BINDER FD-S 56
BINDER FD-S 56 — Drying ovens with Temperature 10…250 °C.
The FD-S 56 is a forced-convection drying and heating oven with a 55 L stainless steel chamber, rated +10 to 250 °C and 1.1 kW on 230 V single-phase. Its 4 h⁻¹ air-change rate and 15 min heat-up to 150 °C distinguish it from natural-convection ED-S siblings. It serves solvent-free drying, thermal treatment and aging where uniform forced airflow matters.
BINDER FD-S Series — Model Comparison & Selection
| Model | Temperature, °C | Air change rate, 1/hour | Depth, mm | Heating up time to 150 c, min | Height, cm | Interior depth, mm |
|---|---|---|---|---|---|---|
| FD-S 56 Current | 10–250 | 4 | 577 | 15 | 62.5 | 310 |
| FD-S 115 | 10–250 | 2 | 616 | 20 | 73.5 | 350 |
Full specifications
| Temperature | 10...250 °C |
| Air change rate | 4 x/h |
| Heating time to 150°C | 15 min |
| Interior volume | 55 L |
| Load per rack | 15 kg (FDS056-230V) |
| Nominal power | 1.1 kW |
| Permitted load | 45 kg (FDS056-230V) |
| Temperature fluctuation | 0.4 K |
| Temperature uniformity | 2.6 K |
| Rear wall clearance | 160 mm (FDS056-230V) |
| Side wall clearance | 100 mm (FDS056-230V) |
| Article Number mfr | 9090 |
| Recovery time after door was opened for 30 s at 150°C mfr | 5 min |
| Depth | 577 mm (FDS056-230V) |
| Height | 625 mm (FDS056-230V) |
| Interior depth | 310 mm (FDS056-230V) |
| Interior height | 440 mm (FDS056-230V) |
| Interior width | 400 mm (FDS056-230V) |
| Width | 614 mm (FDS056-230V) |
| Net weight of the unit (empty) mfr | 35 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Forced-air circulation (FD-S) vs natural convection (ED-S/BD-S); stainless steel 1.4016 (AISI 430) interior, polyester-coated RAL 9003 housing.
- 55 L interior (400 × 440 × 310 mm W×H×D); 15 kg per rack, 45 kg total permitted load.
- R-S microprocessor PID controller with digital display; 1 °C resolution on FD-S (0.1 °C only on BD-S incubators).
- Overtemperature protection class 1 per DIN 12880:2007 standard; safety controller class 2 for FD-S.
Thermal behavior
- Working range +10 to 250 °C; temperature fluctuation 0.4 K, uniformity 2.6 K.
- Heat-up to 150 °C in 15 min; recovery after 30 s door opening at 150 °C is 5 min.
- Air-change rate 4 h⁻¹, adjustable via exhaust air flap.
- Above 150 °C, metallic surfaces may discolour by oxidation; harmless.
Installation and operation
- Wall clearance: 160 mm rear, 100 mm side; do not install in unventilated recesses.
- 230 V 1~ 50 Hz, 1.1 kW, 6.3 A chamber fuse, CEE 7/7 plug; voltage tolerance ±10 %.
- Lift at four lower corners with two people; never use door handle.
- Delayed-off timer function; °C/°F selectable.
Integration & Operating Notes
The chamber is rated 230 V 1~ 50 Hz with a 6.3 A fuse. Verify mains quality before installation; out-of-tolerance voltage can affect controller operation and heating performance.
Insufficient clearance or installation in unventilated recesses prevents heat dispersal and can cause overheating damage. Plan the bench layout around these minimum distances.
The chamber has no explosion protection. Operating temperature must remain below the flash point or sublimation point of the charging material; verify material properties before loading.
Interior is 1.4016 (AISI 430). Acids and halides can corrode it; corrosion damage is excluded from liability. Use only harmless charging materials.
Plan at least yearly maintenance and repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Disconnect power before service; do not remove rear panel; replace door gasket only when cold.
FD-S ovens show temperature in 1 °C steps; sub-degree readout is available only on BD-S incubators. Set and read set-points accordingly.
Selection & Integration Q&A
Pick FD-S 56 for forced convection and 55 L capacity with 4 h⁻¹ air change. Choose ED-S 56 if natural convection suffices; choose FD-S 115 for larger loads. FD-S 56 suits uniform drying where airflow is needed.
230 V 1~ 50 Hz, 1.1 kW, 6.3 A chamber fuse, CEE 7/7 plug, ±10 % voltage tolerance. Provide a grounded single-phase circuit with 6.3 A protection per chamber.
Annual maintenance and electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702. Door gasket is a wear part (replace cold). Racks and exhaust connection are optional accessories; base delivery excludes them.