Peter Huber Kältemaschinenbau CC-104A
Peter Huber Kältemaschinenbau CC-104A — Thermostatic baths.
The CC-104A is a 3-litre immersion circulator with a transparent polycarbonate bath, rated for 15–100 °C with a refrigerator, 20–100 °C with water cooling, or 25–100 °C uncooled. Its variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar, and the Pilot ONE controller holds stability to 0.02 K. It serves as a compact heating/circulation source for external loops and small immersed samples.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-104A Current | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Bath volume | 3 l |
| Dimensions (w × D × H) | 147 x 235 x 330 mm |
| Filling volume | 2 l (minimum) |
| Heating capacity | 1.5 - 2.1 kW |
| Maximum delivery flow rate | 27 l/min (pressure pump; also 0, 7 bar) suction pump; also 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability | 0.02 K |
| Bath opening WxD/BD mfr | ø = 25 / 150 (opening WxD/depth combined) |
| Weight | 6 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-104A is the smallest bath in the CC-10xA series; siblings CC-106A, CC-108A, CC-110A and CC-112A step up in bath volume. This model's 3 l total / 2 l filling capacity and 25 mm diameter × 150 mm deep opening suit small immersed samples and short external loops, not bulk sample immersion.
- The bath is transparent polycarbonate, which caps the usable upper temperature at 100 °C regardless of controller range. Safety classification III / FL per DIN 12876 and adjustable overtemperature protection are built in.
- The pump is a variable-speed pressure-and-suction type with M16x1 male connections; the pressure side reaches 27 l/min at 0.7 bar, the suction side 22 l/min at 0.4 bar.
Thermal behavior and media
- The achievable range is set by the cooling method, not the controller: 25–100 °C without cooling, 20–100 °C with water cooling, 15–100 °C with a refrigerator. Selection must therefore fix the cooling source before the setpoint is specified.
- Temperature stability is 0.02 K with the Pilot ONE controller; an external Pt100 sensor connection is provided for controlling at the sample or loop rather than the bath.
- Heating capacity is 1.5–2.1 kW, giving fast warm-up in the small 2 l filling volume.
Installation, control and operation
- Supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 10 A and a branch fuse between 10 A and 16 A; external branch protection to UL 489 is mandatory. Voltage and frequency tolerances are ±10 %, but a −10 % voltage with +3 % frequency is not permitted.
- Single- and two-phase units ship with a power cable and country-specific plug, which must be specified at ordering.
- Interfaces are Ethernet, USB host and device, and RS232. Ambient must stay within 5–40 °C; protection is IP20, so the unit must be kept dry. EMC classification is Class A, Group 1 to EN55011 (industrial environment).
Integration & Operating Notes
The installer must provide external branch circuit protection listed to UL 489; the circuit must be rated between the 10 A minimum and 16 A maximum fuse. A circuit outside this window either trips on inrush or fails to protect the 10 A maximum current draw.
Voltage and frequency each tolerate ±10 %, but the combination of a 10 % undervoltage with a 3 % overfrequency is explicitly not allowed. Site power quality should be checked before commissioning, since this corner of the envelope can overheat the compressor or pump motor.
Without cooling the unit cannot go below 25 °C; water cooling extends it to 20 °C and a refrigerator to 15 °C. The setpoint specification must therefore be matched to the installed cooling method, and the 100 °C ceiling is set by the polycarbonate bath, not the controller.
The Pilot ONE controller accepts an external Pt100 sensor, so the control point can be placed at the sample or in the external loop instead of the bath. This matters when the 2 l filling volume and small bath would otherwise dominate the measured temperature.
The extended 4-year warranty is granted only on registration; without it the standard coverage applies. Registration should be completed at commissioning so the warranty period is not lost.
Selection & Integration Q&A
Pick the CC-104A when the load fits a 25 mm × 150 mm opening and a 2 l filling volume, or when the external loop is short. Larger siblings add bath volume for bigger immersed samples or more loop inventory; they do not change the 100 °C polycarbonate ceiling.
The loop must be sized against 27 l/min at 0.7 bar on the pressure side and 22 l/min at 0.4 bar on the suction side, with M16x1 male pump connections requiring matching hose adapters. Excessive loop pressure drop will reduce circulation below these maxima.
A 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A fuse and external UL 489 branch protection, in a dry location at 5–40 °C ambient. Avoid −10 % voltage combined with +3 % frequency; specify the country-specific plug at ordering.
The cooling source (water cooling or refrigerator) needed below 25 °C, hose adapters for the M16x1 pump connections, and the external Pt100 sensor if control at the load is required. The power cable and country plug are included only if specified at ordering.