VD 23 drawing
BINDER ·Made in Germany · In Active Production

BINDER VD 23

BINDER VD 23 — Drying ovens with Temperature 9…220 °C.

The BINDER VD 23 is a 24 L vacuum drying chamber with a 16 mm small-flange vacuum connection, operating from +9 to 220 °C with 0.1 K fluctuation and 1.0 K uniformity. As the smallest of the VD 23/56/115 series, it suits benchtop loads up to 50 kg (20 kg per rack) where space and sample volume are limited. It is used for gentle drying of non-flammable solvents under vacuum, with evacuation, ventilation and inert-gas operation managed by the RD4 controller.

Temperature
9…220 °C

BINDER VD Series — Model Comparison & Selection

Model Temperature, °CDepth, mmHeating up time to 100 c, minHeight, cmInterior volume, LPermitted load, kg
VD 23 Current 9–2204138069.82450
VD 115 9–22058113094.211970
VD 56 9–2204618081.55560

Full specifications

Core Performance & Functional Specifications
Temperature 9...220 °C
Heating time to 100°C 80 min
Interior volume 24 L (VD023-230V)
Leak rate 0.01 bar/h
Load per rack 20 kg (VD023-230V)
Nominal power 0.9 kW
Permitted load 50 kg (VD023-230V)
Pressure fluctuation over time 1 mbar
Temperature fluctuation 0.1 K
Temperature uniformity 1.0 K
Rear wall clearance 100 mm (VD023-230V)
Side wall clearance 70 mm (VD023-230V)
Article Number mfr 9630
Pressure fluctuation over time at 400 mbar mfr 70 mbar
Dimensions, Weight & Mechanical Design
Depth 413 mm (VD023-230V)
Height 698 mm (VD023-230V)
Width 523 mm (VD023-230V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1
Interfaces, Connections & Control
Vacuum connection with small flange mfr 16 mm

Description

Construction and series positioning

Behavior under load and vacuum

Installation, control and operation

Integration & Operating Notes

Electrical
Match the voltage variant to the site supply before ordering

The 230 V version (Art. 9630-0001) requires 200...230 V 1~ 50/60 Hz with a CEE 7/7 plug and 6.3 A unit fuse; the UL 120 V version (Art. 9630-0005) uses a NEMA 5-15 plug. Connecting the wrong variant risks damage and invalidates safety certification; check the type plate and supply voltage during commissioning.

Data Sheet Model VD 23 en.pdf; Operating Manual VD (E3.1) 07-2025, chap. 25.4
Process Limits
Do not exceed 50 kg total or 20 kg per rack

The permitted load is 50 kg for the chamber and 20 kg per expansion rack. Exceeding these limits can deform racks or the chamber interior and compromise temperature uniformity. Distribute samples evenly and use only the delivered racks, keeping rack material consistent (no mixing of aluminum and stainless steel).

Factory Specification Sheet; Operating Manual VD (E3.1) 07-2025, chap. 25.4
Process Limits
Use only non-flammable solvents in the chamber

The VD 23 is intended for drying non-flammable solvents. Introducing flammable substances creates a fire or explosion risk because the chamber is not designed for them. Verify solvent compatibility with the stainless-steel interior; corrosive damage from certain ingredients is excluded from BINDER liability.

Data Sheet Model VD 23 en.pdf; Operating Manual VD (E3.1) 07-2025, chap. 1.9
Control & I/O
Configure safety-controller value above the process setpoint

If the “Safety controller” alarm appears, the inner chamber temperature has reached the safety controller value. Set the safety controller value above the intended temperature setpoint and check for heating-system errors. After the alarm, let the chamber cool, acknowledge the alarm, and verify both values before restarting.

Operating Manual VD (E3.1) 07-2025, chap. 25.4
Maintenance
Schedule annual maintenance and electrical safety testing

Maintenance must be performed at least once a year, and an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Maintenance by non-authorized personnel voids the warranty. Replace the door gasket only when cold to avoid damaging it.

Operating Manual VD (E3.1) 07-2025, chap. 25.1 and 25.2
Maintenance
Check the chamber fuse first if the unit stays off

If the chamber is permanently turned off, troubleshooting includes checking the chamber fuse (6.3 A). A blown fuse may indicate an electrical fault; disconnect the power plug before any maintenance work and do not remove the rear panel. Only licensed electricians or BINDER-authorized experts should perform maintenance.

Operating Manual VD (E3.1) 07-2025, chap. 25.2 and 25.4

Selection & Integration Q&A

When should I choose the VD 23 over the VD 56 or VD 115?

Choose the VD 23 for small loads and limited bench space: 24 L interior, 50 kg total load, 20 kg per rack. The VD 56 and VD 115 offer larger interiors for higher throughput but share the same +9 to 220 °C range and vacuum principle. Select the VD 23 when sample volume fits 24 L and portability or footprint matters.

What vacuum pump and connections are required for the VD 23?

The chamber has a 16 mm small-flange vacuum connection. A vacuum pump is not included; BINDER offers a pump module with pump stand and connection kit, and VD connection kits for VP 1 / VP 2 pumps. Ensure the pump is suitable for the 0.01 bar/h leak rate and 1 mbar pressure fluctuation specification.

What maintenance and accessories are needed beyond the base delivery?

Annual maintenance is required, plus electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. The base delivery includes the chamber and expansion racks; a vacuum pump, pump module, connection kits, APT-COM 4 software, analog outputs, Pt 100 object sensor, LED lighting and measuring access port are optional accessories. Door gaskets are consumables replaced when cold.

Dimensional Drawings & Technical Documentation

Dimensional drawings

VD 23 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model VD 23 en
Datasheet ·353,1 КБ ·7 pp.
View PDF
PDF
VD (E3.1) 07 2025 en
Manual ·10,8 МБ ·181 pp.
View PDF