Data Sheet Model MKF 240
Technical Reference Summary
Covered Models & Document Scope
Model MKF 240 dynamic climate chamber for rapid temperature changes with humidity control. Two variants: MKF240-400V (standard, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0380) and MKF240-480V-C (with voltage and frequency converter, 480 V 3~ 60 Hz, Art.-No. 9020-0358).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | -40...180 °C p.3 ↗ | MKF240-400V, 50 Hz |
| Temperature range | -40...180 °C p.3 ↗ | MKF240-480V-C, 60 Hz |
| Temperature uniformity dependent on set | 0.1...1.5 ± K p.3 ↗ | MKF240-400V, 50 Hz |
| Temperature fluctuation dependent on set | 0.1...0.5 ± K p.3 ↗ | MKF240-400V, 50 Hz |
| Cooling down time from 180°C to -40°C | 120 min p.3 ↗ | MKF240-400V, 50 Hz |
| Temperature range with humidity | 10...95 °C p.3 ↗ | MKF240-400V, 50 Hz |
| Humidity range | 10...98 % RH p.3 ↗ | MKF240-400V, 50 Hz |
| Max. heat compensation at 25°C and 90% RH | 400 W p.3 ↗ | MKF240-400V, 50 Hz |
| Rated voltage | 400 V p.3 ↗ | MKF240-400V, 50 Hz |
| Rated voltage | 480 V p.3 ↗ | MKF240-480V-C, 60 Hz |
| Power frequency | 50 Hz p.3 ↗ | MKF240-400V |
| Power frequency | 60 Hz p.3 ↗ | MKF240-480V-C |
| Nominal power | 6,8 kW p.3 ↗ | MKF240-400V, 50 Hz |
| Unit fuse | 16,0 A p.3 ↗ | MKF240-400V, 50 Hz |
| Phase (nominal voltage) | 3~/N/PE p.3 ↗ | MKF240-400V, 50 Hz |
| Phase (nominal voltage) | 3~ p.3 ↗ | MKF240-480V-C, 60 Hz |
| Interior volume | 228 L p.3 ↗ | MKF240-400V, 50 Hz |
| Net weight of the unit (empty) | 360 kg p.3 ↗ | MKF240-400V, 50 Hz |
| Load per rack | 30 kg p.3 ↗ | MKF240-400V, 50 Hz |
| Permitted load | 70 kg p.3 ↗ | MKF240-400V, 50 Hz |
| Sound-pressure level | 65 dB(A) p.4 ↗ | MKF240-400V, 50 Hz |
| Sound-pressure level | 67 dB(A) p.4 ↗ | MKF240-480V-C, 60 Hz |
| Energy consumption at 25°C and 60% RH | 1,500 Wh/h p.4 ↗ | MKF240-400V, 50 Hz |
| Width net | 1,115 mm p.4 ↗ | MKF240-400V, 50 Hz |
| Height net | 1,715 mm p.4 ↗ | MKF240-400V, 50 Hz |
Operating Limits
- Temperature range: -40...180 °C p.3 ↗
- Temperature range with humidity: 10...95 °C p.3 ↗
- Humidity range: 10...98 % RH p.3 ↗
- Max. heat compensation at 25°C and 90% RH: 400 W p.3 ↗
- Permitted load: 70 kg p.3 ↗
- Load per rack: 30 kg p.3 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-3 p.1 ↗ | MKF | Series: dynamic climate chamber for rapid temperature changes with humidity control |
| 4-6 p.1 ↗ | 240 | Model size (interior volume class, 228 L) |
| suffix p.2 ↗ | 400V | Standard mains variant: 400 V 3~/N/PE 50 Hz |
| suffix p.2 ↗ | 480V-C | Mains variant with voltage and frequency converter: 480 V 3~ 60 Hz |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| MKF240-400V p.2 ↗ | 9020-0380 | Standard, 400 V 3~/N/PE 50 Hz | Standard model version | |
| MKF240-480V-C p.2 ↗ | 9020-0358 | with voltage and frequency converter, 480 V 3~ 60 Hz | Option model with voltage and frequency converter |