Overview of MSD hardware, safety, and intended use
Warningpp. 22-33
ESD precautions, hazards, certifications, and intended use
Procedurepp. 34-102
Installing columns on 8890, Intuvo 9000, and 7890 GCs
Procedurepp. 103-144
Operating the MSD in EI mode from the data system
Procedurepp. 145-167
CI autotune, reagent gas setup, and performance verification
Overviewpp. 168-177
Spare parts, tools, and supplies for maintenance
Procedurepp. 178-222
Removing, cleaning, assembling, and installing EI sources
Procedurepp. 223-250
Switching, disassembling, cleaning, and assembling CI sources
p. 149table, 6 rows
p. 36table, 8 rows
p. 56table, 5 rows
p. 78table, 9 rows
p. 199table, 8 rows
p. 200table, 11 rows
p. 210table, 5 rows
p. 211table, 14 rows
p. 18table, 9 rows
p. 156table, 6 rows
Table 13 Default Tune Control Limits, used by CI autotune only
Reagent gas
Methane
Methane
Isobutane
Isobutane
Ammonia
Ammonia
Ion polarity
Positive
Negative
Positive
Negative
Positive
Negative
Abundance target
1x106
1x106
N/A
1x106
N/A
1x106
Peakwidth target
0.6
0.6
N/A
0.6
N/A
0.6
Maximum repeller
4
4
N/A
4
N/A
4
Maximum emission current, µA
240
50
N/A
50
N/A
50
Max electron energy, eV
240
240
N/A
240
N/A
240
N/A: Not available. There are no PFDTD ions formed in PCI with any reagent gas but methane, hence, CI autotune is not available with these configurations.
Table 5 Gas flows
Feature
Gas flows
Gas flows
High vacuum pump
Diffusion
Turbo
Optimal He column flow mL/min
1
1 to 2
Maximum recommended gas flow mL/min*
1.5
4
Maximum gas flow, mL/min†
2
6.5
Max column id
0.53 mm (30 m)
0.53 mm (30 m)
CI capability
No
Yes
CI reagent flows, mL/min
NA
1 to 2
JetClean option H2 flow
NA
0.4 mL/min
*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity.
Table 6 Gas flows
Feature
High vacuum pump: Diffusion
High vacuum pump: Turbo
Optimal He column flow mL/min
1
1 to 2
Maximum recommended gas flow mL/min*
1.5
4
Maximum gas flow, mL/min†
2
6.5
Max column id
0.53 mm (30 m)
0.53 mm (30 m)
JetClean option H2 flow
NA
0.4 mL/min
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity.
Table 7 Gas flows
Feature
Gas flows
Gas flows
High vacuum pump
Diffusion
Turbo
Optimal He column flow mL/min 1
1 to 2
Maximum recommended gas flow mL/min*
1.5
4
Maximum gas flow, mL/min†
2
6.5
Max column id
0.53 mm (30 m)
0.53 mm (30 m)
CI capability
No
Yes
CI reagent flows, mL/min
NA
1 to 2
JetClean option H2 flow
NA
0.4 mL/min
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity.
Table 20 Parts list for the EI SS or Inert source (Figure 64)
Item number
Item description
Part number (SS)
Part number (Inert)
1
Gold plated set screw
G1999-20022
G1999-20022
2
Gold plated screw
G3870-20021
G3870-20021
3
Interface socket
G1099-20136
G1099-20136
4
Source body
G1099-20130
G2589-20043
5
Drawout cylinder
G1072-20008
G1072-20008
6
Drawout plate
05971-20134
G2589-20100
7
4-turn filament
G7005-60061
G7005-60061
8
Spring washer
3050-1374
3050-1374
Table 20 Parts list for the EI SS or Inert source (Figure 64) (continued)
Item number
Item description
Part number (SS)
Part number (Inert)
8
Flat washer
3050-0982
3050-0982
9
Lens insulator
G3170-20530
G3170-20530
10
Entrance lens
G3170-20126
G3170-20126
11
Ion focus lens
05971-20143
05971-20143
12
Repeller insulator
G1099-20133
G1099-20133
13
Repeller
G3870-60172
G3870-60173
14
Flat washer
3050-0627
3050-0627
15
Belleville spring washer
3050-1301
3050-1301
16
Repeller nut
0535-0071
0535-0071
17
Source heater block assembly
G3870-60180
G3870-60179
18
Repeller block insert
G3870-20135
G3870-20135
Table 22 Parts list for the EI standard or EI Inert source (Figure 67)
Item number
Item description
Part number (SSL)
Part number (Inert)
1
Gold plated set screw
G1999-20022
G1999-20022
2
Gold plated screw
G3870-20021
G3870-20021
3
Interface socket
G1099-20136
G1099-20136
4
Source body
G1099-20130
G2589-20043
5
Drawout cylinder
G1072-20008
G1072-20008
Table 22 Parts list for the EI standard or EI Inert source (Figure 67) (continued)
*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity. ‡Turbo pump models are field upgradeable to CI. **Direct insertion probe.
