Contents

  1. Overview pp. 13-18
    Overview of MSD hardware, safety, and intended use
  2. Warning pp. 22-33
    ESD precautions, hazards, certifications, and intended use
  3. Procedure pp. 34-102
    Installing columns on 8890, Intuvo 9000, and 7890 GCs
  4. Procedure pp. 103-144
    Operating the MSD in EI mode from the data system
  5. Procedure pp. 145-167
    CI autotune, reagent gas setup, and performance verification
  6. Overview pp. 168-177
    Spare parts, tools, and supplies for maintenance
  7. Procedure pp. 178-222
    Removing, cleaning, assembling, and installing EI sources
  8. Procedure pp. 223-250
    Switching, disassembling, cleaning, and assembling CI sources
  9. p. 149 table, 6 rows
  10. p. 36 table, 8 rows
  11. p. 56 table, 5 rows
  12. p. 78 table, 9 rows
  13. p. 199 table, 8 rows
  14. p. 200 table, 11 rows
  15. p. 210 table, 5 rows
  16. p. 211 table, 14 rows
  17. p. 18 table, 9 rows
  18. p. 156 table, 6 rows

Table 13 Default Tune Control Limits, used by CI autotune only

Reagent gasMethaneMethaneIsobutaneIsobutaneAmmoniaAmmonia
Ion polarityPositiveNegativePositiveNegativePositiveNegative
Abundance target1x1061x106N/A1x106N/A1x106
Peakwidth target0.60.6N/A0.6N/A0.6
Maximum repeller44N/A4N/A4
Maximum emission current, µA24050N/A50N/A50
Max electron energy, eV240240N/A240N/A240

N/A: Not available. There are no PFDTD ions formed in PCI with any reagent gas but methane, hence, CI autotune is not available with these configurations.

Table 5 Gas flows

FeatureGas flowsGas flows
High vacuum pumpDiffusionTurbo
Optimal He column flow mL/min11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.53 mm (30 m)0.53 mm (30 m)
CI capabilityNoYes
CI reagent flows, mL/minNA1 to 2
JetClean option H2 flowNA0.4 mL/min

*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity.

Table 6 Gas flows

FeatureHigh vacuum pump: DiffusionHigh vacuum pump: Turbo
Optimal He column flow mL/min11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.53 mm (30 m)0.53 mm (30 m)
JetClean option H2 flowNA0.4 mL/min

* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity.

Table 7 Gas flows

FeatureGas flowsGas flows
High vacuum pump
DiffusionTurbo
Optimal He column flow mL/min 11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.53 mm (30 m)0.53 mm (30 m)
CI capabilityNoYes
CI reagent flows, mL/minNA1 to 2
JetClean option H2 flowNA0.4 mL/min

* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity.

Table 20 Parts list for the EI SS or Inert source (Figure 64)

Item numberItem descriptionPart number (SS)Part number (Inert)
1Gold plated set screwG1999-20022G1999-20022
2Gold plated screwG3870-20021G3870-20021
3Interface socketG1099-20136G1099-20136
4Source bodyG1099-20130G2589-20043
5Drawout cylinderG1072-20008G1072-20008
6Drawout plate05971-20134G2589-20100
74-turn filamentG7005-60061G7005-60061
8Spring washer3050-13743050-1374

Table 20 Parts list for the EI SS or Inert source (Figure 64) (continued)

Item numberItem descriptionPart number (SS)Part number (Inert)
8Flat washer3050-09823050-0982
9Lens insulatorG3170-20530G3170-20530
10Entrance lensG3170-20126G3170-20126
11Ion focus lens05971-2014305971-20143
12Repeller insulatorG1099-20133G1099-20133
13RepellerG3870-60172G3870-60173
14Flat washer3050-06273050-0627
15Belleville spring washer3050-13013050-1301
16Repeller nut0535-00710535-0071
17Source heater block assemblyG3870-60180G3870-60179
18Repeller block insertG3870-20135G3870-20135

Table 22 Parts list for the EI standard or EI Inert source (Figure 67)

Item numberItem descriptionPart number (SSL)Part number (Inert)
1Gold plated set screwG1999-20022G1999-20022
2Gold plated screwG3870-20021G3870-20021
3Interface socketG1099-20136G1099-20136
4Source bodyG1099-20130G2589-20043
5Drawout cylinderG1072-20008G1072-20008

Table 22 Parts list for the EI standard or EI Inert source (Figure 67) (continued)

Item numberItem descriptionPart number (SSL)Part number (Inert)
6Drawout plate05971-20134G2589-20100
74-turn filamentG7005-60061G7005-60061
8Spring washer3050-13743050-1374
8Flat washer3050-09823050-0982
9Lens insulatorG3170-20530G3170-20530
10Entrance lensG3170-20126G3170-20126
11Ion focus lens05971-2014305971-20143
12Repeller insulatorG1099-20133G1099-20133
13RepellerG3870-60172G3870-60173
14Flat washer3050-06273050-0627
15Belleville spring washer3050-13013050-1301
16Repeller nut0535-00710535-0071
17Source heater block assemblyG3870-60180G3870-60179
18Repeller block insertG3870-20135G3870-20135

Table 3 5977B Series MSD features

FeatureHigh vacuum pump: DiffusionHigh vacuum pump: Turbo
Optimal He column flow mL/min11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.25 mm (30 m)0.53 mm (30 m)
CI capability‡NoYes
Inert ion sources availableYesYes
GC compatibility9000/8890/8860/7890 Series9000/8890/8860/7890 Series
Foreline pumps availablePfeiffer DuoPfeiffer Duo, Pfeiffer MVP-070-3x, Agilent IDP-3 24V
DIP** capability (3rd party)YesYes

*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity. ‡Turbo pump models are field upgradeable to CI. **Direct insertion probe.

