Data Sheet Model MKF 1020
Technical Reference Summary
Covered Models & Document Scope
BINDER MKF 1020 dynamic climate chamber for rapid temperature changes with humidity control. Two model versions: MKF1020-400V (standard, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0409) and MKF1020-480V-C (with voltage and frequency converter, 480 V 3~ 60 Hz, Art.-No. 9020-0448).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | -40...180 °C p.3 ↗ | MKF1020-400V, 50 Hz |
| Temperature uniformity dependent on set | 0.1...1.8 ± K p.3 ↗ | MKF1020-400V, 50 Hz |
| Temperature fluctuation dependent on set | 0.1...0.5 ± K p.3 ↗ | MKF1020-400V, 50 Hz |
| Cooling down time from 180°C to -40°C | 120 min p.3 ↗ | MKF1020-400V, 50 Hz |
| Humidity range | 10...98 % RH p.3 ↗ | MKF1020-400V, 50 Hz |
| Max. heat compensation at 25°C and 90% RH | 1,500 W p.3 ↗ | MKF1020-400V, 50 Hz |
| Rated voltage | 400 V p.3 ↗ | MKF1020-400V, 50 Hz |
| Power frequency | 50 Hz p.3 ↗ | MKF1020-400V |
| Nominal power | 12,5 kW p.3 ↗ | MKF1020-400V, 50 Hz |
| Unit fuse | 32,0 A p.3 ↗ | MKF1020-400V, 50 Hz |
| Phase (nominal voltage) | 3~/N/PE p.3 ↗ | MKF1020-400V, 50 Hz |
| Interior volume | 991 L p.3 ↗ | MKF1020-400V, 50 Hz |
| Net weight of the unit (empty) | 636 kg p.3 ↗ | MKF1020-400V, 50 Hz |
| Load per rack | 40 kg p.3 ↗ | MKF1020-400V, 50 Hz |
| Permitted load | 200 kg p.3 ↗ | MKF1020-400V, 50 Hz |
| Wall clearance back | 300 mm p.3 ↗ | MKF1020-400V, 50 Hz |
| Wall clearance sidewise | 200 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Width net | 1,580 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Height net | 2,005 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Depth net | 1,480 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Interior width | 1,200 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Interior height | 1,020 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Interior depth | 810 mm p.4 ↗ | MKF1020-400V, 50 Hz |
| Sound-pressure level | 69 dB(A) p.4 ↗ | MKF1020-400V, 50 Hz |
| Number of shelves (std./max.) | 1/11 p.4 ↗ | MKF1020-400V, 50 Hz |
Operating Limits
- Temperature range: -40...180 °C p.3 ↗
- Humidity range: 10...98 % RH p.3 ↗
- Permitted load: 200 kg p.3 ↗
- Load per rack: 40 kg p.3 ↗
- Wall clearance back: 300 mm p.3 ↗
- Wall clearance sidewise: 200 mm p.4 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-3 p.1 ↗ | MKF | Series: dynamic climate chamber for rapid temperature changes with humidity control |
| 4-7 p.1 ↗ | 1020 | Model size (interior volume class, 991 L) |
| 8-11 p.2 ↗ | 400V | Standard model version, 400 V 3~/N/PE 50 Hz |
| 8-13 p.2 ↗ | 480V-C | Option model with voltage and frequency converter, 480 V 3~ 60 Hz |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| MKF1020-400V p.2 ↗ | 9020-0409 | Standard, 400 V 3~/N/PE 50 Hz | Standard model version | |
| MKF1020-480V-C p.2 ↗ | 9020-0448 | with voltage and frequency converter, 480 V 3~ 60 Hz | Option model with voltage and frequency converter |