Shimadzu Altrace
Shimadzu Altrace — Elemental analyzers.
The Shimadzu ALTRACE is an energy-dispersive X-ray fluorescence (EDX) elemental analyzer configured with a 48-sample autosampler, contamination-proof cup, and six automatically exchanged filters. It delivers detection limits of 1 ppm for Pb and Cd in aqueous solution (600 s), a five-fold improvement over the EDX-7200, and covers sub-ppm to percent levels in batch analysis. It is intended for automated multi-element screening in production and environmental laboratories.
Full specifications
| Detection limit (Pb in aqueous solution, 600 seconds measurement time) mfr | 1 ppm (Pb in aqueous solution, 600 seconds measurement time) |
| Detection limit (Pb, EDX-7200 comparison, 600 seconds measurement time) mfr | 5 ppm (Pb, EDX-7200 comparison, 600 seconds measurement time) |
| Detection limit (Cd in aqueous solution, 600 seconds measurement time) mfr | 1 ppm (Cd in aqueous solution, 600 seconds measurement time) |
| Detection limit (Cd, EDX-7200 comparison, 600 seconds measurement time) mfr | 5 ppm (Cd, EDX-7200 comparison, 600 seconds measurement time) |
| Measurement range (batch element analysis) mfr | sub-ppm to % (batch element analysis) |
| Measurement time (detection limit measurements) mfr | 600 seconds (detection limit measurements) |
| Sample replacement mfr | 48-sample autosampler |
| Protective function mfr | Contamination-proof cup |
| Data processor SSD mfr | 500 GB min. |
| Filter #1 energy range (elements Mo,Rh,Pd,Ag,Cd,Sn,Sb) mfr | 14 to 38 keV (elements Mo,Rh,Pd,Ag,Cd,Sn,Sb) |
| Filter #2 energy range (elements S,Cl) mfr | 2 to 4 keV (elements S,Cl) |
| Filter #3 energy range (elements Cr,Mn,Fe,Co,Ni) mfr | 5 to 10 keV (elements Cr,Mn,Fe,Co,Ni) |
| Filter #4 energy range (element Hg) mfr | 9 to 10 keV (element Hg) |
| Filter #5 energy range (elements As,Br,Zr,Hg,Pb,Bi) mfr | 9 to 14 keV (elements As,Br,Zr,Hg,Pb,Bi) |
| Filter #6 energy range (elements Ti,V) mfr | 4 to 5 keV (elements Ti,V) |
| X-ray generator cooling method mfr | Air-cooled (with fan) |
Description
Construction and positioning
- The ALTRACE is a benchtop EDX analyzer with an air-cooled X-ray generator (fan-assisted), eliminating the need for external water cooling.
- Sample handling is automated via a 48-position autosampler; each sample is presented in a contamination-proof cup to protect the instrument from spills and debris.
- Six motorized filters are exchanged automatically to optimize excitation for specific element groups: Filter #1 (14–38 keV) for Mo, Rh, Pd, Ag, Cd, Sn, Sb; Filter #2 (2–4 keV) for S, Cl; Filter #3 (5–10 keV) for Cr, Mn, Fe, Co, Ni; Filter #4 (9–10 keV) for Hg; Filter #5 (9–14 keV) for As, Br, Zr, Hg, Pb, Bi; Filter #6 (4–5 keV) for Ti, V.
- A 500 GB SSD data processor stores spectra and results; no external PC is required for basic operation.
Behavior under load and in different media
- Detection limits are specified for aqueous solutions at 600 s measurement time: 1 ppm for Pb and Cd, compared with 5 ppm for the EDX-7200 under identical conditions.
- The measurement range spans sub-ppm to percent levels for batch element analysis, allowing both trace screening and major-component quantification without hardware change.
- The vacuum measurement unit (referenced in the manual) enables light-element analysis by removing air attenuation; the air-cooled generator supports continuous batch operation without a water chiller.
Installation, connection, and control
- The instrument requires only electrical power and a ventilation path for the air-cooled generator; no facility water or compressed air is specified in the available documentation.
- Sample cups are loaded into the autosampler tray; the contamination-proof cup design is intended to contain liquid or powdered samples and simplify cleaning.
- Filter selection and measurement sequencing are automatic, driven by the method; the operator defines element groups and measurement times in the software.
- The 500 GB SSD provides local storage for methods, spectra, and results; data export and network integration are not detailed in the supplied digest.
Integration & Operating Notes
The 1 ppm Pb/Cd detection limits are specified for 600-second acquisitions in aqueous solution. Shorter measurement times will raise the achievable detection limit; methods must budget sufficient count time per element group to meet trace-level requirements.
Each of the six filters is optimized for a specific energy range (e.g., Filter #2 for S and Cl at 2–4 keV; Filter #5 for As, Br, Zr, Hg, Pb, Bi at 9–14 keV). Using the wrong filter for an element will degrade sensitivity; method setup must assign the correct filter per analyte.
Samples are presented in a contamination-proof cup. Inspect and replace cups as needed to prevent spills from reaching the X-ray chamber; the cup is the primary barrier against contamination that could invalidate results or require service.
The X-ray generator is air-cooled with a fan, so no water chiller or facility cooling loop is required. Ensure adequate clearance around the instrument for airflow and service access; ambient temperature and humidity limits are not specified in the available digest.
The data processor includes a 500 GB SSD for methods, spectra, and results. Network or LIMS connectivity is not described in the supplied documentation; plan data transfer via the available ports or manual export.
Selection & Integration Q&A
Choose the ALTRACE when 1 ppm detection limits for Pb and Cd (600 s, aqueous) are required; the EDX-7200 specifies 5 ppm under the same conditions. The ALTRACE also adds a 48-sample autosampler for unattended batch runs.
Aqueous samples are measured in the contamination-proof cup, which is part of the standard configuration. The cup contains the liquid and protects the chamber; no additional consumables are specified in the available documentation.
The generator is air-cooled, so continuous operation depends on adequate airflow and ambient conditions; specific duty-cycle and temperature limits are not stated in the supplied digest. The 48-sample autosampler supports extended unattended runs.
Inspect and replace contamination-proof cups as needed to prevent chamber contamination. The base delivery includes the 48-sample autosampler and six filters; spare cups, calibration standards, and any external data storage are not listed and should be ordered separately.