Peter Huber Kältemaschinenbau DC32
Peter Huber Kältemaschinenbau DC32 — Recirculating chillers.
The DC32 is a recirculating chiller with a rated operating range of -30 to 50 °C and no integrated pump, so it functions as a cooling stage in an externally pumped loop. Its cooling capacity falls from 0.6 kW at 15 °C to 0.12 kW at -20 °C, making it suitable only for low-load process cooling within that envelope. It uses R-290 refrigerant and an ON-OFF controller.
Peter Huber Kältemaschinenbau Durchflusskuehler Series — Model Comparison & Selection
Full specifications
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | no pump |
| Temperature range | -30...50 °C |
| Dimensions (W x D x H) mfr | 280 x 340 x 465 mm |
| Weight | 30 kg |
| Current consumption | 3 A (max.) |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Controller | ON-OFF-controller |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity | 0.08 kg (1st stage) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Cooling capacity | 0.6 kW (15 °C) 0 °C: 0.47 kW · -20 °C: 0.12 kW |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and positioning
- The DC32 is a compact air-cooled recirculating chiller (280 x 340 x 465 mm, 30 kg) with a single-stage R-290 refrigerant circuit (0.08 kg charge).
- Unlike series siblings DC30 and DC31, this model is supplied without a pump; the fluid circuit terminates in an M16x1 male connection for an external pump or customer loop.
- The controller is an ON-OFF type, not a modulating or PID controller, so temperature is maintained by switching the refrigeration circuit.
Behavior under load
- Cooling capacity is strongly temperature-dependent: 0.6 kW at 15 °C, 0.47 kW at 0 °C, and 0.12 kW at -20 °C. Sizing must use the capacity at the actual operating point, not the 15 °C value.
- The rated operating range is -30 to 50 °C; operation outside this window is not supported.
- Output data are valid at a 20 °C room temperature; higher ambient temperature may reduce cooling capacity.
Installation and operation
- Power supply is 220–240 V, 1~/2~, 50/60 Hz with a maximum current of 3 A. Upstream protection must be between 10 A and 16 A.
- The unit is IP20 and must be installed indoors in a dry, protected environment with ambient temperature between 5 and 40 °C.
- Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
- Space requirements around the unit must be respected as stated in the operating conditions.
Integration & Operating Notes
The refrigeration circuit delivers 0.6 kW at 15 °C but only 0.12 kW at -20 °C. Selecting the DC32 for a load that requires more than 0.12 kW at -20 °C will result in insufficient cooling and possible process deviation. Always size using the capacity at the lowest intended operating temperature.
The datasheet specifies a minimum fuse of 10 A and a maximum fuse of 16 A. A lower-rated breaker may trip during compressor start-up, while a higher-rated one may not protect the unit. Verify the supply voltage is 220–240 V and frequency 50/60 Hz within the allowed tolerances.
The IP20 enclosure is not protected against water or dust ingress, so the installation site must be dry and sheltered. Ambient temperature must remain between 5 and 40 °C; outside this range reliable operation is not guaranteed and cooling capacity may drop further.
The DC32 has no integrated pump and terminates in an M16x1 male thread. The customer must provide an external pump, hoses, and fittings matching this thread. Pump speed cannot be adjusted from the chiller because no pump control is present.
Single/two-phase devices are delivered with a power cable and a country-specific plug. The plug variant must be specified at the time of ordering; otherwise the unit may arrive with a plug unsuitable for the site.
The refrigerant is R-290 (A3, H220), classified as flammable. Installation and service personnel must be aware of the safety measures for A3 refrigerants, including avoiding ignition sources and ensuring adequate ventilation during service.
Selection & Integration Q&A
Choose the DC32 when the loop already has an external pump and you need cooling down to -30 °C with up to 0.6 kW at 15 °C. The DC30 and DC31 are siblings in the same series; select them if their published configurations better match your required temperature range or pump arrangement.
The DC32 has no pump and provides an M16x1 male fluid connection. You must supply an external pump and hoses or fittings with a matching M16x1 female thread. Pump speed cannot be controlled by the chiller, so the external pump must be set independently.
The base delivery includes the chiller with a power cable and a country-specific plug that must be specified at ordering. No pump, hoses, or fittings are included. The R-290 refrigerant charge is factory-filled; service consumables are not part of the scope.