BINDER KB 1060
BINDER KB 1060 — Incubators with Temperature 0…70 °C.
The BINDER KB 1060 is a 1,070 L cooling incubator with a 0–70 °C working range, temperature uniformity of 0.2 K and fluctuation of 0.1 K. As the largest KB series model, it offers 640 W heat compensation and 200 kg permitted load, distinguishing it from smaller siblings. It is used for temperature-controlled storage and incubation of large sample volumes in laboratory processes.
Direct drop-in replacements
BINDER KB Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 1070 L |
| Load per rack | 40 kg |
| Max. heat compensation | 640 W |
| Nominal power | 1.5 kW |
| Permitted load | 200 kg (KB1060-230V / KB1060UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm (KB1060-230V / KB1060UL-120V) |
| Side wall clearance | 180 mm (KB1060-230V / KB1060UL-120V) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Depth | 1175 mm (KB1060-230V / KB1060UL-120V) |
| Height | 1951 mm (KB1060-230V / KB1060UL-120V) |
| Interior width | 1011 mm (KB1060-230V / KB1060UL-120V) |
| Width | 1285 mm (KB1060-230V / KB1060UL-120V) |
| Net weight of the unit (empty) mfr | 315 kg |
| Power frequency | 50 Hz |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and positioning
- The KB 1060 is a floor-standing cooling incubator with 1,070 L interior volume, net dimensions 1,285 mm wide × 1,951 mm high × 1,175 mm deep, and 315 kg empty weight.
- Within the KB series, it is the largest model; siblings KB 65, KB 130, KB 260 and KB 470 share the same 0–70 °C range but have smaller interiors.
- The unit is available in two voltage variants: KB1060-230V (220–240 V, 50/60 Hz, CEE 7/7 plug, Art. 9020-0476) and KB1060UL-120V (120 V, 50/60 Hz, NEMA 5-20 plug, Art. 9020-0504).
Performance and load behavior
- Temperature control is stable with 0.1 K fluctuation and 0.2 K uniformity across the interior.
- Maximum heat compensation is 640 W, allowing rapid recovery after door openings; recovery time after a 30 s door opening at 37 °C is 3 minutes.
- The permitted load is 200 kg total, with a maximum of 40 kg per rack, supporting heavy sample batches.
Installation and operation
- Required wall clearances are 100 mm at the rear and 180 mm at the sides to ensure adequate ventilation.
- The unit operates on a single phase (1~) at 50 Hz with a 16 A fuse; nominal power is 1.5 kW.
- No additional control or interface details are provided in the available documents; refer to the instruction manual for electrical connection and control specifics.
Integration & Operating Notes
Insufficient clearance restricts airflow and can cause overheating, leading to reduced performance or damage. Ensure the installation site allows these minimum distances from walls or other equipment.
The unit requires a 16 A fuse to handle the 1.5 kW nominal power at 220–240 V. Verify that the supply can deliver the required current without nuisance tripping.
Exceeding the per-rack load limit may cause rack deformation or failure, potentially damaging samples or the interior. Distribute weight evenly and consider additional racks for heavy loads.
Selection & Integration Q&A
Choose the KB 1060 when you need maximum capacity (1,070 L) and higher load capacity (200 kg total, 40 kg per rack). Smaller models are suitable for lower volume or when space is limited, but the KB 1060 offers the same temperature range and stability with greater throughput.
Install indoors with 100 mm rear and 180 mm side clearance for ventilation. The permissible ambient temperature range is not specified in the available documents; refer to the manual. Avoid unventilated recesses.
Annual electrical safety inspection is recommended. No accessories are included in base delivery; racks and shelves must be ordered separately. The LED plant light module is an optional accessory for plant growth applications.