Table 2: Capillary connections for 1260 Infinity III systems
p/n
From
To
G7120-60007 (Bottle Head Assembly, long (1730 mm))
Solvent Bottle
Infinity III Pump
5500-1246 (Capillary ST 0.17 mm x 500 mm SI/SI)
Pump
Sampler
5500-1217 (Capillary, ST, 0.17 mm x 900 mm SI/SX)
Pump
Vialsampler with ICC
5500-1246 (Capillary ST 0.17 mm x 500 mm SI/SI)
Multisampler
MCT Valve/Heat Exchanger
5500-1252 (Capillary, ST, 0.17 mm x 400 mm SL/SL)
Vialsampler
MCT Valve/Heat Exchanger
5500-1240 (Capillary ST 0.17 mm x 105 mm SL/SL)
Vialsampler
ICC Heat Exchanger
5500-1250 (Capillary, ST, 0.17 mm x 120 mm SL/SL, long socket)
ICC Heat Exchanger
Column
5500-1193 (InfinityLab Quick Turn Capillary ST 0.17 mm x 105 mm, long socket)
MCT Heat Exchanger
Column
5500-1191 (InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket)
Column/MCT Valve
Detector
5062-8535 (Waste accessory kit (Flow Cell to waste))
VWD
Waste
5062-2462 (Tube PTFE 0.7 mm x 5 m, 1.6 mm od)
DAD/FLD
Waste
G5664-68712 (Analytical tubing kit 0.25 mm i.d. PTFE-ESD)
Detector
Fraction Collector
Capillary and Tubing Connections in Flow Path
p/n
From
To
5500-1596 (Quick Turn Capillary MP35N 0.12 mm x 280 mm)
Column/MCT Valve
Detector (DAD or 1290 FLD)
5500-1598 (Quick Turn Capillary MP35N 0.12 mm x 500 mm)
Column/MCT Valve
Detector (VWD)
5062-8535 (Waste accessory kit (Flow Cell to waste))
VWD
Waste
5062-2462 (Tube PTFE 0.7 mm x 5 m, 1.6 mm od)
DAD or 1290 FLD
Waste
G5664-68712 (Analytical tubing kit 0.25 mm i.d. PTFE-ESD)
Detector
Fraction Collector
Figure 38: 1290 Infinity II Bio Online Sample Manager Accessory Kit
#
Qty.
p/n
Description
1
G3167-68010
1290 Infinity II Bio Online Sample Manager Accessory Kit
1
2
G3167-42000
Single Holder UVD Multi Function
2
2
5043-1356
Column Holder Lamella
3
1
2110-1486
Fuse 2 AT250 V
4
1
5043-0270
Leak plane
5
1
5043-0271
Holder leak plane
6
1
5067-4792
Leak sensor assembly
7
1
5063-6527
Tubing, Silicon Rubber, 1.2 m, ID/OD 6 mm/9 mm
8
1
5500-1156
T-Tube Connector ID6.4
9
1
5320-0003
3-position/6-port FI valve, MP35N, 1300 bar
10
1
5005-0072
Capillary MP35N 0.12 mm x 160 mm SL/S
11
1
5005-0069
Capillary MP35N 0.17 mm x 160 mm SL/SL
12
3
G4220-60007
Bottle Head Assembly
1
8121-3099
CAN Cable, 1 m, flat (not shown)
Figure 42: G3167BA Online Sample Manager Set PM Kit
#
Qty.
p/n
Description
1
G3167-67003
G3167BA Online Sample Manager Set PM Kit
1
1
G7137-87201
Needle Bio-compatible
2
1
G3167-60017
Needle Seat Bio Capillary ST 0.17 x 230 SL/SL (5067-5403 (UHP fitting) is shown as pre-installed but included as a separate part)
3
2
5320-0005
Rotor Seal, 3-position/6-port FI Valve, MP35N, 1300 bar
4
1
G5668-60494
Seal 500 µL Bio-inert
Parts and Consumables — Infinity Lab Quick Connect and Quick Turn Fittings
p/n
Description
5500-1191
InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket
5500-1192
Capillary ST 0.12 mm x 500 mm, long socket
5500-1193
InfinityLab Quick Turn Capillary ST 0.17 mm x 105 mm, long socket
5500-1194
Capillary ST 0.17 mm x 150 mm, long socket
5500-1234
Capillary ST 0.17 mm x 180 mm
5500-1195
Capillary ST 0.17 mm x 200 mm, long socket
5500-1196
Capillary ST 0.17 mm x 280 mm, long socket
5500-1235
Capillary ST 0.17 mm x 380 mm, long socket
5500-1236
Capillary ST 0.17 mm x 400 mm, long socket
5500-1197
Capillary ST 0.17 mm x 500 mm, long socket
5500-1237
Capillary 0.17 mm x 700 mm, ns/ns
5500-1262
Capillary 0.25 mm x 150 mm, ns/ns
5500-1263
Capillary ST 0.25 mm x 400 mm, long socket
5500-1200
Quick Turn Capillary ST 0.12 mm x 130 mm SL/M
5500-1288
Capillary ST 0.12 mm x 150 mm, long socket, M4
5500-1290
Capillary ST 0.17 mm x 150 mm, long socket, M4
Engineering Facts
14 facts · every claim linked to a page
Selection
The system is intended for automated process monitoring with real-time data, so it suits process analytics rather than routine offline QC.
