No photo available yet

Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies TwisTorr 84 FS

Agilent Technologies TwisTorr 84 FS — Turbomolecular Vacuum Pumps.

The TwisTorr 84 FS is a turbomolecular pump with self-positioning aluminium alloy stator disks and parallel spiral pumping channels, available with ISO 63, CFF 4.5", CFF 2.75" or KF 40 inlet flanges. It delivers 67 l/s N2 and 53 l/s H2 at 81,000 rpm, with foreline tolerance of 14 mbar N2/Ar, 12 mbar He and 4 mbar H2. It is intended for high-throughput N2, Ar and lighter gases in leak-free, baked systems requiring base pressure below 5×10⁻¹⁰ mbar.

Direct drop-in replacements

Pfeiffer Vacuum
HiPace 300 with TC 110
47%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Base Pressure With Recommended Forepump 5 mbar
Pumping Speed Ar 66 l/s
Pumping Speed H2 53 l/s
Pumping Speed He 63 l/s
Pumping Speed N2 67 l/s
Rotational speed 81000 rpm
Run-up time 2 minutes
Max gas throughput N2 mfr 100 sccm
Max gas throughput Ar mfr 70 sccm
Compression ratio N2 mfr 1
Compression ratio He mfr 2
Compression ratio H2 mfr 5
Compression ratio Ar mfr 1
Foreline Tolerance N2 mfr 14 mbar
Foreline Tolerance He mfr 12 mbar
Foreline Tolerance H2 mfr 4 mbar
Foreline Tolerance Ar mfr 14 mbar
Operating position mfr Any Any
Lubricant mfr Permanent lubrication lubrication
Installation category mfr II II
Max gas throughput at full rotational speed N2 mfr 100 sccm
Max gas throughput at full rotational speed Ar mfr 70 sccm
N2 mfr 14 mbar
He mfr 12 mbar
H2 mfr 4 mbar
Ar mfr 14 mbar
X3508-64001 mfr 84
X3508-64002 mfr 84 Profibus
X3510-64000 mfr 84 PCB
Description of the TwisTorr 84 FS mfr 291
Pump Description mfr 293
Technical Specification mfr 295
TwisTorr 84 FS Outline mfr 298
Interconnections mfr 314
Earth (Ground) Connection mfr 314
Connection C – Electrical mfr 315
Soft Start mfr 316
Vent Accessories mfr 326
Vibration Isolator Installation mfr 328
Purge Valve Installation mfr 329
Vibration isolator CFF 4.5” mfr 9699376
Vent Valve N.O. 0,5 mm orifice (0,5 m cable) mfr 9699844
Vent Valve extension cable (5 m) mfr 9699941
Vent Screw M5 mfr 3502
Purge Screw mfr 3502
Purge valve 10 SCCM NW16KF - M12 mfr 9699239
Purge valve 10 SCCM 1⁄4 Swagelok - M12 mfr 9699240
Purge valve 20 SCCM NW16KF - M12 mfr 9699241
Purge valve 20 SCCM 1⁄4 Swagelok - M12 mfr 9699242
Purge valve 10 SCCM 1⁄4 Swagelok - 1⁄4 Swagelok mfr 9699232
Purge valve 20 SCCM 1⁄4 Swagelok - 1⁄4 Swagelok mfr 9699236
Metric screws kit mfr 3502
American screws kit mfr 3502
Pump extension cable (3 m) mfr 9699942
Maximum pressure above atmospheric mfr 2 bar
Maximum storage period mfr 12 meses
Dimensions, Weight & Mechanical Design
Weight 2.05 kg
Bottom Mounting mfr 330
CFF4.5 Mounting kit mfr 3502
Electrical, Power & Drive Specifications
Input Voltage mfr 76 Vac
Max input pwr mfr 130 VA
Stand-by pwr mfr 8 W
Max oper. power mfr 70 W
X3509-64000 mfr 84 V
X3509-64001 mfr 84 V
Interfaces, Connections & Control
Inlet flange 4.5 O.D.
Foreline flange mfr 16 NW
Inlet screen installation mfr 317
TwisTorr 84 FS On Board Controller Installation mfr 330
Connection of the High Vacuum Flange mfr 333
Installation of ISO-K Flange mfr 333
Installation of ConFlat Flange (CFF 2 3⁄4) mfr 335
Installation of ConFlat Flange (CFF 4 1⁄2) mfr 335
Mains cable NEMA plug, 3 m long mfr 9699958
Mains cable European plug, 3 m long mfr 9699957
Serial cable and A-Plus software mfr 9699883
Inlet screen CFF4.5” mfr 3502
Inlet screen, KF40 mfr 9699309
Inlet screen CFF 2.75” mfr 9699328
Controller side mounting bracket mfr 3502
Counter-flange torque rating mfr 250 Nm
Materials, Media & Process Compatibility
Coolant Water Minimum Flow 65 l/h
Coolant water pressure mfr 2...4 bar
Operating Conditions, Environment & Protection
Bakeout Temperature 80 max
Operating Altitude 3000 m
Operating temperature 5 °C
Storage temperature 40 °C
Cooling requirements mfr 35 °C
Noise Pressure level (pump only) mfr 40
Pollution degree mfr 2
Air cooling mfr 100 sccm
Water cooling mfr 100 sccm
Air Cooling Kit Installation mfr 321
Water Cooling Kit Connection mfr 323
Metal Water Cooling kit mfr 3502
Plastic Water cooling kit mfr 3502
Air cooling kit (0,5 m cable) mfr 9699290
Air cooling kit extension cable (5 m) mfr 9699940
Standards, Safety & General Data
Inlet screen ISO63 mfr 3502
Vibration isolator ISO 63 mfr 9699375
ISO 63 K clamp tightening torque mfr 22 Nm

