Acquisition Mode MS1 Min Range (m/z) 50 , Max Range (m/z) 500 , Scan Rate (spectra/sec) 3
Instrument Parameters
Source Parameters
Gas Temp
250 °C
Gas Flow
11 L/min
Nebulizer
40 psig
SheathGasTemp
350 °C
SheathGasFlow
12 L/min
Scan Segment
1
Scan Source Parameters
Vcap
3500 V
Nozzle Voltage
300 V
Fragmentor
120
Skimmer1
45
OctopoleRFPeak
750
Engineering Facts
14 facts · every claim linked to a page
Selection
The 2D-LC valve must be hosted in an external valve drive; this constrains the instrument setup.
Determines whether the proposed stack can be used at all; the valve cannot be hosted elsewhere.
Source
“The 2D-LC solution only supports the instrument setup where the 2D-LC valve is hosted in the external valve drive”
Selection
A 1290 Infinity Binary Pump or 1290 Infinity II/III High-Speed Pump is mandatory for the second dimension.
Constrains pump selection for the 2D dimension and prevents using an unsuitable pump.
Source
“In the second dimension, a 1290 Infinity Binary Pump or 1290 Infinity II/III High-Speed Pump is mandatory.”
Selection
MHC and HiRes configurations require additional MHC decks and support both the standard G4236A valve and the ASM G4243A valve head.
Drives hardware bill of materials and stack layout for MHC/HiRes installations.
Source
“2D-LC instruments that are used for Multiple Heart-Cutting or High-Resolution Sampling 2D-LC require additional MHC decks.”
Installation
Different transfer capillaries (6a/6b) must be installed depending on which valve head is used.
Prevents wrong capillary selection and resulting dispersion or connection errors during installation.
Source
“Depending on the valve head that is used, different transfer capillaries (6a/6b) must be installed.”
Installation
A USB hardware dongle is required to activate the 2D-LC solution in addition to software licenses.
Must be ordered and available before commissioning; otherwise the 2D-LC driver cannot be activated.
Source
“Additionally you need a USB hardware dongle to activate the 2D-LC solution.”
Operation
In standard heart-cutting, a second-dimension analysis always finishes, so a later 1D peak may be lost if sampled while 2D is busy.
Explains risk of peak loss and motivates MHC/HiRes or careful timing when using single heart-cutting.
Source
“A started second dimension analysis will always be finished!”
Operation
If all sample loops in the first dimension are occupied, the peak is lost in Multiple Heart-Cutting.
Sets capacity limit of the loop deck and informs method timing and loop count planning.
Source
“If all sample loops in the first dimension are occupied, the peak is lost.”
Operation
Stored cuts are analyzed in reverse order to avoid contamination, and a flush gradient runs before a new parking deck.
Affects sequence logic and carry-over control; engineers must not reorder cuts manually.
Source
“Stored cuts are analysed in backwards order to avoid contamination.”
Operation
In LCxLC the 2D gradient analysis time must be less than or equal to the 1D cut size time.
Defines the modulation constraint; violating it causes incomplete 2D separation and loop conflicts.
Source
“The gradient analysis in the second dimension is less than or equal to the cut size time in the first dimension:”
Performance
ASM dilutes the sample loop content with weak 2D solvent before the 2D column to improve second-dimension separation.
Justifies selecting the ASM valve when 1D solvent strength would distort 2D peaks.
Source
“Active Solvent Modulation (ASM) dilutes the content of the sampling loop (sample and 1D solvent) with weak 2D solvent before it reaches the 2D column”
Performance
The ASM valve has four switching positions; positions 1 and 3 match the standard G4236A valve and positions 2 and 4 add dilution.
Needed to understand valve timing and method programming differences versus the standard valve.
Source
“A full switching cycle of the ASM valve has 4 positions.”
Installation
For a standard 2D-LC with two DADs, the transfer volume is approximately 14 µL.
Sets the expected delay between 1D detection and valve switching; useful for verifying configuration.
Source
“For a standard 2D-LC with two DADs, the transfer volume is approx. 14 µL.”
Operation
If the 2D pump stop time is insufficient to analyze all 2D cuts, an alert appears and the stop time must be extended.
Prevents truncated 2D analysis and lost cuts due to misaligned stop time.
Source
“Most likely the stop time of the 2D pump is not sufficient to analyze all 2D cuts”
Operation
In threshold mode the 2D-LC valve switches when the UV signal rises above the threshold and switches again when it falls below or the cut size elapses.
Defines cut boundaries for peak-triggered methods and helps set threshold and cut size correctly.
Source
“The 2D-LC Valve will switch to the next position when the UV signal falls below this value or the cut size elapse.”