Introduces the CC-104A immersion circulator and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 3 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
max. ambient temperature
40 °C
from Serial-No.:
398007
1.0/20
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump
yes
max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
Bath volume
3 l
Filling capacity
2 l
Bath opening diameter
25 mm
Bath depth
150 mm
Overall dimensions WxDxH **
147x235x330 mm
Net weight
6 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.
Determines whether an external water supply or chiller is required to reach the desired setpoint.
Source
“Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C”
Installation
The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.
Electrical planning must provide a branch circuit rated between 10 A and 16 A at the correct voltage and phase configuration.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A”
Installation
External branch circuit protection according to UL 489 is mandatory.
The installer must provide a UL 489-listed breaker; otherwise the installation does not comply.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10 %, but a −10 % voltage combined with a +3 % frequency is not allowed.
Site power quality must be checked; some voltage/frequency combinations are outside the permitted operating envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be stated on the order to avoid a non-connectable unit on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.
Confirms the safety class for flammable liquids and the presence of adjustable overtemperature cut-out for safe operation.
Source
“Safety classification
III / FL”
Operation
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.
External circuit pressure drop must be checked against these pump curves to ensure adequate circulation.