Induztr
Engineering Digest Document Viewer Related Models

Agilent Cary 635 FTIR Spectrometer

Contents

  1. Overview pp. 3-5
    Introduces the Cary 635 FTIR design and key benefits
  2. Config pp. 6-7
    Shows how to configure the Cary 635 FTIR solution
  3. Procedure pp. 8-11
    Describes spectrum acquisition and liquid analysis workflows
  4. Overview pp. 12-22
    Covers applications in chemical, materials, academic, pharma, life science, food
  5. Spec pp. 23-24
    Provides technical specifications and performance details
  6. Services pp. 25-26
    Describes Agilent support and lab services
  7. p. 9 table, 4 rows
  8. p. 23 table, 6 rows

Single-reflection and multireflection ATR modules

ATR ModuleSpectral RangepH RangeSampling Surface DiameterEffective Path LengthcExample Sample TypesSwivel Press
Single-Reflection Di ATR Module6,300 to 350 cm−1 a ; 5,100 to 600 cm−1 bpH 1–142.5 mm1.1 μm at 4,000 cm−1 2.6 μm at 1,700 cm−1 7.3 μm at 600 cm−1Hard solids, particulates, polymers, pastes, liquids; all pH rangesYes
Single Reflection ZnSe ATR Module5,100 to 600 cm−1pH 5–92.5 mm1.1 μm at 4,000 cm−1 2.6 μm at 1,700 cm−1 7.3 μm at 600 cm−1Neat or concentrated; soft solids, pastes, gels, liquids; not strong acids or basesYes
Single Reflection Ge ATR Module5,100 to 600 cm−1pH 1–141.5 mm0.15 μm at 4,000 cm−1 0.36 μm at 1,700 cm−1 1.02 μm at 600 cm−1Carbon black-filled polymersYes
Multireflection ZnSe ATR Module5,100 to 600 cm−1pH 5–912 mm5.5 μm at 4,000 cm−1 13.0 μm at 1,700 cm−1 36.5 μm at 600 cm−1Lower-level components, more dilute solutions; pastes, gels, liquids; not strong acids or basesNo

a Cary 635 FTIR engine with KBr optics; b Cary 635 FTIR engine with ZnSe optics; c based on a typical sample with refractive index of n = 1.5

Technical Information

CharacteristicSpecification
Spectrometer BodyGlass-fiber reinforced polyamide for durability and structural strength
DetectorHigh-sensitivity thermoelectrically cooled deuterated triglycine sulfate (DTGS) detector
InterferometerPermanently aligned 45˚ Michelson flexure interferometer with 25 mm optical aperture
Spectral Range7,000 to 350 cm−1 with KBr optics 5,100 to 600 cm−1 with high-humidity ZnSe optics
Footprint26.4 cm x 15.7 cm to 34.8 cm x 17.8 cm; depending on attached sampling module
Weight3.5 kg to 4.6 kg; depending on attached sampling module

Engineering Facts

13 facts · every claim linked to a page
Installation

The instrument needs only a small bench footprint, so lab space planning can be minimal.

Determines bench allocation and whether it fits in gloveboxes or fume hoods.

Source
“requiring only 20 × 30 cm of bench space and weighing just 3.8 kg (8 lb)”
Selection

Choose the engine optics based on humidity and required spectral range, since ZnSe and KBr engines differ.

Optics choice sets spectral range and moisture robustness for the application.

Source
“is available with either zinc selenite (ZnSe) or potassium bromide (KBr) optics”
Selection

The all-ZnSe engine is more moisture-resistant than ZnSe-coated alternatives, which matters for humid sites.

Moisture resistance drives engine selection in tropical or humid environments.

Source
“the Cary 635 FTIR ZnSe engine uses optical components made entirely of ZnSe for a more robust, moisture-resistant system”
Installation

Sampling modules attach without user alignment, simplifying setup and module swaps.

Eliminates alignment procedures during installation and accessory changes.

Source
“Interchangeable sampling accessories require no user alignment.”
Selection

ATR crystal choice must match sample pH and chemistry; ZnSe is limited to pH 5–9 and not for strong acids or bases.

Prevents crystal damage and wrong module selection for aggressive samples.

Source
“Neat or concentrated; soft solids, pastes, gels, liquids; not strong acids or bases”
Selection

Diamond ATR covers the full pH 1–14 range for hard solids, polymers, and pastes.

Guides selection of the diamond ATR module for corrosive or abrasive samples.

Source
“Hard solids, particulates, polymers, pastes, liquids; all pH ranges”
Operation

DialPath uses factory-calibrated fixed path lengths selectable by rotation, avoiding cell assembly.

Simplifies liquid transmission setup and removes path-length calibration work.

Source
“Instant selection from three factory-calibrated, fixed path lengths (30 to 1,000 microns)”
Operation

DialPath needs only microliter sample volumes, reducing sample consumption.

Critical when sample volume is limited or expensive.

Source
“Quantitative and qualitative results from as little as 3 µL of sample”
Maintenance

A user-replaceable desiccant and long-life lamp reduce maintenance effort and cost.

Affects planned maintenance routines and cost of ownership.

Source
“a user-replaceable desiccant and long-lived lamp source minimize maintenance costs”
Compliance

MicroLab Pharma provides audit trails and user privileges to support data integrity regulations.

Required for GMP and 21 CFR Part 11 regulated environments.

Source
“offers secure data storage, user privilege management, and audit trails to facilitate data integrity regulations”
Compliance

Performance verification aligns with multiple pharmacopeial standards for regulated labs.

Confirms suitability for pharmacopeial compliance in pharma QC.

Source
“Performance verification with the Cary 635 FTIR instrument aligns with standards defined by the European, US, Indian, Chinese, Japanese, etc. Pharmacopeias.”
Performance

The 25 mm aperture and short internal path give lab-grade performance in a compact unit.

Sets expectations for sensitivity versus larger benchtop systems.

Source
“A large 25 mm optical aperture and short internal optical path in the interferometer provide performance levels typically associated with much larger lab FTIR systems.”
Maintenance

The interferometer is insensitive to vibration, reducing site vibration requirements.

Relaxes installation site vibration isolation requirements.

Source
“the interferometer is insensitive to vibration”