BINDER KBF PRO 470
BINDER ·Made in Germany · In Active Production

BINDER KBF PRO 470

BINDER KBF PRO 470 — Climate chambers with Temperature -20…100 °C.

The KBF PRO 470 is a 472 L constant climate chamber with a -20…100 °C range and 10…98 % RH humidity control, using a steam humidifier and R600A refrigeration circuit. It is the mid-size standard model in the KBF PRO series, distinguished from smaller 130/260 and larger 1060/1600 siblings by its 472 L interior and 40 kg per-rack load capacity. It is intended for reproducible temperature and humidity testing of components and materials in laboratory and industrial qualification processes.

Temperature
-20…100 °C

BINDER KBF PRO Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF PRO 470 Current -20–100472408001.60.3
KBF PRO 1060 -20–10010704010001.90.3
KBF PRO 130 -20–100127304501.90.2
KBF PRO 1600 -20–10016104010001.90.3
KBF PRO 260 -20–100262408001.60.2
KBF PRO 720 -20–1007274010001.80.3

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 472 L
Load per rack 40 kg
Maximum heat compensation 800 W
Nominal power 1.6 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 4 min
Dimensions, Weight & Mechanical Design
Net weight of the unit (empty) mfr 249 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1
Operating Conditions, Environment & Protection
Humidity fluctuation 1 % RH
Humidity range 10...98 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Mounting
Wall spacer and tilt protection are mandatory for the 470

The chamber must be installed with the specified wall distance (required from size 260 upward) to ensure proper airflow and operation. The flexible tilt protection kit, supplied for sizes 130 and 470, must be mounted to prevent tipping. Failure to provide the rear clearance can cause overheating and invalidate performance; omitting the tilt protection creates a safety hazard.

Operating Manual, Section 4.1 and 4.2 Installation and Connections
Process Limits
Humidity control limited to 10…90 °C chamber temperature

Although the chamber operates from -20…100 °C, active humidity control is only permitted between 10 °C and 90 °C. Operating the humidifier outside this temperature band can lead to condensation, inaccurate humidity or damage to the system. Process recipes must respect this combined temperature-humidity envelope.

Data Sheet Model KBF PRO 470, Technical Data
Process Limits
Sample heat load must not exceed 800 W continuous

The chamber can compensate a maximum heat input of 800 W from loaded test items. Exceeding this value will prevent the chamber from holding the set temperature and may cause a temperature alarm. When designing a test, calculate the total dissipated power of all powered samples and ensure it stays below this limit, especially at low setpoints.

Data Sheet Model KBF PRO 470, Technical Data
Process Limits
Humidity sensor is sensitive to specific chemical vapors

The humidity sensor must not be exposed to ammonia, acetone, ethanol, formaldehyde, ozone, hydrochloric acid, hydrogen sulphide, nitrogen oxides or sulphur dioxide above the tolerated concentrations listed in the manual (e.g. NH3 permanent load 5500 ppm, HCl 300 ppm, H2S 350 ppm). Tests involving these substances require prior evaluation of concentration levels to avoid sensor damage and invalid humidity readings.

Operating Manual, Section 1.14 Humidity Sensor Resistance Table
Electrical
Single-phase supply must be rated for 16 A unit fuse

The chamber is rated at 208…240 V, 1~, 50 Hz with a nominal power of 1.6 kW and a 16 A unit fuse. The site supply and upstream overcurrent protection must be sized accordingly. Verify that the actual site voltage and frequency match the chamber rating before connection; incorrect supply can damage the controller and refrigeration system.

Data Sheet Model KBF PRO 470, Electrical Data; Operating Manual, Section 1.6 Type Plate
Process Media
Water connections must withstand up to 11 bar

The freshwater and wastewater connections for the steam humidifier must be rated for at least 11 bar maximum operating pressure. The humidifying system design and wastewater handling differ by serial number (up to 20250000010225 and from 20250000010226), so the correct connection method and spare parts must be selected based on the chamber serial number.

Operating Manual, Section 1.6 Type Plate and Section 18.1 Humidification / Dehumidification System
Maintenance
Annual maintenance and twice-yearly condenser cleaning required

Regular maintenance must be performed at least once a year, including the steam humidifier. Condensers should be cleaned at least twice a year by a qualified technician; with high dust in ambient air, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and must be reset by BINDER Service. Refrigeration work requires personnel trained to EN 13313:2010.

Operating Manual, Section 23.2 Maintenance and Service

Selection & Integration Q&A

When should I choose the KBF PRO 470 over the smaller KBF PRO 260 or the larger KBF PRO 1060?

Choose the 470 when your sample batches fit within 472 L and each rack load stays under 40 kg, with total heat dissipation below 800 W. The 260 offers less volume for smaller loads; the 1060 provides more capacity but requires more floor space and services. All share the -20…100 °C range.

What are the requirements for the water supply and drainage when integrating the humidifier?

The freshwater and wastewater connections must be rated for at least 11 bar. The humidifying system and condensate handling differ by serial number (up to 20250000010225 and from 20250000010226), so verify the chamber serial number to select the correct connection method and spare parts.

What mounting and environmental conditions are mandatory for the KBF PRO 470?

Install with the specified wall spacer (required from size 260 upward) and mount the supplied flexible tilt protection kit. The chamber is IP 20, so do not spray with water. Ambient conditions must allow the specified rear clearance for proper airflow and cooling.

What maintenance intervals and accessories are required beyond the base delivery?

Plan annual maintenance including the steam humidifier, condenser cleaning at least twice a year, and annual electrical safety tests per EN 50678/VDE 0701. The maintenance reminder appears after 8760 h or two years and requires BINDER Service reset. Optional accessories include APT-COM 4 software, analog outputs, alarm contacts, LED light, door lock, Pt 100 sensor, and external water cans.