50 °C. The base delivery does not include the inlet screen (mod. 9699300/9699309, reduces speed ~10%), the optional vent valve for P1, the FRG-700 gauge for J5, or serial/I/O cables; these must be ordered separately.