Confirms the intended application before selecting this platform for a process monitoring task.
Source
“The 1260 Infinity III Prime Online LC System is a process analysis system enabling automated process monitoring, providing real-time data for greater control and faster understanding of processes.”
Selection
The Online Sample Manager supports automated dilutions up to 1:1000, which defines the maximum sample dilution capability.
Determines whether the module can handle concentrated process samples without manual prep.
Source
“The Online Sample Manager provides automated dilutions of up to 1:1000, retain-sample functionality, and direct analysis of the process samples, as well as automated sample preparation.”
Selection
The DAD WR covers 190–950 nm with a 120 Hz acquisition rate, so it fits fast LC and broad UV-Vis detection needs.
Guides detector choice when fast acquisition or wide wavelength range is required.
Source
“With its 120 Hz data acquisition rate the detector is perfectly suited for fast LC applications.”
Selection
The MCT thermostats columns from 20 °C below ambient up to 110 °C, defining the usable column temperature window.
Needed to verify the column oven can reach the required method temperatures.
Source
“facilitates precise column thermostatting over a broad temperature range with cooling down to 20 °C below ambient temperature and heating up to 110 °C.”
Installation
Stack configuration should be chosen to minimize delay volume, so the flow path layout is a performance decision.
Affects how modules are physically arranged to keep delay volume low.
Source
“These configurations optimize the system flow path, ensuring minimum delay volume.”
Installation
A single stack is limited to a maximum of four modules, which constrains how many modules can be stacked together.
Determines whether the planned module count fits a single-stack layout.
Source
“Single Stack (maximal 4 modules, in a bench rack or directly on the bench)”
Installation
Two stacks are recommended when stack height becomes excessive, for example with an additional detector.
Informs the physical layout decision when adding detectors.
Source
“To avoid excessive height of the stack (for example when using the system in combination with an additional detector) it is recommended to form two stacks.”
Safety
The solvent cabinet holds a maximum of 8 L total and no individual bottle may exceed 2 L.
Sets capacity limits for solvent storage during installation.
Source
“The solvent cabinet is designed to store a maximum volume of 8 L solvent. The maximum volume for an individual bottle stored in the solvent cabinet should not exceed 2 L.”
Safety
The leak sensor stops the running system once the leak detection level is reached, so leak handling is automatic.
Explains the automatic shutdown behavior relied on for safe operation.
Source
“The leak sensor in the leak plane stops the running system as soon as the leak detection level is reached.”
Safety
DMF must not be used as mobile phase because it corrodes the PVDF leak sensor.
Prevents solvent selection that would damage the leak sensor.
Source
“Do not use DMF as mobile phase.”
Installation
Waste drainage must run straight into the waste container without restriction at bends or joints.
Guides waste tubing routing to avoid blockages and backflow.
Source
“The waste drainage must go straight into the waste containers. The waste flow must not be restricted at bends or joints.”
Operation
The ERI interface timeout defines how long the software waits for a sample delivery device answer before marking the sample missed.
Needed to configure reliable communication with the sample delivery hardware.
Source
“The Timeout [min] value defines the time, how long the software waits for an answer from the sample delivery device.”
Safety
Solvents with auto-ignition temperature below 200 °C or boiling point below 56 °C must not be used.
Constrains mobile phase selection for safe operation.
Source
“Do not use solvents with an auto-ignition temperature below 200 °C (392 °F). Do not use solvents with a boiling point below 56 °C (133 °F).”
Safety
Bottles exceeding the maximum permissible volume of 2.5 L must not be used.
Sets a bottle size limit for safe solvent handling.
Source
“Do not use bottles that exceed the maximum permissible volume (2.5 L).”