Peter Huber Kältemaschinenbau CC-E xd
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-E xd

Peter Huber Kältemaschinenbau CC-E xd — Thermostatic baths.

The CC-E xd is a 195 mm immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 200–240 V single- or two-phase supply. Its 25–200 °C base range extends to 20–200 °C with water cooling or −30–200 °C with a refrigerator, positioning it as the cold-capable member of the Einhaengethermostate series. It circulates heat transfer fluid through external circuits at up to 22 l/min.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Einhaengethermostate Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minuteWeight, kgAmbient temperature limit, °CCurrent consumption, A
CC-E xd Current Pilot ONEEthernet, USB (Host u. D1755–4010
KISS E KISSUSB (Device), RS232 Inte1445–4010

Full specifications

Core Performance & Functional Specifications
Maximum delivery flow rate 22 l/min (pressure pump max.; 0, 4 bar)
suction pump max.; 0, 25 bar: 17 l/min
Pump connection M16x1 AG male (Option, #39593) (Option, #39593)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 01 K
Dimensions (W x D x H) mfr 132 x 159 x 360 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Dimensions, Weight & Mechanical Design
Weight 5 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz V, Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-E xd is an immersion circulator (Einhaengethermostat) with a 195 mm immersion depth, 132 × 159 × 360 mm housing and 5 kg weight. Unlike the plain CC-E sibling, the xd modification carries a variable-speed pressure and suction pump, enabling both delivery and return-side suction in closed external circuits. The KISS E is the simpler neighbor in the same series. Heating capacity scales with site voltage: 1.5 kW at 200 V, 1.6 kW at 208 V, 1.8 kW at 220 V, 2.0 kW at 230 V and 2.1 kW at 240 V.

Thermal behavior and media

Control, installation and operation

The Pilot ONE controller provides Ethernet, USB host/device and RS232 interfaces. Supply is 200–240 V 1~/2~ 50/60 Hz with maximum current consumption of 10 A; voltage tolerance is ±10% provided the frequency tolerance does not run in the opposite direction. External branch circuit protection to UL 489 is required, and single/two-phase units ship with a power cable and country-specific plug that must be specified at order. Enclosure rating is IP20, safety class III/FL, and EMC classification is Class A, Group 1 to EN55011. Ambient operating temperature must remain between 5 and 40 °C.

Integration & Operating Notes

Electrical
Branch protection to UL 489 must be provided externally

The unit does not carry its own branch circuit protection to UL 489; the electrical panel design must include it. Plan the feeder and protective device accordingly, since the unit draws up to 10 A.

Datasheet CC-E xd Immersion Circulator, page 2
Process Limits
Low-temperature limit depends on the cooling method chosen

Without auxiliary cooling the range starts at 25 °C; water cooling lowers it to 20 °C and a refrigerator to −30 °C. The cooling option must be decided at selection, because it defines the achievable process temperature and cannot be inferred from the base unit.

Datasheet CC-E xd Immersion Circulator, page 1
Cooling & Media
Polyglycol is not permissible with the Huber pump adapter

When the Huber pump adapter is used, polyglycol must not be selected as heat transfer fluid. Confirm fluid compatibility with the pump circuit before filling, since an incompatible fluid can damage the pump.

Datasheet CC-E xd Immersion Circulator, page 2
Electrical
Country-specific plug must be specified at order

Single- and two-phase units are delivered with a power cable and a country-specific plug. State the destination country when ordering so the correct plug is supplied; otherwise the unit may arrive with a plug unsuitable for the site.

Datasheet CC-E xd Immersion Circulator, page 2

Selection & Integration Q&A

When should I choose the CC-E xd rather than the plain CC-E or KISS E in the same series?

Pick the xd when the process needs external circuit circulation with suction-side return and cold operation down to −30 °C with a refrigerator. The plain CC-E and KISS E are the simpler neighbors; the xd's variable-speed pressure/suction pump and Pilot ONE interfaces are its differentiators.

What supply and protection must be provided for the CC-E xd, and what fluid restrictions apply to the pump circuit?

Provide 200–240 V 1~/2~ 50/60 Hz with ±10% voltage tolerance not opposed by frequency deviation, plus external UL 489 branch protection; the unit draws up to 10 A. Specify the country plug at order. Do not use polyglycol with the Huber pump adapter.

What is not included in the base delivery that must be planned for?

The base unit ships with a power cable and country-specific plug only when specified at order; external UL 489 branch protection, the cooling source (water cooling or refrigerator) needed for sub-25 °C operation, and the Pt100 external sensor are separate items to plan.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-E xd Immersion Circulator
Datasheet ·17,7 КБ ·2 pp.