* There are no PFDTD ions formed with any reagent gas but methane. Tune with methane and use the same parameters for the other gas. † There are no negative reagent gas ions formed. To pretune in negative mode, use background ions: 17 (OH-), 35 (Cl-), and 235 (ReO3-). These ions cannot be used for reagent gas flow adjustment. Set the flow to 40% for NCI and adjust as necessary to get acceptable results for your application.
Engineering Facts
14 facts · every claim linked to a page
Selection
Choose the high-vacuum pump type based on required column flow and CI capability, since diffusion and turbo models have different limits.
Pump choice constrains column flow, column id, and whether CI can be used at all.
Source
“Optimal He column flow mL/min”
Selection
Diffusion-pump models cannot be upgraded to CI, while turbo models can be field upgraded.
Determines whether the instrument can later support chemical ionization.
Source
“Turbo pump models are field upgradeable to CI.”
Selection
The Intuvo 9000 GC cannot be used with CI on this MSD.
Rules out CI workflows if the Intuvo GC is part of the configuration.
Source
“If you are using the Agilent Intuvo 9000 Gas Chromatograph with your MSD, Chemical Ionization (CI) is not currently supported.”
Installation
Never condition a capillary column with hydrogen; use inert gas first to avoid oven explosion risk.
Prevents an explosion hazard during column conditioning before MSD connection.
Source
“Do not condition your capillary column with hydrogen.”
Installation
The column insertion depth into the interface depends on source type: about 1 mm for XTR/SS/Inert/CI and 4 to 5 mm for EI HES.
Wrong insertion depth degrades sensitivity and can damage the source.
Source
“For an EI HES Installation (See Figure 22 on page 86.), the column extends 4 to 5 mm.”
Installation
Do not install the interface tip seal, spring, and retainer with an EI SS or Inert source because the socket will not fit.
Prevents improper assembly and sealing problems on SS/Inert sources.
Source
“Do not use the interface tip seal, spring, and knurled seal tip retainer with an EI SS or Inert source.”
Installation
If the analyzer will not close freely, do not force it; misalignment will damage the seal, interface, or ion source.
Avoids costly damage during interface tip seal installation and alignment.
Source
“Forcing the analyzer closed if these parts are misaligned will damage the seal or the interface or the ion source, or keep the sideplate from sealing.”
Installation
On Intuvo, tighten the compression bolt only until the torque driver clicks once; over-tightening damages the flow path and causes leaks.
Correct torque is required to avoid flow-path damage and leaks.
Source
“Over-tightening can damage the flow path, strip the fittings, and cause leaks.”
Operation
After pumpdown, wait 2 hours for thermal equilibrium before acquiring data to ensure reproducibility.
Data taken before equilibrium may not be reproducible.
Source
“wait 2 hours for the MSD to reach thermal equilibrium.”
Operation
Do not heat any GC zone until carrier gas flow is on, or the column will be damaged.
Prevents column damage from heating without carrier gas.
Source
“Do not turn on any GC heated zones until carrier gas flow is on.”
Operation
After switching from EI to CI or venting, bake out the MSD at least 2 hours before tuning.
Avoids poor tune results and contamination after mode changes.
Source
“the MSD must be baked out for at least 2 hours before tuning.”
Operation
Do not run CI autotune with an air leak or high water, as it will severely contaminate the ion source.
Prevents source contamination that requires venting and cleaning.
Source
“Continuing with CI autotune if the MSD has an air leak or large amounts of water will result in severe ion source contamination.”
Maintenance
CI operation requires more frequent ion source cleaning because higher source pressures accelerate contamination.
Plans maintenance intervals and downtime for CI workflows.
Source
“In CI operation, the ion source chamber is subject to more rapid contamination than in EI operation because of the higher source pressures required for CI.”
Safety
The standard foreline pump oil trap does not filter toxic chemicals and must be removed for toxic solvents or CI MSDs.
Requires proper exhaust routing to protect operators from toxic vapors.
Source
“The oil trap supplied with the standard foreline pump stops only foreline pump oil. It does not trap or filter out toxic chemicals.”