BINDER KBF PRO 1600 — Climate chambers with Temperature -20…100 °C.

The KBF PRO 1600 is a 1,610 L constant climate chamber with a -20…100 °C range and 10…98 % RH, the largest chamber in the KBF PRO series. It is distinguished from smaller siblings by its 40 kg per-rack load capacity and 1,000 W max heat compensation, enabling high-throughput testing of bulky or heat-dissipating samples. It serves as a programmable test platform for reproducible temperature/humidity profiles in laboratory and industrial qualification.

Temperature
-20…100 °C

BINDER KBF PRO Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF PRO 1600 Current -20–10016104010001.90.3
KBF PRO 1060 -20–10010704010001.90.3
KBF PRO 130 -20–100127304501.90.2
KBF PRO 260 -20–100262408001.60.2
KBF PRO 470 -20–100472408001.60.3
KBF PRO 720 -20–1007274010001.80.3

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 1610 L
Load per rack 40 kg
Maximum heat compensation 1000 W
Nominal power 1.9 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 3 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 7 min
Dimensions, Weight & Mechanical Design
Net weight of the unit (empty) mfr 450 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1
Operating Conditions, Environment & Protection
Humidity fluctuation 1 % RH
Humidity range 10...98 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, control and operation

Integration & Operating Notes

Mounting
Wall spacer required for 1600 size to ensure airflow

The manual specifies a wall spacer from chamber size 260 onward. For the 1600, maintain the prescribed distance to walls and adjacent equipment to prevent recirculation of exhaust air and ensure uniform temperature distribution. Failure to do so may degrade uniformity and overload the refrigeration system.

Operating Manual KBF/KBF PRO (E7), Section 4.1 Installation
Process Limits
Humidity sensor must not exceed chemical tolerance limits

The humidity sensor has defined maximum permanent load concentrations for substances such as ammonia (5500 ppm), formaldehyde (2400 ppm) and hydrochloric acid (300 ppm). Exceeding these limits can permanently damage the sensor and invalidate humidity readings. Review the resistance table in the manual before testing materials that emit vapors.

Operating Manual Section 1.14 Humidity Sensor Resistance Table
Maintenance
Annual service and twice-yearly condenser cleaning required

Regular maintenance must be performed at least once a year, including the steam humidifier. Condensers should be cleaned at least twice a year by a qualified technician; with dusty ambient air, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and requires reset by BINDER Service.

Operating Manual Section 23.2 Maintenance and Service
Process Media
Freshwater and wastewater connections must be rated for 11 bar

The chamber requires freshwater and wastewater connections. The maximum operating pressure is 11 bar, so all supply fittings and hoses must be rated accordingly. The humidifying and wastewater systems differ by serial number; identify the chamber serial number to select the correct connection method and spare parts.

Operating Manual Section 1.6 Type Plate; Sections 18.1.1–18.1.5

Selection & Integration Q&A

When should I choose the KBF PRO 1600 over the smaller KBF PRO 470 or 1060?

Choose the 1600 when you need the 1,610 L interior volume and a 40 kg per-rack load capacity for large or heavy test batches. The 470 and 1060 share the same temperature range but offer less volume and lower load ratings, so they are suitable for smaller sample sets.

What maintenance schedule and service intervals apply to the KBF PRO 1600?

Perform maintenance at least annually, including the steam humidifier. Clean condensers at least twice a year (more often in dusty environments). The controller displays a maintenance reminder after 8760 operating hours or two years, which must be reset by BINDER Service.

What accessories or consumables are not included with the base unit that I may need to order?

The base delivery does not include the wall spacer (required from size 260), freshwater/wastewater cans, APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, or flexible Pt 100 sensor. These must be ordered separately as options or accessories.