Peter Huber Kältemaschinenbau KISS E
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS E

Peter Huber Kältemaschinenbau KISS E — Thermostatic baths.

The KISS E is an immersion circulator with a pressure/suction pump and 150 mm immersion depth, delivering 14 l/min at 0.25 bar. Its base range is 25–200 °C, extended to 20–200 °C with water cooling or −30–200 °C with a refrigerator. It serves as the temperature-control head for open baths in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Einhaengethermostate Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minuteWeight, kgAmbient temperature limit, °CCurrent consumption, A
KISS E Current KISSUSB (Device), RS232 Inte1445–4010
CC-E xd Pilot ONEEthernet, USB (Host u. D1755–4010

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Dimensions (w × D × H) 132 x 163 x 312 mm
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Dimensions, Weight & Mechanical Design
Weight 4 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and media

Installation and operation

Integration & Operating Notes

Electrical
Supply circuit must be rated 10–16 A with UL 489 breaker

The unit draws up to 10 A at 200–240 V; the upstream fuse must be at least 10 A and no more than 16 A. An external branch circuit protection device according to UL 489 is mandatory, so the panel or installation must include a listed breaker sized within this window.

Datasheet, Technical Data, page 1–2
Process Limits
Cooling method sets the achievable lower temperature

Without auxiliary cooling the unit operates from 25 °C upward; water cooling extends the range to 20 °C, and a refrigerator to −30 °C. Select the cooling configuration before specifying the process window, because the base unit alone cannot go below 25 °C.

Datasheet, Overview, page 1
Cooling & Media
Avoid acetone and polyglycols with the plastic pump

The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection, use a water mixture with glysantine or ethylene glycol. Fluid choice must be checked against pump wetted materials before commissioning.

Datasheet, page 2
Mounting
Respect 150 mm immersion depth in the bath

The thermostat must be mounted so that the 150 mm immersion depth is maintained in the bath. This ensures correct temperature sensing and pump operation; insufficient depth may cause inaccurate control or pump cavitation.

Datasheet, Technical Data, page 1
Control & I/O
Autostart after power failure may restart heating unattended

An autostart function for power failure is included. For processes where automatic restart is hazardous, the restart behavior must be assessed and external interlocks or supervision provided; the unit will resume operation when power returns.

Datasheet, page 1
Environment
Install only in dry, protected locations within 5–40 °C ambient

IP20 protection means the unit is not dust-tight or water-resistant; it must be installed in a dry, protected environment. Ambient temperature must remain between 5 °C and 40 °C for proper operation.

Datasheet, page 1

Selection & Integration Q&A

When should I choose the KISS E instead of a sibling model like CC-E or CC-E xd?

Choose KISS E when you need a basic immersion circulator with 25–200 °C range (extendable with cooling) and a pressure/suction pump. Sibling models may offer different features; verify their specifications for extended ranges or interfaces not covered here.

What are the requirements for the electrical supply and branch protection?

Provide 200–240 V 1~/2~ 50/60 Hz with ±10% tolerance (not opposing). Use a branch circuit breaker per UL 489 rated between 10 A and 16 A; the unit draws up to 10 A.

What maintenance and accessories are needed beyond the base unit?

The base delivery does not include the optional M16x1 AG male pump connection (#19606). For freeze protection, use glysantine or ethylene glycol mixtures. No specific maintenance schedule is given in the provided digest; consult the manual.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS E Immersion Circulator
Datasheet ·110,9 КБ ·2 pp.