BINDER KBF 130

BINDER KBF 130 — Climate chambers with Temperature 0…70 °C.

The BINDER KBF 130 is a 127 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, using a steam humidifier and R600A refrigeration circuit. As the smallest standard KBF variant (Art. 9020-0478, 230 V), it is distinguished from larger siblings by its compact 130-size cabinet and tilt-protection kit, while sharing the MB2 program controller and DIN 12880 class 2/3.3 safety architecture. It is intended for reproducible conditioning and stability testing of samples at defined temperature and humidity in laboratory and quality-control environments.

Temperature
0…70 °C

BINDER KBF Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF 130 Current 0–70127302701.60.2
KBF 1060 0–701070406401.60.3
KBF 1600 0–701610406401.70.4
KBF 260 0–70262404101.40.3
KBF 470 0–70472405301.40.2
KBF 720 0–70727406401.60.3

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 127 L (KBF130-230V / KBF130UL-120V)
Load per rack 30 kg (KBF130-230V / KBF130UL-120V)
Maximum heat compensation 270 W
Nominal power 1.6 kW (KBF130-230V / KBF130UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 3 min
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (KBF130-230V / KBF130UL-120V)
Rated voltage 220...240 V
Unit fuse 16.0 A (KBF130-230V / KBF130UL-120V)
Operating Conditions, Environment & Protection
Humidity fluctuation 1.5 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must match 220...240 V single-phase with 16 A fuse

The KBF 130-230V draws 1.6 kW nominal on a 1~ 220...240 V, 50/60 Hz supply and requires a 16.0 A unit fuse. Connecting to a 120 V circuit or a supply outside this band will prevent correct operation; the UL variant (9020-0494) is the separate 120 V model with NEMA 5-20 plug.

Data Sheet Model KBF 130 en.pdf; Factory Specification Sheet
Process Limits
Humidity control only within 10...70 °C, not full 0...70 °C range

Although the chamber temperature range is 0...70 °C, humidity control is specified only for 10...70 °C. Operating the humidifier below 10 °C is outside the validated humidity range and may produce condensation or control instability; plan humidity tests within the 10...70 °C window.

Factory Specification Sheet (Temperature range with humidity
Process Limits
Do not exceed 30 kg per rack or 270 W heat compensation

Each rack is limited to 30 kg load, and the chamber can compensate a maximum of 270 W of heat input from samples. Exceeding either limit compromises temperature uniformity (0.2 K) and may prevent the chamber from holding set-point, especially at low temperatures where refrigeration capacity is already engaged.

Factory Specification Sheet; Operating Manual, Technical Description
Maintenance
Annual maintenance and twice-yearly condenser cleaning required

Regular maintenance must be performed at least once a year, including the steam humidifier; condensers should be cleaned at least twice a year (more often in dusty ambient air). Refrigeration work requires personnel certified per EN 13313:2010. After 8760 operating hours or two years a maintenance message appears every two weeks until reset by BINDER Service.

Operating Manual, Section 23.2, p.133-134
Process Media
Water connections must be rated for at least 11 bar

The chamber's maximum operating pressure is 11 bar. Freshwater supply and wastewater connections must be rated accordingly and the pressure regulated if the site water main exceeds this value. The humidifier's maintenance interval also depends on the available water quality and steam output, so water hardness should be checked at commissioning.

Operating Manual, Section 1.6 Type plate, p.12; Section 23.2, p.134

Selection & Integration Q&A

When should I choose the KBF 130 rather than a larger KBF sibling such as the KBF 260 or KBF 470?

Pick the KBF 130 for 127 L interior volume and loads within 30 kg per rack and 270 W heat compensation. Larger siblings (260, 470, 1060, 1600) share the 0...70 °C range but add volume and, from size 260, require a wall spacer; the 130 uses the compact triangular panel and tilt-protection kit.

What are the electrical and water-supply requirements for integrating the KBF 130 into my lab?

The 230 V variant needs a 1~ 220...240 V, 50/60 Hz supply with 16 A fuse and CEE 7/7 plug (1.6 kW nominal). Water connections must withstand 11 bar; the humidifier needs freshwater and wastewater routing, with the connection scheme depending on whether the serial number is up to or above 20250000010225.

What mounting and environmental limits apply to the KBF 130 in a laboratory?

The 130-size unit requires the flexible tilt-protection kit and is rated IP 20, so it must not be sprayed with water. Unlike sizes 260 and larger, no wall spacer is specified for the 130. Ambient dust increases condenser cleaning frequency beyond the standard twice-yearly interval.

What maintenance schedule and accessories should I budget for beyond the base chamber delivery?

Plan annual maintenance including the steam humidifier, twice-yearly condenser cleaning, and electrical safety testing per EN 50678/EN 50699. The maintenance message appears at 8760 h or two years and requires BINDER Service reset. Optional accessories include APT-COM 4, analog outputs, zero-voltage alarm contacts, Pt 100 object sensor, and external freshwater/wastewater cans.

Dimensional Drawings & Technical Documentation