Agilent Technologies TPS mini Pumping Station — Liquid chromatography with Motor power 0.22 kW, Noise level 56.4 dB.

The TPS mini Pumping Station is an integrated high-vacuum unit combining a turbomolecular pump, solid-state frequency-converter controller and oil-free four-head diaphragm forepump in one 7.9 kg package. It delivers 37 l/s N2 pumping speed down to 5 mbar base pressure, with a 70,000 RPM rotor and 2-minute run-up. It serves as a compact, maintenance-free vacuum source for analytical instruments and laboratory chambers.

Motor power
0.22 kW
Noise level
56.4 dB

Full specifications

Core Performance & Functional Specifications
Motor power 220 VA
Noise level 56.4 dB
Base Pressure 5 mbar
Maximum operating pressure 2 pressure
Pumping Speed H2 32 l/s
Pumping Speed He 44 l/s
Pumping Speed N2 48 l/s
Rotational speed 70000 RPM
Run-up time 2 minutes
Operating position mfr any any
Lubricant mfr permanent lubrication lubrication
Compliance mfr 55011
Installation category mfr II II
Power cable mfr 2 long
Indoor use mfr Indoor Use Only Only
N2 mfr 37 l/s
He mfr 29 l/s
H2 mfr 23 l/s
Compliance with: mfr 55011
DIY Full Range Gauge Kit / 74 FS KF 40 mfr 9699190
DIY Full Range Gauge Kit / 74 FS CFF 4.5” mfr 9699193
FRG-700 3 m cable mfr 9699960
FRG-700 5 m cable mfr 9699961
J1 IN-OUT mfr Input-output connector to remote control the TPS
STATUS mfr Blue LED to show the system operating condition
P1 VENT mfr Connector to control the optional vent valve val
J5 GAUGE mfr Connector to control external EyeSys Gauge Gauge
Minimum free space around system mfr 20 cm
Minimum power cable wire section mfr 0.75
Shelf life mfr 12 months
Dimensions, Weight & Mechanical Design
Weight 7.9 kg
Electrical, Power & Drive Specifications
Input voltage mfr 100...240 Vac
Input frequency mfr 50...60 Hz
Stand-by power mfr 10 W
Protection fuse mfr 2.5 A
Input voltage: mfr 100...240 Vac
Input frequency: mfr 50...60 Hz
Stand-by power: mfr 10 W
Interfaces, Connections & Control
Inlet flange 4.5 NW
Serial communication (kit) mfr 232 software
Mains cable NEMA plug [3m] mfr 9699958
Mains cable EU plug [3m] mfr 9699957
Serial cable and T-Plus Software mfr 9699883
Inlet screen DN40 mfr 9699309
Inlet screen DN63 mfr 9699300
J2 SERIAL mfr 232 connection
Counter-flange torque capability mfr 50 Nm
Operating Conditions, Environment & Protection
Bakeout Temperature 120 °C
Noise level 56.4 dB
Storage temperature 20 °C
Max altitude mfr 2000 m
Pollution degree mfr 2
Standards, Safety & General Data
DIY Full Range Gauge Kit / 74 FS ISO 63 mfr 9699192
ISO 63 mfr 10 clamps

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Mounting
Counter-flange must resist 50 Nm rotor-seizure torque

The TPS mini can be installed in any position but cannot be fixed by its base. The inlet flange must be connected to a counter-flange able to withstand 50 Nm around its axis, so the chamber or system mounting must be designed for this rotational load in case of rotor seizure.

User Manual, Installation, p.258
Electrical
Supply cable must be three-wire, minimum 0.75 mm² (AWG18)

The power supply cable must be three wires (Ph+N+Earth) with minimum cross-section 0.75 mm² (AWG18). Ensure proper earth connection and adequate conductor size for the 220 VA input; a 2.5 A protection fuse is specified.

User Manual, Installation, p.259
Process Limits
Bakeout temperature differs by flange type: 120 °C CF, 80 °C ISO

Do not exceed the bakeout temperature at the inlet flange: maximum 120 °C for CF flange variants and maximum 80 °C for ISO flange variants. Exceeding these limits during chamber bakeout can damage the pump.

User Manual, Use, p.261
Process Limits
Never exceed 2 bar above atmospheric pressure

Maximum operating pressure is 2 bar above atmospheric pressure. Do not exceed this pressure during operation or venting; the integrated diaphragm forepump and turbopump are designed for vacuum service only.

User Manual, Installation, p.258
Control & I/O
SOFT START can extend ramp-up to 20 minutes

The SOFT START input can extend ramp-up time up to 20 minutes. Use it for gentle acceleration, e.g. after long storage or for delicate processes; the default run-up is 2 minutes.

User Manual, J1 – IN-OUT, p.289
Control & I/O
Serial link: 8-N-1, baud 600–9600, shielded cable max 30 m

Configure the host for 8 data bits, no parity, 1 stop bit, and baud rate 600/1200/2400/4800/9600 programmable. Use shielded cable of maximum 30 m for serial and I/O port connections; longer or unshielded cables may cause communication errors or EMI issues.

User Manual, RS 232/RS 485, p.298; Interconnections, p.286
Maintenance
Pumps stored over 12 months must be returned to factory

Shelf life of the turbomolecular pump is 12 months from shipping date. Pumps stored longer than 12 months must be returned to the factory before use. If stored below 5 °C, wait until the TPS mini reaches operating temperature before use.

User Manual, Storage, p.256
Cooling & Media
Keep 20 cm free space around the unit for air circulation

Minimum free space around the system is 20 cm, required for proper air circulation and cooling. Operating ambient temperature must remain within +5 °C to +35 °C.

User Manual, Installation, p.260; Technical Specification, p.273

Selection & Integration Q&A

When should I choose the TPS mini Pumping Station rather than a separate turbopump and controller combination?

Choose the TPS mini when you need a compact, maintenance-free vacuum source with integrated controller and oil-free diaphragm forepump in one 7.9 kg unit. It suits analytical instruments and laboratory chambers where space is limited and a base pressure of 5 mbar with 37 l/s N2 is sufficient. Separate components are preferable only if higher pumping speeds or different flange sizes are required.

What are the requirements for the counter-flange and mating hardware at the inlet?

The counter-flange must withstand 50 Nm around its axis. For ISO 63, use 4 M10 clamps at 22 Nm or 4 M8 claws at 11 Nm. For ConFlat, use Agilent hardware, apply Fel-pro C-100 high-temperature lubricant to threads and between nuts and flange, and tighten sequentially; the 2.75" CF flange has six blind holes.

What maintenance is required, and what accessories are not included in the base delivery?

The TPS mini is maintenance-free; all service must be by authorized personnel. Before service, disconnect mains, vent, wait for rotor stop and surface temperature below 50 °C. The base delivery does not include the inlet screen (mod. 9699300/9699309, reduces speed ~10%), the optional vent valve for P1, the FRG-700 gauge for J5, or serial/I/O cables; these must be ordered separately.