Table 15 Reagent gases

Reagent gas/modeReagent ion massesPFDTD Calibrant ionsFlow adj ions: Ratio EI/PCI/NCI MSD Turbo pump Recommended flow: 20% PCI 40% NCI
Methane/PCI17, 29, 41*41, 267, 59928/27: 1.5 – 5.0
Methane/NCI17, 35, 235†185, 351, 449N/A
Isobutane/PCI39, 43, 57N/A57/43: 5.0 – 30.0
Isobutane/NCI17, 35, 235185, 351, 449N/A
Ammonia/PCI18, 35, 52N/A35/18: 0.1 – 1.0
Ammonia/NCI17, 35, 235185, 351, 517N/A

* There are no PFDTD ions formed with any reagent gas but methane. Tune with methane and use the same parameters for the other gas. † There are no negative reagent gas ions formed. To pretune in negative mode, use background ions: 17 (OH-), 35 (Cl-), and 235 (ReO3-). These ions cannot be used for reagent gas flow adjustment. Set the flow to 40% for NCI and adjust as necessary to get acceptable results for your application.

Engineering Facts

14 facts · every claim linked to a page
Selection

Choose the high-vacuum pump type based on required column flow and CI capability, since diffusion and turbo models have different limits.

Pump choice constrains column flow, column id, and whether CI can be used at all.

Source
“Optimal He column flow mL/min”
Selection

Diffusion-pump models cannot be upgraded to CI, while turbo models can be field upgraded.

Determines whether the instrument can later support chemical ionization.

Source
“Turbo pump models are field upgradeable to CI.”
Selection

The Intuvo 9000 GC cannot be used with CI on this MSD.

Rules out CI workflows if the Intuvo GC is part of the configuration.

Source
“If you are using the Agilent Intuvo 9000 Gas Chromatograph with your MSD, Chemical Ionization (CI) is not currently supported.”
Installation

Never condition a capillary column with hydrogen; use inert gas first to avoid oven explosion risk.

Prevents an explosion hazard during column conditioning before MSD connection.

Source
“Do not condition your capillary column with hydrogen.”
Installation

The column insertion depth into the interface depends on source type: about 1 mm for XTR/SS/Inert/CI and 4 to 5 mm for EI HES.

Wrong insertion depth degrades sensitivity and can damage the source.

Source
“For an EI HES Installation (See Figure 22 on page 86.), the column extends 4 to 5 mm.”
Installation

Do not install the interface tip seal, spring, and retainer with an EI SS or Inert source because the socket will not fit.

Prevents improper assembly and sealing problems on SS/Inert sources.

Source
“Do not use the interface tip seal, spring, and knurled seal tip retainer with an EI SS or Inert source.”
Installation

If the analyzer will not close freely, do not force it; misalignment will damage the seal, interface, or ion source.

Avoids costly damage during interface tip seal installation and alignment.

Source
“Forcing the analyzer closed if these parts are misaligned will damage the seal or the interface or the ion source, or keep the sideplate from sealing.”
Installation

On Intuvo, tighten the compression bolt only until the torque driver clicks once; over-tightening damages the flow path and causes leaks.

Correct torque is required to avoid flow-path damage and leaks.

Source
“Over-tightening can damage the flow path, strip the fittings, and cause leaks.”
Operation

After pumpdown, wait 2 hours for thermal equilibrium before acquiring data to ensure reproducibility.

Data taken before equilibrium may not be reproducible.

Source
“wait 2 hours for the MSD to reach thermal equilibrium.”
Operation

Do not heat any GC zone until carrier gas flow is on, or the column will be damaged.

Prevents column damage from heating without carrier gas.

Source
“Do not turn on any GC heated zones until carrier gas flow is on.”
Operation

After switching from EI to CI or venting, bake out the MSD at least 2 hours before tuning.

Avoids poor tune results and contamination after mode changes.

Source
“the MSD must be baked out for at least 2 hours before tuning.”
Operation

Do not run CI autotune with an air leak or high water, as it will severely contaminate the ion source.

Prevents source contamination that requires venting and cleaning.

Source
“Continuing with CI autotune if the MSD has an air leak or large amounts of water will result in severe ion source contamination.”
Maintenance

CI operation requires more frequent ion source cleaning because higher source pressures accelerate contamination.

Plans maintenance intervals and downtime for CI workflows.

Source
“In CI operation, the ion source chamber is subject to more rapid contamination than in EI operation because of the higher source pressures required for CI.”
Safety

The standard foreline pump oil trap does not filter toxic chemicals and must be removed for toxic solvents or CI MSDs.

Requires proper exhaust routing to protect operators from toxic vapors.

Source
“The oil trap supplied with the standard foreline pump stops only foreline pump oil. It does not trap or filter out toxic chemicals.”