Agilent Technologies VHS-6 Diffusion Pump
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies VHS-6 Diffusion Pump

Agilent Technologies VHS-6 Diffusion Pump — Vacuum technologies with Motor power 1.45 kW, Pumping speed 4320 m³/h.

The VHS-6 is a 4-stage, self-aligning stainless-steel jet diffusion pump in the VHS small series, rated 1200 l/s air with standard cold cap and 1.2 Torr·l/s maximum throughput. It sits between the VHS-4 and VHS-250, requiring a 17 cfm backing pump and 0.25 gpm cooling water at 15 psi. It serves as the high-vacuum stage in systems needing stable operation below 10⁻³ Torr.

Motor power
1.45 kW
Pumping speed
4320 m³/h

Agilent Technologies Small Diffusion Pumps Series — Model Comparison & Selection

Model Motor power, kWPumping speed, m³/hCooldown time, minutesCooling water pressure, psiDiffusion pump pumping speed, L/sWarm up time, minutes
VHS-6 Diffusion Pump Current 1.45432010155010
VHS-250 Diffusion Pump 1.45432010155010
VHS-4 Diffusion Pump 1.45432010155010

Full specifications

Core Performance & Functional Specifications
Motor power 1450 Watts
Pumping speed 1200 l/s
Cooldown Time 10 minutes
Diffusion Pump Pumping Speed 50 L/s
Warm-up Time 10 minutes
Vacuum gauge measurement range mfr 6 pressure
Dry scroll pump typical size mfr 3...30 h
Turbomolecular pump pumping speed mfr 70...3 L/s
Optimum Operating Range mfr 1 Torr
Maximum Forepressure mfr 0.65 Torr
Backstreaming Rate at Pump Inlet mfr with std cold cap cap
Backing Pump Size mfr 10 cfm
Recommended Jet Assembly mfr 4 steel
Interfaces, Connections & Control
Turbomolecular pump controller protection class mfr 54
Operating Conditions, Environment & Protection
Cooling Water Pressure 15 psi
Cooling Water Requirements mfr 60 °F
Cooling Water Flow mfr 0.25 gpm
Thermal Switches mfr 300 °F

Description

Construction and series positioning

Behavior under load and operating range

Installation, connection, and control

Integration & Operating Notes

Process Limits
Foreline pressure must stay below 0.65 Torr for jet action

The maximum forepressure (no load) is 0.65 Torr (0.87 mbar); if the backing pump cannot maintain this, the diffusion jet stalls and high-vacuum performance collapses. Size the backing line and pump accordingly, and monitor foreline pressure during high gas loads.

User Manual, Table 1-1, page 19
Cooling & Media
Provide 0.25 gpm cooling water at 15 psi and 60–80 °F inlet

Cooling water flow for the VHS-6 is 0.25 gpm at 15 psi, with maximum inlet temperature 60/80 °F (15–26 °C). Insufficient flow or excessive inlet temperature will trip the thermal switch or degrade pumping performance; install a flow switch interlocked with the heater circuit.

User Manual, Table 1-1, page 20
Electrical
Pump will not start if safety circuits keep contactor open

The pump will not start if safety circuits and/or protective devices prevent the contactor from staying closed. Check utilities, flow switches, interlocks, and thermostat operation before troubleshooting the heater or motor; a manual-reset thermal switch at 300 °F (149 °C) must be reset after cooling.

User Manual, Table 4-1, page 51

Selection & Integration Q&A

When should I choose the VHS-6 over the VHS-4 or VHS-250?

Choose the VHS-6 when you need 1200 l/s air pumping speed and can supply a 17 cfm backing pump and 0.25 gpm cooling water. The VHS-4 also delivers 1200 l/s but needs only a 10 cfm backing pump; the VHS-250 is for higher gas loads (3700 l/s air, 4600 l/s helium) and also requires 17 cfm backing.

What backing pump and foreline components are required for the VHS-6?

A backing pump of at least 17 cfm is required to achieve maximum throughput, and the foreline must be sized to keep pressure below 0.65 Torr (no-load maximum forepressure). Include a foreline valve and gauge; check foreline connections for leaks as part of commissioning.

What maintenance and accessories are needed beyond the base pump?

The base delivery does not include pump fluid, cooling-water fittings, or a cold cap (standard cold cap is referenced for ratings). Maintenance includes checking inlet/foreline connections, drain and fill plugs, and fittings for leaks; the thermal switch is manual-reset at 300 °F. Replacement parts and jet assemblies are listed in the manual.

Dimensional Drawings & Technical Documentation

Technical documentation