BINDER KBF PRO 130
BINDER KBF PRO 130 — Climate chambers with Temperature -20…100 °C.
The KBF PRO 130 is a 127 L constant climate chamber with a -20…100 °C temperature range and 10…98 % RH humidity range, distinguished within the KBF PRO series by its compact 130-size cabinet and triangular instrument panel. It is intended for precise, reproducible climatic testing of small to medium samples, with 0.1 K temperature fluctuation and 0.2 K uniformity.
BINDER KBF PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW | Temperature uniformity, K |
|---|---|---|---|---|---|---|
| KBF PRO 130 Current | -20–100 | 127 | 30 | 450 | 1.9 | 0.2 |
| KBF PRO 1060 | -20–100 | 1070 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 1600 | -20–100 | 1610 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 260 | -20–100 | 262 | 40 | 800 | 1.6 | 0.2 |
| KBF PRO 470 | -20–100 | 472 | 40 | 800 | 1.6 | 0.3 |
| KBF PRO 720 | -20–100 | 727 | 40 | 1000 | 1.8 | 0.3 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 127 L |
| Load per rack | 30 kg |
| Maximum heat compensation | 450 W |
| Nominal power | 1.9 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 3 min |
| Net weight of the unit (empty) mfr | 142 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity fluctuation | 1 % RH |
| Humidity range | 10...98 % RH |
Description
Construction and series positioning
- The KBF PRO 130 is the smallest model in the KBF PRO series, which also includes sizes 260, 470, 1060 and 1600. All share the same -20…100 °C temperature range, but the 130 is the only size with a triangular instrument panel (sizes 470 and above use a signal triangle and ergonomically adjustable control terminal).
- The 127 L interior volume and 30 kg per-rack load capacity make it suited to bench-scale or small-batch testing where floor space is limited. The empty weight is 142 kg.
- Unlike larger sizes, the 130 does not require the wall spacer (mandatory from size 260) but does use the flexible tilt protection kit (shared with size 470).
Behavior under load and in different media
- Temperature control is stable to 0.1 K fluctuation and 0.2 K uniformity across the working range. Humidity control spans 10…98 % RH with 1 % RH fluctuation, but the temperature range with humidity active is restricted to 10…90 °C.
- Recovery after a 30 s door opening is fast: 2 minutes at 25 °C/60 % RH and 3 minutes at 40 °C/75 % RH, supporting frequent access during test runs.
- The chamber handles a maximum heat compensation of 450 W, allowing samples with moderate heat dissipation to be tested without exceeding the control envelope.
- The humidity sensor has a tolerated error of ±2 % RH; readings should be interpreted with this tolerance in mind.
Installation, connection and control
- Electrical supply is single-phase 208…240 V, 50 Hz, with a nominal power of 1.9 kW and a 16 A unit fuse. The standard model (Art. No. 9020-0439) is fitted with a CEE 7/7 plug.
- The MB2 chamber controller provides fixed-value, time-program and week-program operating modes, a chart-recorder display, and alarm/notification functions. Temperature safety devices per DIN 12880 (class 1 overtemperature protection and class 2/3.3 safety controller) are integrated.
- Humidity control requires a freshwater supply and wastewater/condensate handling; the design of these systems differs by serial number (up to 20250000010225 and from 20250000010226), so the correct manual section and spare parts must be selected accordingly.
- The chamber is rated IP 20, so it must not be sprayed with water. Refrigerant is R600A (flammable hydrocarbon), and maintenance on the refrigeration system must be performed by qualified personnel trained to EN 13313:2010.
Integration & Operating Notes
The flexible tilt protection kit is specified for chamber sizes 130 and 470. It must be mounted to prevent tipping during operation or relocation. Unlike larger sizes from 260 upward, no wall spacer is required for the 130, so the chamber can be placed closer to walls, but the tilt protection still applies.
The chamber is rated for 208…240 V, 50 Hz, 1.9 kW, with a 16 A unit fuse. Connecting to a supply outside this range or with insufficient overcurrent protection can damage the unit or cause unsafe operation. Confirm the site voltage and that the circuit breaker or fuse matches the 16 A rating.
Although the chamber temperature range is -20…100 °C, humidity control is only permitted between 10 and 90 °C. Operating humidity outside this window may lead to condensation, inaccurate control or equipment stress. Plan test profiles so that humidification phases stay within this temperature band.
The maximum heat compensation is 450 W. Samples dissipating more than this will exceed the chamber's ability to maintain setpoint, potentially causing temperature deviation or safety device activation. Calculate the total heat output of all samples per test and stay well below this limit.
Regular maintenance is required at least once a year, including the steam humidifier. Condensers should be cleaned at least twice a year by a qualified technician; in dusty environments, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and repeats every two weeks until reset by BINDER Service.
The humidity sensor must not be exposed to substances above the listed tolerated concentrations. For example, ammonia permanent load is limited to 5500 ppm, hydrochloric acid to 300 ppm, and ozone to 0.5 ppm. Exceeding these limits can degrade sensor accuracy or damage it. Review sample outgassing before testing.
The humidifying system and wastewater/condensate handling differ between chambers up to serial number 20250000010225 and from 20250000010226. Identify the serial number before ordering spare parts or planning plumbing. The maximum operating pressure for water connections is 11 bar.
Selection & Integration Q&A
Choose the 130 for small samples or limited bench space: it offers 127 L and 30 kg per rack, with the same -20…100 °C range as larger models. Step up to 260 or 470 only when you need more volume or rack capacity; note that from size 260 a wall spacer is required, while the 130 uses a tilt protection kit instead.
Annual maintenance is mandatory, including the steam humidifier, and condensers should be cleaned twice a year. The base delivery does not include APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, flexible Pt 100 sensor, external freshwater/wastewater cans, or BINDER Pure Aqua Service; these must be ordered separately.
Yes, the 130 does not require the wall spacer that is mandatory from size 260 upward. However, the flexible tilt protection kit must be installed on the 130 (and 470) to prevent tipping. Ensure the floor is level and capable of supporting the 142 kg empty weight.