  1. Overview pp. 1-6
    Document purpose, what's new in version 6.9, table of contents
  2. Overview pp. 7-9
    Procedural notes, safety notes, and prerequisites
  3. Spec pp. 10-17
    Lists of required and optional equipment, reagents, and software
  4. Procedure pp. 18-26
    Spike mix preparation, storage, and sample prep steps
  5. Procedure pp. 27-30
    Prepare and incubate the cRNA labeling reaction
  6. Procedure pp. 31-34
    Purify labeled/amplified RNA and quantify cRNA yield
  7. Procedure pp. 35-40
    Prepare hybridization samples and assemble hybridization
  8. Procedure pp. 41-47
    Prewarm wash buffers and wash microarray slides
  9. Procedure pp. 48-56
    Scanner settings and data extraction with Feature Extraction software
  10. Procedure pp. 57-72
    Alternative purification, thermocycler protocol, and labeling kit prep
  11. Procedure pp. 73-78
    Assess RNA and cRNA quality with Bioanalyzer and NanoDrop
  12. Procedure pp. 79-83
    Stabilization and drying solution wash to prevent ozone damage
  13. pp. 84-96
    Kit contents, supplemental guides, microarray handling tips, related reagents
  14. p. 13 table, 7 rows
  15. p. 11 table, 8 rows
  16. p. 16 table, 10 rows
  17. p. 10 table, 8 rows
  18. p. 12 table, 23 rows
  19. p. 14 table, 18 rows
  20. p. 74 table, 4 rows
  21. p. 79 table, 3 rows
  22. p. 96 table, 3 rows
  23. p. 15 table, 9 rows

Table 8 Required Equipment

DescriptionVendor and part number
Agilent Microarray ScannerAgilent p/n G4900DA, G2565CA or G2565BA
Hybridization Chamber, stainlessAgilent p/n G2534A
Hybridization gasket slides 1 microarray/slide, 5 slides/boxAgilent p/n G2534-60003
Hybridization gasket slides 2 microarrays/slide, 5 slides/boxAgilent p/n G2534-60002
Hybridization gasket slides 4 microarrays/slide, 5 slides/boxAgilent p/n G2534-60011
Hybridization gasket slides 8 microarrays/slide, 5 slides/boxAgilent p/n G2534-60014
Go to www.agilent.com/genomics to see all available kit configurations.

Table 2 Catalog SurePrint G3 and HD Microarrays - Mouse

Part NumberDescription
G4852BG3 Mouse GE 8×60K v2 Microarray Kit (3 slides)
G4858A-074809G3 Mouse GE 8×60K v2 Microarray (1 slide)
G4852AG3 Mouse GE 8×60K Microarray Kit (3 slides)
G4858A-028005G3 Mouse GE 8×60K Microarray (1 slide)
G4846AHD Mouse GE 4×44K v2 Microarray Kit (5-slides)
G2519F-026655HD Mouse GE 4×44K v2 Microarray (1 slide)
G4122FHD Whole Mouse Genome Microarray Kit, 4×44K (5 slides)
G2519F-014868HD Whole Mouse Genome Microarray Kit, 4×44K (1 slide)

Table 10 Optional Equipment/Reagents

DescriptionVendor and part number
2100 BioanalyzerAgilent p/n G2939AA
RNA 6000 Nano Assay Kit (RNA Series II Kit)Agilent p/n 5067-1511
Quick Amp Labeling Kit, Two-ColorAgilent p/n 5190-0444
Stabilization and Drying Solution*Agilent p/n 5185-5979
Ozone-Barrier Slide Cover*Agilent p/n G2505-60550
Absolutely RNA Nanoprep KitAgilent p/n 400753
slide boxCorning p/n 07201629
acetonitrileSigma p/n 271004-1L
sulfolaneSigma p/n T22209
thermal cyclerPCR 96-well plate or 0.2 mL PCR tubes

* Recommended when processing microarrays in high ozone environment.

