BINDER ED 400
BINDER ED 400 — Drying ovens with Temperature 5…300 °C.
The BINDER ED 400 is a 400 L natural-convection drying and heating chamber with a +5 to 300 °C working range, built for three-phase 400 V operation. It is the largest standard ED model, distinguished by its 1,000 mm interior width and 90 kg total load capacity. It serves batch drying, heating, and thermal conditioning where uniform, fan-free air circulation is required.
BINDER ED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 10 x/h |
| Heating time to 150°C | 71 min |
| Interior volume | 400 L |
| Load per rack | 35 kg (ED400-400V-I) |
| Nominal power | 3.4 kW |
| Permitted load | 90 kg (ED400-400V-I) |
| Temperature fluctuation | 0.5 K |
| Temperature uniformity | 3.0 K |
| Rear wall clearance | 160 mm (ED400-400V-I) |
| Side wall clearance | 100 mm (ED400-400V-I) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 31 min |
| Depth | 765 mm (ED400-400V-I) |
| Height | 1025 mm (ED400-400V-I) |
| Interior depth | 520 mm (ED400-400V-I) |
| Interior height | 800 mm (ED400-400V-I) |
| Interior width | 1000 mm (ED400-400V-I) |
| Width | 1235 mm (ED400-400V-I) |
| Net weight of the unit (empty) mfr | 125 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
Description
Construction and series positioning
- The ED 400 uses natural convection (no fan) to circulate air through the 400 L interior, with a defined air-change rate of 10 x/h. This fan-free architecture minimizes turbulence and particle entrainment, making it suitable for applications where forced airflow would disturb lightweight samples.
- Within the ED series, the ED 400 is the largest standard model, offering a 1,000 mm interior width and 800 mm height. Sibling models (ED 23, ED 56, ED 115, ED 260) share the same +5 to 300 °C range but differ in chamber volume and load capacity. The ED 400 is chosen when batch size or sample dimensions exceed the capacity of smaller units.
- The interior is designed to accept racks with a maximum load of 35 kg per rack and a total permitted load of 90 kg, supporting multiple shelves of dense samples.
Thermal behavior and performance
- Temperature uniformity is specified at 3.0 K and fluctuation at 0.5 K, ensuring consistent conditions across the chamber for reproducible drying or heating. Heating-up time to 150 °C is 71 minutes, and recovery after a 30-second door opening at 150 °C is 31 minutes, reflecting the thermal inertia of the 400 L volume.
- The natural convection design results in gentle airflow, which may extend drying times for moisture-laden loads compared to forced-convection ovens, but avoids disturbing fine powders or lightweight parts.
Installation and operation
- The unit requires a 400 V, 3-phase/N/PE, 50 Hz supply with a 16 A fuse and a CEE 16/6 H 5-pin plug (variant ED400-400V-I). Nominal power is 3.4 kW. A 208 V 60 Hz variant (ED400UL-208V-I) is available for North American installations.
- Wall clearances must be maintained: 160 mm at the rear and 100 mm on the sides to ensure adequate ventilation and safe operation. The unit weighs 125 kg empty and has net dimensions of 1,235 mm width, 765 mm depth, and 1,025 mm height.
- The RS 422 interface (option model 422) allows integration with BINDER controllers or external monitoring systems for temperature programming and data logging.
Integration & Operating Notes
The ED 400 draws 3.4 kW and must be connected to a 3 N/PE 400 V 50 Hz supply protected by a 16 A fuse. A CEE 16/6 H 5-pin plug is provided. Ensure the supply can handle the starting current and that the earth connection is reliable to avoid damage or safety hazards.
Insufficient clearance restricts airflow around the chamber, causing overheating of external components and reduced temperature uniformity. Install the unit on a level floor with at least 160 mm behind and 100 mm on each side. Do not block the ventilation slots.
Exceeding the permitted load can deform the racks or damage the chamber interior. Distribute samples evenly across shelves and ensure no single rack carries more than 35 kg. The total load includes racks and accessories. Overloading may also impair temperature uniformity.
The RS 422 option (model 422) allows connection to a PC or BINDER controller for programming temperature profiles and logging data. Use shielded cables and ensure proper termination to avoid communication errors. This interface is not included in the base model unless specified.
To maintain temperature accuracy and avoid contamination, clean the interior regularly with mild detergent. Periodically verify temperature uniformity and calibration, especially after moving the unit or changing loads. Refer to the operating manual for detailed maintenance intervals.
Selection & Integration Q&A
Choose the ED 400 when your batch requires more than 260 L capacity or interior dimensions exceeding those of the ED 260. The ED 400 offers 400 L, 1,000 mm width, and 90 kg total load, making it suitable for large or numerous samples. For smaller loads, the ED 260 or ED 115 may be more energy-efficient and space-saving.
The ED 400 requires a 400 V, 3-phase/N/PE, 50 Hz supply with a 16 A fuse and a CEE 16/6 H 5-pin plug. Nominal power is 3.4 kW. For 208 V 60 Hz networks, the ED400UL-208V-I variant with NEMA L21-20 plug is available. Ensure the supply is properly grounded and can handle the load.
Regular cleaning of the interior and periodic calibration are recommended. The RS 422 interface is an option (model 422) for external control. Racks and shelves are available as accessories and must not exceed 35 kg each. No consumables are included in base delivery; order racks and interface cables separately as needed.