Peter Huber Kältemaschinenbau CC-130A Visco 3 Visco-Thermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-130A Visco 3 Visco-Thermostat

Peter Huber Kältemaschinenbau CC-130A Visco 3 Visco-Thermostat — Thermostatic baths.

The CC-130A Visco 3 is a 30-litre viscometer thermostat bath built around a variable-speed pressure/suction pump and the Pilot ONE controller. Its 28–100 °C native range extends down to 15 °C only when water cooling or a refrigerator is added. It holds ±0.01 K at 70 °C for capillary viscometry and density work.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Bath volume 30 l
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 28...100 °C
with cooling unit: 15...100 °C · with water cooling: 15...100 °C
Temperature stability mfr ±0, 01 K
Dimensions (W x D x H) mfr 500 x 240 x 490 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity pressure pump max. mfr 27 l/min; 0, 7 bar l/min; bar
Bath opening WxD/BD mfr 90 x 90 / 310 mm
Filling volume min. mfr 25, 5 l
Dimensions, Weight & Mechanical Design
Weight 14 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 200-240V 1~/2~ 50/60Hz V, Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection and a 10 A minimum fuse

The unit draws up to 10 A on a 200–240 V single- or two-phase supply and requires external branch circuit protection according to UL 489. Undersized protection will trip during heat-up; the supply must also stay within ±10% on voltage and frequency without the two deviations running in opposite directions.

CC-130A Visco 3 Immersion Circulator Datasheet, pages 1–2
Cooling & Media
Sub-ambient setpoints need water cooling or a refrigerator

Without cooling the bath only reaches 28 °C; 15 °C operation requires the integral cooling coil plumbed to a cooling water supply or a refrigerator. Plan the cooling water connection and its temperature before specifying a method that runs below ambient.

CC-130A Visco 3 Immersion Circulator Datasheet, page 1
Maintenance
Order drain valve and viscometer holder separately

A drain valve and a holder for Ubbelohde viscometers (#9586) are optional accessories and are not part of the base delivery. Without the drain valve, emptying the 30 l bath for fluid changes must be done another way; without the holder, viscometers must be supported independently.

CC-130A Visco 3 Immersion Circulator Datasheet, page 2

Selection & Integration Q&A

When should I choose the CC-130A Visco 3 rather than another bath in the Visco series?

Pick it when you need a 30 l open bath with three 90 × 90 mm openings for viscometers and 0.01 K stability at 70 °C. If your method runs below 28 °C, budget for water cooling or a refrigerator to reach 15 °C.

What must the external circuit and power supply provide to connect this bath?

A 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10 A minimum fuse and UL 489 branch protection, plus a country-specific plug specified at order. External loop connects via M16×1 male, up to 27 l/min at 0.7 bar.

What site and duty conditions limit where this bath can be installed?

Ambient must stay 5–40 °C and the IP20 enclosure must be kept dry. Supply voltage and frequency may deviate ±10% only if not in opposite directions. Overtemperature cut-out must be set before unattended runs.

What is not included in the base delivery and must be ordered or planned separately?

The drain valve and the Ubbelohde viscometer holder (#9586) are optional accessories. The country-specific power plug must be specified at ordering. Cooling water supply or a refrigerator is needed for 15 °C operation.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-130A Visco 3 Immersion Circulator Datasheet
Datasheet ·73,9 КБ ·2 pp.