Table 1. Catalog SurePrint G3 and HD Microarrays - Human

Part NumberDescription
G4851CG3 Human Gene Expression 8×60K v3 Microarray Kit (3 slides)
G4858A-072363G3 Human Gene Expression 8×60K v3 Microarray (1 slide)
G4851BG3 Human Gene Expression 8×60K v2 Microarray Kit (3 slides)
G4858A-039494G3 Human Gene Expression 8×60K v2 Microarray (1 slide)
G4851AG3 Human GE 8×60K Microarray Kit (3 slides)
G4858A-028004G3 Human GE 8×60K Microarray (1 slide)
G4845AHD Human GE 4×44K v2 Microarray Kit (5 slides)
G2519F-026652HD Human GE 4×44K v2 Microarray (1 slide)

Table 4 Catalog SurePrint HD Microarrays and Microarray Kits - Model Organisms/Non-Human

Part NumberDescription
G2519F-021169Arabidopsis (V4) Gene Expression Microarray, 4×44K (1 slide)
G2519F-021623Barley Gene Expression Microarray, 4×44K (1 slide)
G2519F-023647Bovine (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-022520Brassica Gene Expression Microarray, 4x44K (1 slide)
G2519F-021193Canine (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-020186C. elegans (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-026441Chicken (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-022523Cotton Gene Expression Microarray, 4×44K (1 slide)
G2519F-021791Drosophila Gene Expression Microarray, 4×44K (1 slide)
G2519F-021322Horse Gene Expression Microarray, 4×44K (1 slide)
G2519F-015060Magnaporthe (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-022524Medicago Gene Expression Microarray, 4×44K (1 slide)
G2519F-020449Mosquito Gene Expression Microarray, 4×44K (1 slide)
G2519F-026440Porcine (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-026806Rhesus Macaque (V2) Gene Expression Microarray, 4×44K (1 slide)
G2519F-020908Rabbit Gene Expression Microarray, 4×44K (1 slide)
G2519F-015241Rice Gene Expression Microarray, 4×44K (1 slide)
G2519F-020938Salmon Gene Expression Microarray, 4×44K (1 slide)
G4813A-019921Sheep Gene Expression Microarray, 8×15K (1 slide)
G2519F-021113Tobacco Gene Expression Microarray, 4×44K (1 slide)
G2519F-022297Wheat Gene Expression Microarray, 4×44K (1 slide)
G4813A-016322Yeast (V2) Gene Expression Microarray, 8×15K (1 slide)
G2519F-026437Zebrafish (V3) Gene Expression Microarray, 4×44K (1 slide)

Table 8 Required Equipment (continued)

DescriptionVendor and part number
Hybridization oven; temperature set at 65°CAgilent p/n G2545A
Hybridization oven rotator for Agilent Microarray Hybridization ChambersAgilent p/n G2530-60029
nuclease-free 1.5 mL microfuge tubeAmbion p/n 12400 or equivalent
magnetic stir bar (×2)Corning p/n 401435 or equivalent
magnetic stir plate (×2)Corning p/n 6795-410 or equivalent
circulating water baths or heat blocks set to 37°C, 40°C, 60°C, 65°C, 70°C, and 80°C,Corning p/n 6795-420 or equivalent
microcentrifugeEppendorf p/n 5417R or equivalent
sterile storage bottleNalgene 455-1000 or equivalent
spectrophotometerNanoDrop p/n ND-1000 UV-VIS or equivalent
micropipettorPipetman P-10, P-20, P-200, P-1000 or equivalent
slide-staining dish, with slide rack (×3)Thermo Shandon p/n 121 or equivalent
clean forceps
ice bucket
powder-free gloves
sterile, nuclease-free aerosol barrier pipette tips
vortex mixer
timer
nitrogen purge box for slide storage

Table 30 Compatible Assays

DescriptionCompatible Assay
RNA 6000 Nano LabChip KitEukaryote Total RNA Nano Assay Qualitative range 5 to 500 ng/µL
RNA 6000 Nano LabChip KitmRNA Nano Assay* Qualitative range 25 to 250 ng/µL
RNA 6000 Pico LabChip KitEukaryote Total RNA Pico Assay Qualitative range 50 to 5000 pg/µL in water
RNA 6000 Pico LabChip KitmRNA Pico Assay* Qualitative range 250 to 5000 pg/µL in water

* The mRNA assays are suitable for analysis of cRNA as well.

Table 31

Ozone level in your labWash ProcedureOzone-Barrier Slide Cover
< 5 ppbWash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 44No
> 5 ppb and < 10 ppbWash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 44Yes
> 10 ppbWash Procedure with Stabilization and Drying Solution. See "Step 1. Prepare the Stabilization and Drying Solution" on page 80 and "Step 2. Wash with Stabilization and Drying Solution" on page 81.Yes

Wash procedure to follow

Related Microarray Reagents

DescriptionVendor and part number
Universal Human Reference RNAAgilent p/n 740000
Universal Mouse Reference RNAAgilent p/n 740100
Universal Rat Reference RNAAgilent p/n 740200

