Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model KB 1060

Technical Reference Summary

Covered Models & Document Scope

Model KB 1060 cooling incubator, 1,070 L interior volume. Two standard variants: KB1060-230V (220...240 V 1~ 50/60 Hz, CEE 7/7 plug, Art.-No. 9020-0476) and KB1060UL-120V (120 V 1~ 50/60 Hz, NEMA 5-20 plug, Art.-No. 9020-0504).

Technical Data

Parameter Value Condition
Temperature range 0...70 °C p.2 ↗ KB1060-230V / KB1060UL-120V
Temperature uniformity at 37°C 0.2 ± K p.2 ↗ KB1060-230V / KB1060UL-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ KB1060-230V / KB1060UL-120V
Max. heat compensation at 37°C 640 W p.2 ↗ KB1060-230V / KB1060UL-120V
Recovery time after door was opened 3 min p.2 ↗ KB1060-230V / KB1060UL-120V
Rated voltage 220...240 V p.2 ↗ KB1060-230V
Rated voltage 120 V p.2 ↗ KB1060UL-120V
Power frequency 50/60 Hz p.2 ↗ KB1060-230V / KB1060UL-120V
Nominal power 1,5 kW p.2 ↗ KB1060-230V / KB1060UL-120V
Unit fuse 16,0 A p.2 ↗ KB1060-230V / KB1060UL-120V
Phase (nominal voltage) 1~ p.2 ↗ KB1060-230V / KB1060UL-120V
Interior volume 1,070 L p.2 ↗ KB1060-230V / KB1060UL-120V
Net weight of the unit (empty) 315 kg p.2 ↗ KB1060-230V / KB1060UL-120V
Load per rack 40 kg p.2 ↗ KB1060-230V / KB1060UL-120V
Permitted load 200 kg p.2 ↗ KB1060-230V / KB1060UL-120V
Wall clearance back 100 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Wall clearance sidewise 180 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Width net 1,285 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Height net 1,951 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Depth net 1,175 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Interior width 1,011 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Interior height 1,250 mm p.2 ↗ KB1060-230V / KB1060UL-120V
Interior depth 847 mm p.3 ↗ KB1060-230V / KB1060UL-120V
Sound-pressure level at 25°C 52 dB(A) p.3 ↗ KB1060-230V / KB1060UL-120V
Number of shelves (std./max.) 2/16 p.3 ↗ KB1060-230V / KB1060UL-120V

Operating Limits

  • Temperature range: 0...70 °C p.2 ↗
  • Max. heat compensation at 37°C: 640 W p.2 ↗
  • Permitted load: 200 kg p.2 ↗
  • Load per rack: 40 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 180 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ KB Product family: cooling incubator
3-6 p.1 ↗ 1060 Model/size: 1060 (1,070 L interior volume)
suffix p.1 ↗ -230V Mains variant: 220...240 V 1~ 50/60 Hz, CEE 7/7 plug
suffix p.1 ↗ UL-120V Mains variant: 120 V 1~ 50/60 Hz, NEMA 5-20 plug

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
KB1060-230V p.1 ↗ 9020-0476 220...240 V 1~ 50/60 Hz, CEE 7/7, Standard
KB1060UL-120V p.1 ↗ 9020-0504 120 V 1~ 50/60 Hz, NEMA 5-20, Standard

Related models