BINDER VD 56
BINDER VD 56 — Drying ovens with Temperature 9…220 °C.
The BINDER VD 56 is a 55 L vacuum drying chamber with a 16 mm small-flange vacuum connection, operating from 9 to 220 °C at 1.4 kW. It is the mid-size unit in the VD series (VD 23 / VD 56 / VD 115), distinguished by a 60 kg total load capacity and 20 kg per rack. It is used for gentle, oxidation-free drying of non-flammable solvents under vacuum in laboratory and production processes.
BINDER VD Series — Model Comparison & Selection
Full specifications
| Temperature | 9...220 °C |
| Heating time to 100°C | 80 min |
| Interior volume | 55 L (VD 56) |
| Leak rate | 0.01 bar/h |
| Load per rack | 20 kg (VD 56) |
| Nominal power | 1.4 kW |
| Permitted load | 60 kg (VD 56) |
| Pressure fluctuation over time | 1 mbar |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 1.2 K |
| Rear wall clearance | 100 mm (VD 56) |
| Side wall clearance | 70 mm (VD 56) |
| Article Number mfr | 9630 |
| Pressure fluctuation over time at 400 mbar mfr | 70 mbar |
| Depth | 461 mm (VD 56) |
| Height | 815 mm (VD 56) |
| Width | 638 mm (VD 56) |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 8.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Vacuum connection with small flange mfr | 16 mm |
Description
Construction and series positioning
- The VD 56 is the middle chamber of the VD series; siblings VD 23 and VD 115 share the same 9...220 °C range but differ in volume and load capacity.
- Interior volume is 55 L with net dimensions 638 mm wide × 815 mm high × 461 mm deep; the inner chamber is made of especially corrosion-resistant stainless steel.
- The standard controller is the RD4 microprocessor program controller; the MB2 controller is optional. Both are covered by the operating manual.
- Two electrical variants exist: VD056-230V (200...230 V, 1~, 50/60 Hz, CEE 7/7 plug, Art. 9630-0002) and VD056UL-120V (100...120 V, 1~, 50/60 Hz, NEMA 5-15 plug, Art. 9630-0006).
Behavior under process conditions
- Temperature uniformity is 1.2 K and temperature fluctuation 0.1 K; heating-up time to 100 °C is 80 min.
- The vacuum system supports evacuation, ventilation after drying (ambient air or inert gas), ventilation/breaking vacuum after power failure, and inert-gas operation.
- Leak rate is 0.01 bar/h; pressure fluctuation over time is 1 mbar (70 mbar at 400 mbar per extra specs).
- The chamber is equipped with a temperature safety device class 2.0 per DIN 12880 (overtemperature protective device class 2 thermal switch and overtemperature safety controller class 2).
Installation, connection and control
- Required wall clearances: 100 mm at the rear and 70 mm at the sides.
- The vacuum connection is a 16 mm small flange; inert gas supply and pump module connections are described in the manual.
- The RD4 controller provides Ethernet and USB interfaces; APT-COM 4 software enables networking. Network settings (IP, subnet, gateway, DNS) and USB data transfer are configurable.
- The chamber has IP protection 20 per EN 60529 and must not become wet during operation or maintenance.
Integration & Operating Notes
Maintenance work must be performed at least once a year; warranty becomes void if maintenance is conducted by non-authorized personnel. An annual repeat inspection and a test prior to initial startup and after maintenance or repair are required for electrical safety (EN 50678/VDE 0701 and EN 50699/VDE 0702 recommended). The door gasket must be replaced only when cold.
If the chamber is permanently turned off, troubleshooting includes checking the chamber fuse (8.0 A unit fuse) before contacting BINDER Service. Also verify the supply voltage matches the variant (115 V or 230 V). Do not remove the rear panel; disconnect the power plug before any maintenance work.
Selection & Integration Q&A
Choose the VD 56 for a 55 L interior with 60 kg total load and 20 kg per rack. The VD 23 is smaller; the VD 115 larger. All share the 9...220 °C range, so select by volume and load, not temperature.
The chamber has a 16 mm small-flange vacuum connection. A pump module with pump stand and connection kit is available as an accessory; VD connection kits for VP 1 / VP 2 pumps are listed. Plan inert gas supply if ventilation with gas is needed.
Annual maintenance and electrical safety testing are required. The base delivery includes factory calibration and expansion racks; options include APT-COM 4 software, analog outputs, measuring access port, Pt 100 object sensor, LED lighting, and pump module. Door gasket replacement only when cold.