BINDER KBF 260
BINDER KBF 260 — Climate chambers with Temperature 0…70 °C.
The BINDER KBF 260 is a 262 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 260-size cabinet with a triangular instrument panel. It is the mid-series 230 V single-phase model (220...240 V, 1.4 kW, 16 A fuse) distinguished from smaller 130 and larger 470/1060/1600 siblings by its 40 kg-per-rack load capacity and 410 W max heat compensation. It is used for stable, reproducible temperature and humidity conditioning of samples in laboratory and industrial test processes.
BINDER KBF Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 262 L (KBF260-230V / KBF260UL-120V) |
| Load per rack | 40 kg (KBF260-230V / KBF260UL-120V) |
| Maximum heat compensation | 410 W |
| Nominal power | 1.4 kW (KBF260-230V / KBF260UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 3 min |
| Power frequency | 50/60 Hz (KBF260-230V / KBF260UL-120V) |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A (KBF260-230V / KBF260UL-120V) |
| Humidity fluctuation | 1.5 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and series positioning
- The KBF 260 (Article 9020-0479) is the 260-size member of the KBF constant climate chamber series, which also includes the 130, 470, 1060 and 1600 sizes; all share the 0...70 °C temperature range.
- This model uses the triangular instrument panel layout specific to sizes 130 and 260, rather than the signal triangle and ergonomically adjustable control terminal of the 470 and larger sizes.
- The 262 L interior volume and 40 kg load per rack place it between the compact 130 and the larger 470; the 410 W maximum heat compensation defines the maximum sample heat load that can be balanced.
- Two mains variants exist: KBF260-230V (220...240 V, 1~, 50/60 Hz, CEE 7/7 plug) and KBF260UL-120V (120 V, 1~, 50/60 Hz, NEMA 5-20 plug).
Behavior under load and in humidity mode
- Temperature fluctuation is 0.1 K and temperature uniformity 0.3 K; humidity fluctuation is 1.5 % RH.
- With humidity active, the temperature range is restricted to 10...70 °C; the full 0...70 °C range applies only without humidity control.
- Recovery after a 30 s door opening is 2 min at 25 °C / 60 % RH and 3 min at 40 °C / 75 % RH, so short access does not require long re-stabilization.
- The chamber uses a steam humidifier and a refrigeration system with R600A refrigerant (charge -0.095 kg); defrosting occurs during refrigerating operation and anti-condensation protection is provided.
Installation, connection and control
- From size 260 upward a wall spacer is required to maintain the specified wall distance; the flexible tilt protection kit applies only to sizes 130 and 470, not to this model.
- The rear connection panel carries freshwater and wastewater connections; the maximum operating pressure is 11 bar, so supply fittings must be rated accordingly.
- Control is via the MB2 chamber controller with fixed-value, time-program and week-program modes, chart-recorder display, alarm and notification functions, and user management with password-protected authorization levels.
- Temperature safety is provided by an over-temperature protective device (class 1) and an over-temperature safety controller class 2/3.3 (selectable as temperature limiter or temperature protection), plus undertemperature protection class 3.2, per DIN 12880:2007.
- The chamber is rated IP 20 according to EN 60529 and must not be sprayed with water.
Integration & Operating Notes
The 230 V variant requires 220...240 V single-phase 50/60 Hz with a 16 A unit fuse and CEE 7/7 plug; the UL variant requires 120 V with NEMA 5-20. Connecting a chamber to a supply outside its type-plate rating risks damage and invalidates safe operation, so verify the type plate before installation.
From chamber size 260 onward a spacer for wall distance is required; without it, heat rejection from the rear of the chamber is impaired. The flexible tilt protection kit is not used on this size — it applies only to sizes 130 and 470.
The maximum heat compensation of 410 W defines how much heat the chamber can remove from operating samples. Exceeding it prevents the chamber from holding the set temperature, so calculate the total dissipated power of the load before designing a test.
The chamber's maximum operating pressure is 11 bar. Freshwater and wastewater fittings, tubing and valves connected to the rear panel must be rated for at least this pressure to avoid leakage or rupture when connected to a pressurized water main.
Regular maintenance is required at least once a year, and the steam humidifier must also be serviced at least annually. Condensers should be cleaned at least twice a year by a qualified technician, and more often in dusty ambient air. Maintenance by non-authorized personnel voids the warranty.
The refrigeration system contains flammable R600A hydrocarbon refrigerant. Any maintenance on it must be performed by qualified personnel trained in accordance with EN 13313:2010 (e.g. refrigeration technician certified per Regulation (EC) No. 303/2008).
Replacing the door gasket while the chamber is warm may damage the new gasket. Allow the chamber to cool to ambient before attempting replacement, and include electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 in annual service.
Selection & Integration Q&A
Choose the 260 for 262 L interior volume and 40 kg per rack when the 130 is too small and the 470 is unnecessary. All share 0...70 °C; the 260 uses the triangular instrument panel of sizes 130/260, while 470 and larger use the signal triangle and adjustable control terminal.
From size 260 a wall spacer is required to maintain the specified wall distance; the tilt protection kit does not apply to this size. The chamber is IP 20 per EN 60529 and must not be sprayed with water. Ambient dust increases condenser cleaning frequency beyond twice a year.
Annual maintenance and annual steam humidifier service are required, plus condenser cleaning at least twice a year; BINDER Service must reset the maintenance message after 8760 h or two years. Options not included: APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, Pt 100 object sensor, external freshwater/wastewater cans, BINDER Pure Aqua Service.