Waters SFC Prep 150 MGM System for Scalable Sample Purification
Waters ·Made in USA · In Active Production

Waters SFC Prep 150 MGM System for Scalable Sample Purification

Waters SFC Prep 150 MGM System for Scalable Sample Purification — Chromatographs with Flow rate 150 milliliter/minute, Max pressure 30 megapascal.

The Waters SFC Prep 150 MGM System is a preparative supercritical fluid chromatography (SFC) instrument configured with a Mass-Guided (MGM) fraction collection system, enabling scalable sample purification. It delivers flow rates from 0.50 to 150.00 mL/min (combined CO2 and co-solvent) at pressures up to 6000 psi (roll-off to 5000 psi at 150 mL/min), with a maximum pressure rating of 10,000 psi. This modification is distinguished by its MGM detection and automated fraction collection, making it suitable for high-recovery purification of compounds from complex mixtures.

Flow rate
150 milliliter/minute
Max pressure
30 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 150 mL/min
Max pressure 300 bar
Injection volume 9500 µL
Operating flow rate range mfr 20 to 150 mL/min, combined CO2 and co-solvent
Recovery mfr >90% (within total flow rate range)
Pressure rating mfr 10,000 psi
Max operating pressure mfr 8700 psi
Flow range mfr 20 to 200 g/min
Wetted materials mfr 316 stainless-steel, 440C, 17–4PH stainless-stee
Number of solvents mfr Four solvents on the Solvent Select Valve (SSV)
Solvent conditioning mfr Sparging one channel per solvent
Operating flow rate range mfr 0.50 to 150.00 mL/min, in 0.01 mL/min increments
Maximum operating pressure mfr 6000 psi up to 100 mL/min, roll off to 5000 psi
Plunger seal wash mfr Integral, active, programmable
Delay volume mfr <6.5 mL
Compressibility compensation mfr Automatic and continuous
Priming mfr Self prime from solvent container on floor with
Compositional accuracy mfr +/-3% absolute (full scale) from 5% to 95%; and
Compositional precision mfr Retention time variation <0.15 min SD (uracil/ca
Default syringe mfr 5 mL
Default loop size mfr 2 mL (Optional; 5 mL, and 10 mL)
Dimensions, Weight & Mechanical Design
Height 12.3 inch
Weight 84.0 lbs
Width 14.3 inch
Operating Conditions, Environment & Protection
Ambient operating temperature mfr 15 to 40 °C
Ambient operating humidity mfr 20 to 80% RH at 40 °C (104 °F)

Description

Construction and Positioning

Integration & Operating Notes

Process Limits
Do not exceed 6000 psi at flows above 100 mL/min

The maximum operating pressure is 6000 psi up to 100 mL/min, but rolls off to 5000 psi at 150 mL/min. Exceeding these limits can cause overpressure, potentially damaging seals or the column. Always verify backpressure at the target flow rate, especially when using small-particle columns or high-viscosity modifiers.

Factory Specification Sheet
Process Limits
Recovery >90% requires operation within total flow range

The system achieves >90% recovery only when the total flow rate (combined CO2 and co-solvent) is within the specified range of 20 to 150 mL/min. Operating outside this range may reduce recovery and compromise purification yield. Ensure method flow rates are set accordingly.

Factory Specification Sheet
Maintenance
Enable programmable plunger seal wash to extend seal life

The integral, active, programmable plunger seal wash must be configured and maintained to prevent seal wear and leaks. Regularly check wash solvent levels and program wash cycles according to the instruction manual. Neglecting this can lead to premature seal failure and downtime.

Instruction Manual, Section on Maintenance

Selection & Integration Q&A

When should I choose the SFC Prep 150 MGM System over other SFC Prep 150 series variants?

Choose the MGM variant when mass-guided fraction collection is needed for automated purification of unknown or complex samples. For simpler, UV-based purification, a non-MGM sibling may suffice. The MGM adds a mass spectrometer for peak-triggered collection, enhancing recovery and purity.

What are the requirements for connecting the SFC Prep 150 MGM to a mass spectrometer and fraction collector?

The system includes an integrated mass spectrometer for MGM operation. Ensure the MS is compatible with SFC flow rates (up to 150 mL/min) and pressures. The fraction collector must handle the outlet flow and be triggered by the MS signal. Refer to the manual for specific interface connections.

What maintenance items and accessories are not included with the base system?

The base system includes a 5 mL syringe and 2 mL loop; optional 5 mL and 10 mL loops are available separately. Consumables such as columns, solvents, and replacement seals are not included. Regular maintenance includes seal wash solvent replacement and pump seal inspection per the manual.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Prep SFC 150 MultiGram System Site Preparation Guide
Manual ·474,6 КБ ·27 pp.
View PDF