Waters SYNAPT XS
Waters ·Made in USA · In Active Production

Waters SYNAPT XS

Waters SYNAPT XS — Mass spectrometers.

The Waters SYNAPT XS is a high-resolution mass spectrometer platform combining ion mobility spectrometry with time-of-flight mass analysis. It is positioned for advanced analytical laboratories requiring multidimensional separation and accurate mass measurement. The system serves as a primary detector in LC-MS workflows for complex mixture characterization.

Direct drop-in replacements

Shimadzu
Axima Assurance
0%
match
⚖️ Compare
Shimadzu
Axima Confidence
0%
match
⚖️ Compare
0%
match
⚖️ Compare
Shimadzu
GC-MS Qp2020 NX
0%
match
⚖️ Compare

Full specifications

Description

Construction and Architecture

The SYNAPT XS is a benchtop mass spectrometer integrating ion mobility separation (IMS) with quadrupole time-of-flight (QTOF) mass analysis. The architecture typically includes an electrospray ionization source, a quadrupole mass filter, a traveling-wave ion mobility cell, and a reflectron TOF analyzer. This configuration enables acquisition of both mass-to-charge (m/z) and ion mobility (drift time) data, providing an additional dimension of separation for isobaric and isomeric species. ### Behavior Under Load and Modes

The instrument supports multiple acquisition modes, including MS, MS/MS, and ion mobility-enhanced MS. In IMS mode, ions are separated based on their collision cross section (CCS) as they travel through a gas-filled mobility cell under the influence of a traveling wave. This allows for the resolution of species that would otherwise overlap in m/z alone. The TOF analyzer provides high mass resolution and mass accuracy, typically in the low ppm range, which is essential for elemental composition assignment. ### Installation, Connection, and Control

The SYNAPT XS requires a stable laboratory environment with controlled temperature and humidity, and a supply of high-purity nitrogen for the source and mobility cell. It is controlled via Waters' MassLynx software, which manages acquisition, calibration, and data processing. The system interfaces with liquid chromatography systems, typically via an electrospray source, and requires a compatible LC gradient system for sample introduction. Electrical requirements include a dedicated power supply and proper grounding. The instrument should be installed on a vibration-isolated bench to prevent spectral artifacts.

Integration & Operating Notes

Mounting
Install on a vibration-isolated bench to avoid spectral artifacts

The TOF analyzer is sensitive to mechanical vibrations, which can degrade mass accuracy and resolution. The instrument must be placed on a rigid, level surface free from nearby sources of vibration such as pumps or centrifuges. Adequate clearance for ventilation and service access is also required.

Waters SYNAPT XS Site Preparation Guide
Electrical
Requires a dedicated, stable power supply with proper grounding

The system draws significant current and is sensitive to voltage fluctuations. A dedicated circuit with a clean ground is necessary to prevent electrical noise from affecting detector performance. Uninterruptible power supply (UPS) is recommended to protect against power interruptions.

Waters SYNAPT XS Site Preparation Guide
Process Limits
Nitrogen gas supply must be high-purity and at stable pressure

The source and ion mobility cell require high-purity nitrogen (≥99.999%) at a regulated pressure. Fluctuations or impurities can cause signal instability, reduced sensitivity, or contamination of the ion optics. A gas generator or Dewar supply with appropriate regulators and filters is essential.

Waters SYNAPT XS Operator's Guide

Selection & Integration Q&A

When should I choose the SYNAPT XS over other Waters QTOF models?

Choose the SYNAPT XS when ion mobility separation is needed for resolving isobars/isomers or measuring collision cross sections. For routine accurate mass without IMS, a standard QTOF like the Xevo G2-XS may suffice.

What are the requirements for connecting the SYNAPT XS to an LC system?

The SYNAPT XS connects to an LC via an electrospray source. It requires a compatible LC system with flow rates typically in the 0.1–1.0 mL/min range, and the LC must be controlled by MassLynx for synchronized acquisition.

What environmental conditions are required for the SYNAPT XS?

The instrument requires an air-conditioned laboratory with temperature stability (±2°C) and humidity control (40–60% RH). Avoid direct sunlight and drafts. A vibration-isolated bench is mandatory for optimal performance.

What maintenance and consumables are needed for the SYNAPT XS?

Regular maintenance includes source cleaning, ion optics cleaning, and replacement of the nitrogen filters. Consumables include ESI emitters, sample vials, and calibration solutions. A service contract is recommended for annual preventive maintenance.

Dimensional Drawings & Technical Documentation

Dimensional drawings

SYNAPT XS — drawing 1

Click drawing to inspect vector dimensions

Technical documentation