Agilent Technologies Turbo V 81
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Turbo V 81

Agilent Technologies Turbo V 81 — Turbomolecular Vacuum Pumps with Motor power 0.05 kW, Noise level 45 dB, Pumping speed 7.2 m³/h.

The Turbo-V 81-T is a turbomolecular pump with a 9-stage aluminium turbine on permanently lubricated ceramic ball bearings, driven by a 54 Vac three-phase motor at up to 1350 Hz. It reaches 5×10⁻⁹ mbar base pressure with a recommended DS 42 forepump and delivers 50 l/s N₂ through its 63 mm inlet. Four flange variants (ISO 63, KF 40, CF 4.5", CF 2.75") cover clean high-vacuum duties.

Motor power
0.05 kW
Noise level
45 dB
Pumping speed
7.2 m³/h

Direct drop-in replacements

Pfeiffer Vacuum
HiPace 300 with TC 110
43%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Motor power 50 W
Noise level 45 dB
Pumping speed 2 l/s
Base Pressure 5 mbar
Base Pressure With Recommended Forepump 5 mbar
Compression ratio 2 dimensionless
Maximum pressure 2 pressure
Pumping Speed H2 50 l/s
Pumping Speed He 65 l/s
Pumping Speed N2 77 l/s
Rotational speed 1350 Hz
Run-up time 60 seconds
Operating position mfr any any
Input mfr 60 max
Lubricant mfr permanent lubrication lubrication
Installation category mfr II II
Compliance with: mfr 292
Heater band 220 V mfr 9699801
Heater band 120 V mfr 9699802
Pump extension cable mfr 9699942
Vibration damper CFF 4.5" mfr 9699376
Mechanical pump DS 102 mfr 9499315
Mechanical pump DS 42 mfr 9499309
Dry scroll SH 100 mfr 01001 UNIV
Dry scroll SH 110 mfr 01101 UNIV
PCB 24 V mfr 9699538
Dimensions, Weight & Mechanical Design
Weight 1.83 kg
Weight kg (lbs): mfr 63
Interfaces, Connections & Control
Inlet flange 63 O.D.
Foreline flange mfr 16 NW
Inlet screen, DN 40 mfr 9699309
Inlet screen, DN 63 mfr 9699300
Vibration damper DN 63 mfr 9699375
Rack controller 81-AG base mfr 9698988
Rack controller 81-AG RS232/485 mfr 9698989
Rack controller 81-AG Profibus mfr 9698990
Turbo-V 81-AG Navigator Controller 24 Vdc mfr 9698995
Turbo-V 81-AG Navigator Controller 100-240 Vac mfr 9698996
Turbo-V 81-AG Navigator Controller (cable 0.7 m) (any version) mfr 9699844
Turbo-V 81-AG Rack Controller (cable 0.7 m) (any version) mfr 9699844
Controller TV70 Navigator (P/N 9698970, 9698971) mfr 9699834
Turbo-V 70 Rack controller (P/N 9699405, 9699505) mfr 9699843
Materials, Media & Process Compatibility
Coolant water flow mfr 10 l/h
Coolant water pressure mfr 2...4 bar
Coolant water mfr 2...4 bar
Operating Conditions, Environment & Protection
Bakeout Temperature 120 °C
Noise level 45 dB
Storage temperature -20 °C
Cooling requirements mfr Natural air convection Forced air or water optio
Pollution degree mfr 2
Metallic water cooling kit mfr 9699823
Plastic water cooling kit mfr 9699824
Air cooling kit (0.5 m cable) mfr 9699290
Air cooling kit extension cable (5 m) mfr 9699940
Standards, Safety & General Data
Pumping speed With ISO 63 or CFF 4.5” mfr 2 l/s
ISO 63 mfr 10 clamps
Fixing torque for ISO 63 flange mfr 22 Nm

Description

Construction and series positioning

Performance behaviour

Installation, control and operation

Integration & Operating Notes

Mounting
Counter-flange must resist 250 Nm; base support is not allowed

The pump is fixed exclusively through its inlet flange, and the chamber flange plus support structure must be designed for 250 Nm around the axis. Supporting the pump by its base is not permitted, so the chamber interface carries the full cantilever load.