Table 9 Required Reagents

DescriptionVendor and part or catalog number
Low Input Quick Amp Labeling Kit, Two-ColorAgilent p/n 5190-2306
RNA Spike-In Kit, Two-ColorAgilent p/n 5188-5279
Gene Expression Hybridization KitAgilent p/n 5188-5242
Gene Expression Wash Buffer KitAgilent p/n 5188-5327
DNase/RNase-free distilled waterInvitrogen p/n 10977-015
RNeasy Mini KitQiagen p/n 74104
ethanol (95% to 100% molecular biology grade)Sigma-Aldrich p/n E7023-6×500ML
Milli-Q water or equivalent
isopropyl alcohol (molecular biology grade)

Engineering Facts

14 facts · every claim linked to a page
Selection

The kit supports a total RNA input range of 10–200 ng or a minimum of 5 ng poly A+ RNA for two-color processing, which defines the minimum sample you can load.

Determines whether your available sample amount is compatible with this labeling kit before you commit to the workflow.

Source
“with a sample input RNA range between 10 ng and 200 ng of total RNA or a minimum of 5 ng of poly A+ RNA for two-color processing”
Performance

Impure template RNA (low A260/A230) degrades results and must be purified before use.

Sets a purity acceptance criterion for input RNA that must be checked before labeling.

Source
“RNA that is not pure, as measured by A260/A230 ratio, can lead to poor results and must be purified.”
Safety

The Stabilization and Drying Solution is toxic and flammable and must be handled in a fume hood with gloves and eye/face protection.

Dictates lab ventilation and PPE requirements for the drying step installation.

Source
“Stabilization and Drying Solution is toxic and flammable and must be used in a suitable fume hood.”
Safety

LiCl in the 2× Hi-RPM Hybridization Buffer is toxic and a potential teratogen, requiring suitable PPE.

Requires PPE planning when handling hybridization buffer.

Source
“LiCl is toxic and a potential teratogen.”
Operation

Spike A must always be labeled with Cy3 and Spike B with Cy5, and this orientation is hard-coded in the software.

Prevents dye-swap errors that would invert the LogRatio output and require manual correction.

Source
“Always label Spike A Mix with cyanine 3 and Spike B Mix with cyanine 5.”
Operation

Fragmentation must be stopped after exactly 30 minutes at 60°C by cooling and adding hybridization buffer.

Over-fragmentation degrades cRNA and compromises hybridization signal.

Source
“Do not incubate sample in the next step for more than 30 minutes.”
Operation

Do not vortex the hybridization mix because the surfactant forms bubbles; mix only by careful pipetting.

Bubbles in the hybridization mix cause array artifacts and uneven hybridization.

Source
“Do not mix on a vortex mixer; mixing on a vortex mixer introduces bubbles.”
Maintenance

Adding 0.005% Triton X-102 to wash buffers reduces array wash artifacts and is highly recommended.

Improves wash quality and reproducibility of array results.

Source
“The addition of 0.005% Triton X-102 (10%) to the Gene Expression wash buffers reduces the possibility of array wash artifacts.”
Operation

For uniform washing, disassemble no more than eight microarray slides per wash group.

Sets a batch-size limit that affects throughput planning and wash consistency.

Source
“For uniform washing, do up to a maximum of eight disassembly procedures yielding eight microarray slides.”
Operation

Do not scan G3 microarrays with HD format settings because resolution will be too low for analysis.

Prevents selecting the wrong scanner profile that would yield unusable image resolution.

Source
“Do not scan G3 microarrays with HD format settings.”
Selection

HD formats require 5 µm scan resolution while G3 formats require 3 µm.

Determines scanner resolution setting matched to the array format chosen.

Source
“Scan resolution”
Performance

Default Feature Extraction protocols may not be optimal for custom arrays lacking enough replicated or validated negative control probes.

Custom array designs may need protocol tuning for reliable data extraction.

Source
“If custom arrays without enough replicated probes are used, or arrays with custom probes designated as Negative Control probes are used, the default FE Protocols may not be optimal.”
Operation

Using a thermocycler instead of water baths can slightly lower cRNA yield.

Informs the trade-off when choosing thermocycler versus water bath incubation.

Source
“The use of a thermocycler can slightly lower the yield of cRNA when compared to the use of water baths.”
Operation

Avoid pipetting volumes below 2 µL to ensure accuracy when preparing spike mixes.

Sets a minimum pipetting volume to maintain dilution accuracy.

Source
“Avoid pipetting volumes less than 2 µL to ensure accuracy.”