BINDER FP 56
BINDER ·Made in Germany · In Active Production

BINDER FP 56

BINDER FP 56 — Drying ovens with Temperature 10…300 °C.

The BINDER FP 56 is a 60 L forced-convection drying and heating chamber with a +10 to 300 °C working range, 1.2 kW single-phase heating, and an 80 h⁻¹ air-change rate. As the smallest standard FP variant, it delivers 0.4 K fluctuation and 1.4 K uniformity at 150 °C with a 4-minute recovery after a 30-second door opening. It is intended for reproducible drying, heating, and thermal conditioning of non-explosive, non-corrosive loads in laboratory and industrial process chains.

Temperature
10…300 °C

BINDER FP Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FP 56 Current 10–300805961562.260
FP 115 10–300326361973.2116
FP 260 10–30097861996.2259
FP 720 12–300581625159741

Full specifications

Core Performance & Functional Specifications
Temperature 10...300 °C
Air change rate 80 x/h
Heating time to 150°C 15 min
Interior volume 60 L
Load per rack 30 kg (FP056-230V / FP056UL-120V)
Nominal power 1.2 kW
Permitted load 70 kg (FP056-230V / FP056UL-120V)
Temperature fluctuation 0.4 K
Temperature uniformity 1.4 K
Rear wall clearance 160 mm (FP056-230V / FP056UL-120V)
Side wall clearance 100 mm (FP056-230V / FP056UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 596 mm (FP056-230V / FP056UL-120V)
Height 622 mm (FP056-230V / FP056UL-120V)
Width 560 mm (FP056-230V / FP056UL-120V)
Net weight of the unit (empty) mfr 41 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 10.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior under load and door cycling

Installation, control, and operation

Integration & Operating Notes

Mounting
Maintain 160 mm rear and 100 mm side clearance

The chamber dissipates heat through the rear and sides; installing it in a recess or closer than the prescribed distances prevents adequate ventilation and can cause overheating damage. Verify clearance before positioning, especially when the unit is placed under a shelf or beside another heat source.

Operating Manual FP (E3.1), chap. 3.4 Installation, p. 22
Electrical
Supply must stay within ±10% of nominal at 50 Hz

The 230 V variant is rated 1.2 kW at 230 V 1~ 50 Hz with a 10.0 A unit fuse. Out-of-tolerance voltage or the wrong frequency affects temperature performance and safety; size the branch circuit for the full unit fuse rating and confirm the supply frequency before connection.

Operating Manual FP (E3.1), chap. 1.6 Type plate, p. 9
Process Limits
Load temperature must remain below flash or sublimation point

The chamber has no explosion protection. Operating temperature must lie below the flash point or sublimation point of the charging material, and no explosive mixture with air may form. Verify solvent and material properties before loading; corrosive acids and halides that attack stainless steel, aluminum, or copper are also excluded.

Operating Manual FP (E3.1), chap. 1.8 and 1.9, pp. 11-12
Control & I/O
Select overtemperature class 2 or 3.1 before commissioning

The temperature safety device per DIN 12880 can be configured as a class 2 temperature limiter or a class 3.1 safety controller. The selection changes alarm behavior and reset requirements, so it must be agreed with the process owner and set before the chamber is released to production.

Operating Manual FP (E3.1), chap. 11.2.1, p. 38
Control & I/O
Alarm contact is zero-voltage; analog output is optional

A zero-voltage switching contact is available on the rear SUB-D 'ALARM / SWITCH CONTACT' socket for external monitoring. An analog temperature output requires the optional SUB-D 'ANALOG OUTPUT' socket. Only connect external devices compliant with EN 61010-1:2010 or EN 60950-1:2006 mod.

Operating Manual FP (E3.1), chap. 2.2 and 2.5, pp. 19-20
Maintenance
Annual maintenance and electrical repeat inspection are mandatory

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; warranty is void if non-authorized personnel perform it. An electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required at commissioning, after service, and annually. The controller shows 'Service due!' after one year.

Operating Manual FP (E3.1), chap. 21.1-21.3, pp. 65-66
Maintenance
Replace the door gasket only when the chamber is cold

Hot gasket replacement can damage the gasket and the door seal seat. Allow the chamber to cool and disconnect power before service. Never remove the rear panel or allow water to penetrate the chamber during cleaning or maintenance.

Operating Manual FP (E3.1), chap. 21.2, p. 65

Selection & Integration Q&A

When should I choose the FP 56 instead of the FP 115 or FP 260?

Pick the FP 56 when 60 L and 70 kg total load cover the batch and the 560 mm width fits the bench or floor slot. Move to FP 115 or FP 260 only when larger batches or more rack area are required; all three share the +10 to 300 °C range and forced convection, so the choice is volumetric, not thermal.

What supply and external devices does the FP 56 need for integration?

The 230 V variant needs 230 V 1~ 50 Hz within ±10%, sized for the 10.0 A unit fuse. External alarm wiring uses the zero-voltage SUB-D 'ALARM / SWITCH CONTACT'; analog temperature output requires the optional 'ANALOG OUTPUT' socket. Ethernet and USB devices must comply with EN 61010-1:2010 or EN 60950-1:2006 mod.

What mounting and duty-cycle limits apply to the FP 56?

Install indoors only, with 160 mm rear and 100 mm side clearance, and never in an unventilated recess. The chamber is IP 20 and must not be sprayed or operated wet. Loads must stay below the flash or sublimation point, and total load must not exceed 70 kg with no more than 30 kg per rack.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by authorized personnel and annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required; the controller shows 'Service due!' after one year. Optional accessories include APT-COM 4 software, flexible Pt100 object sensor, analog output, HEPA fresh air filter, and inert gas connection; these are not part of the standard scope.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model FP 56 en
Datasheet ·521,8 КБ ·7 pp.
View PDF
PDF
FP (E3.1) 06 2026 en
Manual ·7,9 МБ ·93 pp.
View PDF