User Manual, Installation, p.195
Electrical
Pump will not start with temperature sensor disconnected

The 6-pin controller connector carries the motor phases on pins B, C, D and ground on pin E; pins A and F are reserved for Agilent. The thermistor circuit must be continuous at commissioning, because an open sensor prevents start-up and the same circuit cuts power above 60 °C.

User Manual, Connection C - Electrical, p.225
Process Limits
Bakeout only below 10⁻⁴ mbar and with water cooling active

Heater bands 9699801/9699802 heat the casing to about 80 °C, but bakeout is permitted only when inlet pressure is below 10⁻⁴ mbar and water cooling is running. Inlet flange temperature must stay within 120 °C (ConFlat) or 80 °C (ISO), otherwise flange and pump damage can result.

User Manual, Heater Band Installation, p.213; Use, p.196
Cooling & Media
Water cooling needs 10 l/h at 2–4 bar and low conductance

Coolant must be +15 °C to +35 °C at 2–4 bar inlet pressure with 10 l/h flow and conductance ≤500 µs/cm; closed circuits above that conductance should use up to 20 % Ethyl-Glycole. Natural convection limits ambient to +5…+25 °C; forced air or water extends it to +35 °C.

User Manual, Technical Specification, p.204; Water Cooling Kit Installation, p.218
Maintenance
No field maintenance; return to Agilent for reconditioning

The Turbo-V 81-T requires no user maintenance and any work must be performed by authorized personnel. Before any intervention, disconnect mains, vent the pump, wait for rotor stop and surface temperature below 50 °C. A reconditioned replacement is available through Agilent service.

User Manual, Maintenance, p.197
Mounting
Rigid foreline piping requires bellows to isolate vibration

The KF 16 NW foreline may be connected by hose or vacuum-approved pipe; if a rigid pipe is used, bellows must eliminate vibration from the mechanical forepump. A suitable trap is advisable with oil-sealed forepumps to prevent oil backstreaming into the turbopump.

User Manual, Connection B - Fore-Vacuum Pump, p.225

Selection & Integration Q&A

When should I choose the ISO 63 version of the Turbo-V 81-T rather than the KF 40 or ConFlat variants?

Pick the ISO 63 model (9698905) for 63 mm O.D. clamp-type high-vacuum connections with 4 M10 clamps at 22 Nm; choose KF 40 (9698906) for small elastomer-sealed foreline-style ports, or ConFlat 4.5"/2.75" (9698907/9698908) where metal-sealed UHV joints and 120 °C bakeout are required. Pumping-speed data differ per flange variant.

What controller and forepump does this pump require, and can it be connected directly to mains?

No direct mains connection: the 6-pin connector supplies a 54 Vac three-phase motor (1350 Hz max, 50 W max) and must be driven by an Agilent Navigator or Standard Rack controller. Recommended forepumps are Agilent DS 42–DS 102 mechanical pumps or the SH 100 dry pump; the 5×10⁻⁹ mbar base pressure is specified with that forepump.

What ambient and mounting conditions must the installation site satisfy?

Mount in any position but fix only via the inlet flange to a counter-flange rated for 250 Nm; the base cannot carry the pump. Natural convection allows +5…+25 °C ambient; above 25 °C add the 24 Vdc air cooling kit or water cooling (+5…+35 °C). Protect with a ferromagnetic shield in magnetic fields and keep vent pressure below 2 bar above atmospheric.

What maintenance is required and which accessories or consumables are not included in the base delivery?

The pump needs no field maintenance; service is by Agilent authorized personnel, with a reconditioned replacement available. Not included: inlet screens 9699300/9699309 (≈10 % speed loss), heater bands 9699801/9699802, air cooling kit 9699290, water cooling kits 9699823/9699824, vent valves with their own control units, vibration isolators 9699375/9699376, and the controller and forepump.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Turbo-V 81-T Turbomolecular Pump Instruction Manual
Manual ·234 pp.
View PDF