ED-S 56 drawing
BINDER ·Made in Germany · In Active Production

BINDER ED-S 56

BINDER ED-S 56 — Drying ovens with Temperature 7…250 °C.

The ED-S 56 is a 62 L natural-convection drying and heating oven (Solid.Line) with a stainless steel interior and R-S microprocessor controller, operating from +7 to 250 °C. As the smallest chamber in the ED-S series, it delivers 1.05 kW heating with 0.8 K fluctuation and 3.4 K uniformity, reaching 150 °C in 60 minutes. It is intended for gentle drying, heating, and thermal treatment of harmless materials where forced airflow is not required.

Temperature
7…250 °C

BINDER ED-S Series — Model Comparison & Selection

Model Temperature, °CDepth, mmHeating up time to 150 c, minHeight, cmInterior depth, mmInterior volume, L
ED-S 56 Current 7–2505776062.535062
ED-S 115 7–2506165573.5390118

Full specifications

Core Performance & Functional Specifications
Temperature 7...250 °C
Heating time to 150°C 60 min
Interior volume 62 L
Load per rack 15 kg
Nominal power 1.05 kW
Permitted load 45 kg (EDS056-230V)
Temperature fluctuation 0.8 K
Temperature uniformity 3.4 K
Rear wall clearance 160 mm (EDS056-230V)
Side wall clearance 100 mm (EDS056-230V)
Article Number mfr 9090
Recovery time after door was opened for 30 s at 150°C mfr 45 min
Dimensions, Weight & Mechanical Design
Depth 577 mm (EDS056-230V)
Height 625 mm (EDS056-230V)
Interior depth 350 mm (EDS056-230V)
Interior height 440 mm (EDS056-230V)
Interior width 400 mm (EDS056-230V)
Width 614 mm (EDS056-230V)
Net weight of the unit (empty) mfr 36 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, control, and operation

Integration & Operating Notes

Electrical
Supply must be 230 V ±10 % with 6.3 A fusing

The 230 V variant requires a single-phase supply with protective earth and a 6.3 A chamber fuse. Voltage tolerance is ±10 % at 50/60 Hz; out-of-range supply can cause controller faults or heater damage. Verify mains quality before installation and use the type plate to confirm the delivered voltage variant.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 4.3, p. 19-20; type plate, p. 8
Mounting
Keep 160 mm rear and 100 mm side clearance

Natural convection relies on heat dissipation from the housing. Insufficient clearance or installation in unventilated recesses can cause overheating and damage. Ensure the specified distances to walls and adjacent equipment are maintained, and provide adequate room ventilation.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 3.4, p. 17
Process Limits
Do not process materials above their flash point

The chamber has no explosion protection. Charging materials must be harmless and the operating temperature must lie below the flash point or sublimation point. Corrosive substances (acids, halides) can attack the stainless steel interior and are excluded from liability.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 1.9, p. 10-11
Control & I/O
Safety controller class 2 must be set for ED-S

The overtemperature safety device is class 1 per DIN 12880, with a class 2 safety controller for ED-S ovens. The safety controller must be configured to the application's maximum permissible temperature; an alarm triggers when the limit is exceeded. Do not rely on the standard controller alone for unattended operation.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 9, p. 24
Maintenance
Annual electrical inspection and maintenance required

Regular maintenance at least once a year, plus repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after repair. Disconnect the power plug before maintenance; do not remove the rear panel. Replace the door gasket only when cold.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 12.1-12.2, p. 29-30
Mounting
Lift at four corners with two people; never by the door

The chamber weighs 36 kg net. Lifting or transporting by the door handle or door can damage the hinges and gasket. Use two people and lift at the four lower corners. After cold storage, wait at least one hour at ambient temperature before start-up to avoid condensation.

Instruction Manual BD-ED-FD-S (E1) 06-2026, chap. 3.1-3.3, p. 16-17

Selection & Integration Q&A

When should I choose the ED-S 56 over the larger ED-S 115?

Choose the ED-S 56 for smaller batches up to 62 L and 45 kg total load, where bench space is limited. The ED-S 115 offers a larger chamber but the same +7 to 250 °C range and natural convection; select it when sample volume exceeds 62 L or when more rack space is needed.

What are the electrical supply requirements for the ED-S 56 230 V variant?

The 230 V variant requires a single-phase 230 V supply with protective earth, 50/60 Hz, within ±10 % tolerance, and a 6.3 A fuse. The unit is fitted with a CEE 7/7 plug. Verify the type plate to confirm the voltage variant before connecting.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance and electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. The base delivery does not include racks or shelves; these must be ordered separately. The door gasket is a wear part and should be replaced only when cold.

Dimensional Drawings & Technical Documentation

Dimensional drawings

ED-S 56 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model ED S 56 en
Datasheet ·399,7 КБ ·5 pp.
View PDF
PDF
BD ED FD S (E1) 06 2026 en
Manual ·10,8 МБ ·65 pp.
View PDF