BINDER ED-S 56
BINDER ED-S 56 — Drying ovens with Temperature 7…250 °C.
The ED-S 56 is a 62 L natural-convection drying and heating oven (Solid.Line) with a stainless steel interior and R-S microprocessor controller, operating from +7 to 250 °C. As the smallest chamber in the ED-S series, it delivers 1.05 kW heating with 0.8 K fluctuation and 3.4 K uniformity, reaching 150 °C in 60 minutes. It is intended for gentle drying, heating, and thermal treatment of harmless materials where forced airflow is not required.
BINDER ED-S Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, mm | Heating up time to 150 c, min | Height, cm | Interior depth, mm | Interior volume, L |
|---|---|---|---|---|---|---|
| ED-S 56 Current | 7–250 | 577 | 60 | 62.5 | 350 | 62 |
| ED-S 115 | 7–250 | 616 | 55 | 73.5 | 390 | 118 |
Full specifications
| Temperature | 7...250 °C |
| Heating time to 150°C | 60 min |
| Interior volume | 62 L |
| Load per rack | 15 kg |
| Nominal power | 1.05 kW |
| Permitted load | 45 kg (EDS056-230V) |
| Temperature fluctuation | 0.8 K |
| Temperature uniformity | 3.4 K |
| Rear wall clearance | 160 mm (EDS056-230V) |
| Side wall clearance | 100 mm (EDS056-230V) |
| Article Number mfr | 9090 |
| Recovery time after door was opened for 30 s at 150°C mfr | 45 min |
| Depth | 577 mm (EDS056-230V) |
| Height | 625 mm (EDS056-230V) |
| Interior depth | 350 mm (EDS056-230V) |
| Interior height | 440 mm (EDS056-230V) |
| Interior width | 400 mm (EDS056-230V) |
| Width | 614 mm (EDS056-230V) |
| Net weight of the unit (empty) mfr | 36 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The ED-S 56 uses natural convection (no fan) for gentle, draft-free heating; the interior is stainless steel 1.4016 (AISI 430) with a polyester-coated RAL 9003 housing.
- Within the ED-S series, the 56 is the compact 62 L model; the sibling ED-S 115 offers a larger chamber but identical temperature range (+7 to 250 °C). Choose the 56 when bench space and sample volume are limited.
- The R-S microprocessor controller provides a digital display with 1 °C resolution (typical for ED-S/FD-S ovens) and PID control; overtemperature protection is class 1 per DIN 12880 with a class 2 safety controller.
Behavior under load and in operation
- Heating-up time to 150 °C is 60 minutes; recovery after a 30 s door opening at 150 °C is 45 minutes, reflecting the natural-convection design and the thermal mass of the chamber.
- Temperature fluctuation is 0.8 K and uniformity 3.4 K, suitable for drying and heating tasks that do not demand forced-air homogeneity.
- The chamber supports a total load of 45 kg, with a maximum of 15 kg per rack; loads should be distributed evenly to avoid local overheating.
- Above 150 °C, metallic surfaces may discolour by oxidation; this is harmless and does not affect function.
Installation, control, and operation
- The unit requires a 230 V, 1-phase, 50/60 Hz supply with protective earth and a 6.3 A fuse; the CEE 7/7 plug is standard for the 230 V variant (EDS056-230V).
- Wall clearances of 160 mm at the rear and 100 mm at the sides must be observed; do not install in unventilated recesses.
- The exhaust air flap allows adjustment of the air change rate; the R-S controller supports set-point entry, °C/°F selection, and a delayed-off timer.
- The chamber is not explosion-protected; only harmless materials may be processed, and the operating temperature must remain below the flash point or sublimation point of the charge.
Integration & Operating Notes
The 230 V variant requires a single-phase supply with protective earth and a 6.3 A chamber fuse. Voltage tolerance is ±10 % at 50/60 Hz; out-of-range supply can cause controller faults or heater damage. Verify mains quality before installation and use the type plate to confirm the delivered voltage variant.
Natural convection relies on heat dissipation from the housing. Insufficient clearance or installation in unventilated recesses can cause overheating and damage. Ensure the specified distances to walls and adjacent equipment are maintained, and provide adequate room ventilation.
The chamber has no explosion protection. Charging materials must be harmless and the operating temperature must lie below the flash point or sublimation point. Corrosive substances (acids, halides) can attack the stainless steel interior and are excluded from liability.
The overtemperature safety device is class 1 per DIN 12880, with a class 2 safety controller for ED-S ovens. The safety controller must be configured to the application's maximum permissible temperature; an alarm triggers when the limit is exceeded. Do not rely on the standard controller alone for unattended operation.
Regular maintenance at least once a year, plus repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial start-up and after repair. Disconnect the power plug before maintenance; do not remove the rear panel. Replace the door gasket only when cold.
The chamber weighs 36 kg net. Lifting or transporting by the door handle or door can damage the hinges and gasket. Use two people and lift at the four lower corners. After cold storage, wait at least one hour at ambient temperature before start-up to avoid condensation.
Selection & Integration Q&A
Choose the ED-S 56 for smaller batches up to 62 L and 45 kg total load, where bench space is limited. The ED-S 115 offers a larger chamber but the same +7 to 250 °C range and natural convection; select it when sample volume exceeds 62 L or when more rack space is needed.
The 230 V variant requires a single-phase 230 V supply with protective earth, 50/60 Hz, within ±10 % tolerance, and a 6.3 A fuse. The unit is fitted with a CEE 7/7 plug. Verify the type plate to confirm the voltage variant before connecting.
Annual maintenance and electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. The base delivery does not include racks or shelves; these must be ordered separately. The door gasket is a wear part and should be replaced only when cold.