BINDER KBF PRO 470
BINDER KBF PRO 470 — Climate chambers with Temperature -20…100 °C.
The KBF PRO 470 is a 472 L constant climate chamber with a -20…100 °C range and 10…98 % RH humidity control, using a steam humidifier and R600A refrigeration circuit. It is the mid-size standard model in the KBF PRO series, distinguished from smaller 130/260 and larger 1060/1600 siblings by its 472 L interior and 40 kg per-rack load capacity. It is intended for reproducible temperature and humidity testing of components and materials in laboratory and industrial qualification processes.
BINDER KBF PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW | Temperature uniformity, K |
|---|---|---|---|---|---|---|
| KBF PRO 470 Current | -20–100 | 472 | 40 | 800 | 1.6 | 0.3 |
| KBF PRO 1060 | -20–100 | 1070 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 130 | -20–100 | 127 | 30 | 450 | 1.9 | 0.2 |
| KBF PRO 1600 | -20–100 | 1610 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 260 | -20–100 | 262 | 40 | 800 | 1.6 | 0.2 |
| KBF PRO 720 | -20–100 | 727 | 40 | 1000 | 1.8 | 0.3 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 472 L |
| Load per rack | 40 kg |
| Maximum heat compensation | 800 W |
| Nominal power | 1.6 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 4 min |
| Net weight of the unit (empty) mfr | 249 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity fluctuation | 1 % RH |
| Humidity range | 10...98 % RH |
Description
Construction and series positioning
- The KBF PRO 470 (Art. No. 9020-0441, designation 470-230) is a floor-standing constant climate chamber with a 472 L usable volume and a net weight of 249 kg. It uses a steam humidifier for humidification and an R600A refrigerant circuit for cooling, with a nominal power of 1.6 kW at 208…240 V, 1~, 50 Hz.
- Within the KBF PRO series, the 470 sits between the 260 and 1060 models. All siblings share the -20…100 °C temperature range, but the 470 offers a larger interior than the 130/260 while retaining a 40 kg per-rack load limit. The 470 is one of the sizes that uses the signal triangle instrument panel (sizes 470, 720, 1060, 1600) rather than the triangular panel of the 130/260, and it requires the flexible tilt protection kit (sizes 130 and 470).
- The chamber is equipped with a temperature safety device according to DIN 12880:2007, class 2/3.3, and an over-temperature protective device class 1. Humidity is measured by a sensor with a tolerated error of ±2 % r.h.
Behavior under load and in different media
- Temperature fluctuation is 0.1 K and temperature uniformity is 0.3 K, with a humidity fluctuation of 1 % RH. Recovery after a 30 s door opening is 2 min at 25 °C/60 % RH and 4 min at 40 °C/75 % RH, indicating fast re-stabilization after access.
- The chamber can compensate up to 800 W of heat introduced by loaded samples, which defines the maximum continuous dissipated power of test items that can be held at setpoint. The 40 kg per-rack load limit applies to the rack system; heavier loads require distribution across multiple racks or a different model.
- Humidity control is available from 10…98 % RH, but the temperature range with active humidity is restricted to 10…90 °C. Operation outside this band with humidity enabled is not permitted. The humidity sensor must not be exposed to ammonia, acetone, ethanol, formaldehyde, ozone, hydrochloric acid, hydrogen sulphide, nitrogen oxides or sulphur dioxide above the tolerated concentrations listed in the manual (e.g. NH3 permanent load 5500 ppm, HCl 300 ppm, H2S 350 ppm).
Installation, connection and control
- The chamber requires a wall spacer from size 260 upward, so the 470 must be installed with the specified rear clearance. The flexible tilt protection kit is supplied for this size and must be mounted to prevent tipping. The rear connection panel provides freshwater and wastewater connections for the humidifier; the water supply must be rated for at least 11 bar maximum operating pressure.
- Electrical connection is single-phase 208…240 V, 50 Hz, with a 16 A unit fuse. The chamber is IP 20 and must not be sprayed with water. The MB2 chamber controller provides fixed-value and program operation (time and week programs), a chart recorder display, alarm and notification functions, user management, and Ethernet/USB interfaces for data transfer. Optional analog outputs and zero-voltage alarm contacts are available for integration with external monitoring.
- The humidifying system and wastewater/condensate handling differ by serial number (up to 20250000010225 and from 20250000010226), so the correct spare parts and connection method must be selected based on the chamber serial number.
Integration & Operating Notes
The chamber must be installed with the specified wall distance (required from size 260 upward) to ensure proper airflow and operation. The flexible tilt protection kit, supplied for sizes 130 and 470, must be mounted to prevent tipping. Failure to provide the rear clearance can cause overheating and invalidate performance; omitting the tilt protection creates a safety hazard.
Although the chamber operates from -20…100 °C, active humidity control is only permitted between 10 °C and 90 °C. Operating the humidifier outside this temperature band can lead to condensation, inaccurate humidity or damage to the system. Process recipes must respect this combined temperature-humidity envelope.
The chamber can compensate a maximum heat input of 800 W from loaded test items. Exceeding this value will prevent the chamber from holding the set temperature and may cause a temperature alarm. When designing a test, calculate the total dissipated power of all powered samples and ensure it stays below this limit, especially at low setpoints.
The humidity sensor must not be exposed to ammonia, acetone, ethanol, formaldehyde, ozone, hydrochloric acid, hydrogen sulphide, nitrogen oxides or sulphur dioxide above the tolerated concentrations listed in the manual (e.g. NH3 permanent load 5500 ppm, HCl 300 ppm, H2S 350 ppm). Tests involving these substances require prior evaluation of concentration levels to avoid sensor damage and invalid humidity readings.
The chamber is rated at 208…240 V, 1~, 50 Hz with a nominal power of 1.6 kW and a 16 A unit fuse. The site supply and upstream overcurrent protection must be sized accordingly. Verify that the actual site voltage and frequency match the chamber rating before connection; incorrect supply can damage the controller and refrigeration system.
The freshwater and wastewater connections for the steam humidifier must be rated for at least 11 bar maximum operating pressure. The humidifying system design and wastewater handling differ by serial number (up to 20250000010225 and from 20250000010226), so the correct connection method and spare parts must be selected based on the chamber serial number.
Regular maintenance must be performed at least once a year, including the steam humidifier. Condensers should be cleaned at least twice a year by a qualified technician; with high dust in ambient air, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and must be reset by BINDER Service. Refrigeration work requires personnel trained to EN 13313:2010.
Selection & Integration Q&A
Choose the 470 when your sample batches fit within 472 L and each rack load stays under 40 kg, with total heat dissipation below 800 W. The 260 offers less volume for smaller loads; the 1060 provides more capacity but requires more floor space and services. All share the -20…100 °C range.
The freshwater and wastewater connections must be rated for at least 11 bar. The humidifying system and condensate handling differ by serial number (up to 20250000010225 and from 20250000010226), so verify the chamber serial number to select the correct connection method and spare parts.
Install with the specified wall spacer (required from size 260 upward) and mount the supplied flexible tilt protection kit. The chamber is IP 20, so do not spray with water. Ambient conditions must allow the specified rear clearance for proper airflow and cooling.
Plan annual maintenance including the steam humidifier, condenser cleaning at least twice a year, and annual electrical safety tests per EN 50678/VDE 0701. The maintenance reminder appears after 8760 h or two years and requires BINDER Service reset. Optional accessories include APT-COM 4 software, analog outputs, alarm contacts, LED light, door lock, Pt 100 sensor, and external water cans.