Labconco Basic Laboratory Hoods
Labconco · In Active Production

Labconco Basic Laboratory Hoods

Labconco Basic Laboratory Hoods — Fume hoods.

The Protector Basic 47 is a 4-foot (47-inch) low-profile fume hood with 25-inch depth, epoxy-coated steel construction, and by-pass airflow rated at 785 CFM. It is the narrowest of the Basic series (47 and 70 inch), available assembled or unassembled, with or without blower, in 115 V/60 Hz or 220 V/50 Hz. It serves as a general-purpose containment enclosure for routine laboratory work where a 10-inch exhaust duct and 18-inch sash stop are acceptable.

Full specifications

Core Performance & Functional Specifications
Airflow 785 CFM
Dimensions, Weight & Mechanical Design
Width 4 ft
Sash stop height mfr 18.0 in
Work surface depth mfr 1.25 in
Overall width (A) mfr 72.0 in
Dimension B mfr 67.5 in
Dimension C mfr 32.5 in
Dimension D mfr 36.0 in
Electrical, Power & Drive Specifications
Electrical requirements mfr 208 A

Description

Construction and series positioning

Airflow and containment behavior

Installation, control and operation

Integration & Operating Notes

Mounting
Duct weight must not bear on hood or blower housing

The exhaust ductwork must be supported independently by hangers or brackets attached to the building structure. Allowing duct weight to rest on the blower scroll or hood panels can distort the housing, misalign the blower, and compromise containment. Verify support before connecting the 10-inch duct.

Instruction Manual, Chapter 2 Installation Factors and Chapter 3, p.14–16
Process Limits
Perchloric acid and radioisotopes are excluded from this hood

The manual explicitly states perchloric acid should not be used and radioisotope materials are not recommended. Flammable or explosive materials are permitted only when the hood is equipped with explosion-proof components. Selecting this standard model for those applications would violate the manufacturer's operating limits.

Instruction Manual, Chapter 4 Normal Operation, p.18
Process Limits
Service-line pressure must be regulated to 40 PSI

Service fixtures require 1/4-inch OD copper (brass for gas) tubing with an accessible shut-off valve. Recommended operating pressure is 40 PSI; maximum allowable is 200 PSI. If building supply exceeds 40 PSI, install a pressure regulator upstream to protect fixtures and maintain consistent flow.

Instruction Manual, Chapter 2 Service Line Requirements, p.15
Maintenance
Lubricate blower motor annually after first two years

The exhaust blower motor is factory-lubricated for two years. Thereafter apply 20 to 30 drops of SAE 20 automotive oil annually. Check belt tension at the same interval: approximately 5 pounds pressure should deflect the belt 1/2 inch from its normal position. These actions extend motor and belt life and maintain rated airflow.

Instruction Manual, Appendix B Routine Maintenance, p.24
Control & I/O
Sash weight ties must be released before first operation

During shipment the sash weight is secured with two plastic wire ties. Before operation, cut the ties, remove cardboard packing, slide the weight up, and attach the sash cables into their slots. Operating with the weight still tied will prevent proper sash balance and may damage the cable system.

Instruction Manual, Chapter 2 Sash Weight Release, p.13
Mounting
Lift only by sides and back panel

The sash foil is not directly attached to the hood body and will detach if used as a lifting point. Rig or lift the hood only by its sides and back panel to avoid damaging the foil or cabinet structure during placement.

Instruction Manual, Chapter 2 Installation Factors, p.12

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Educational Fume Hoods & Enclosures Catalog
Brochure ·6,9 МБ ·32 pp.
View PDF
PDF
Basic Laboratory Hoods Manual
Manual ·1,3 МБ ·29 pp.
View PDF