Description

Construction and series positioning

Behavior under load and media

Installation, cooling and control

Integration & Operating Notes

Cooling & Media
Cooling must be installed and active before first start

The pump must always be cooled before use; running without the air cooling kit 9699290 or a water cooling kit X3502-68002/X3502-68003 can damage internal components. Water cooling requires at least 65 l/h at 15–25 °C and 2–4 bar. During bakeout, water cooling must remain active.

User Manual, Description of the TwisTorr 84 FS / Use of TwisTorr 84 FS
Process Limits
Foreline pressure must stay below gas-specific tolerance

The backing pump must maintain foreline pressure below 14 mbar for N2 and Ar, 12 mbar for He and 4 mbar for H2; exceeding these values causes loss of compression. Recommended forepumps are Agilent DS 42, DS 102, IDP-7 and IDP-3. Maximum gas throughput at full speed is 100 sccm N2 and 70 sccm Ar.

User Manual, Technical Specification
Mounting
Inlet flange fixing must withstand 250 Nm around axis

The pump can be installed in any position but must be fixed by its inlet flange with a connection able to withstand 250 Nm around its axis. For ISO 63, use 4 × M10 high-strength steel clamps (e.g. IC63250DCMZ) at 22 Nm fixing torque to prevent detachment in case of rotor failure.

User Manual, Pump Fixing
Electrical
Pump will not start if temperature sensor is disconnected

The 6-pin controller connector uses pins B, C, D for the 3-phase motor supply and pin E for pump ground. If the temperature sensor is disconnected, the pump will not start; the sensor automatically cuts off power above 60 °C. Verify sensor wiring if the pump fails to start.

User Manual, Connection C – Electrical
Control & I/O
Soft Start is disabled by default and must be enabled

Soft Start uses frequency steps 450, 1100, 1200, 1300 and 1350 Hz and is disabled by default. Enable it via software or Remote I/O command after a protracted stop to improve grease distribution in the bearings before reaching full rotational speed.

User Manual, Soft Start
Maintenance
No user maintenance; service only by authorized personnel

The TwisTorr 84 FS requires no maintenance. Any work must be carried out by authorized personnel. Before any work, disconnect from mains, vent the pump, wait for the rotor to stop and for surface temperature to fall below 50 °C. A Request for Return sheet is required before returning the pump for repair or advanced exchange.

User Manual, Maintenance

Selection & Integration Q&A

When should I choose the TwisTorr 84 FS with CFF 4.5" flange instead of the ISO 63 variant?

Choose CFF 4.5" when you need bakeout to 120 °C and metal-sealed UHV joints for base pressure below 5×10⁻¹⁰ mbar; it weighs 3.50 kg. Choose ISO 63 for lower bakeout (80 °C max) and lighter weight (2.05 kg) with elastomer seals. KF 40 and CFF 2.75" are also available for other foreline/inlet standards.

Which controller and backing pump are required for the TwisTorr 84 FS?

Use only a compatible Agilent controller providing 54 Vac three-phase at 1350 Hz: X3508-64001 (AG Rack RS232/485), X3508-64002 (AG Rack Profibus), X3509-64000 (on board 110/220 V), X3509-64001 (on board 24 V) or X3510-64000 (PCB). Recommended forepumps are Agilent DS 42, DS 102, IDP-7 or IDP-3; the foreline must stay below 14 mbar N2/Ar, 12 mbar He and 4 mbar H2.

What are the environmental and duty limits for installing the TwisTorr 84 FS?

Install indoors at +5 °C to +35 °C ambient, 0–90 % RH non-condensing, max altitude 3000 m. Vent port pressure must be less than 2 bar above atmospheric. Bakeout is limited to 80 °C (ISO 63) or 120 °C (CFF 4.5") with water cooling active. Storage is -40 °C to +70 °C with 12-month shelf life.

What accessories are not included in the base delivery and what maintenance is required?

Cooling is mandatory but not included: order air cooling kit 9699290 or water cooling kit X3502-68002/X3502-68003. Inlet screens X3502-68000, X3502-68001, 9699309 or 9699328 are optional and reduce pumping speed by 20% Ar, 18% N2, 7% He, 5% H2. The pump requires no maintenance; service only by authorized personnel after venting and cooling below 50 °C.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TwisTorr 84 FS Turbomolecular Pump Instruction Manual
Manual ·349 pp.
View PDF