Product identification and general description of the CC-108A unit
Specp. 2
Technical data per DIN 12876 including temperature and performance ratings
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connenction (optional)
M16x1 male
Bath volume
6 l
min. filling capacity
4 l
Width bath opening WxD/ bath depth
130x210/ 150 mm
Overall dimensions WxDxH **
147x407x330 mm
Net weight
6 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.
Determines whether the bath can reach the required low temperatures and which cooling option must be specified.
Source
“Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A and a fuse rated between 10 A and 16 A.
Electrical supply and branch-circuit protection must be sized accordingly before installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A”
Installation
Voltage and frequency tolerances are ±10 % only if the frequency deviation does not run in the opposite direction, so a −10 % voltage with +3 % frequency is not allowed.
Site power quality must be checked; combined out-of-tolerance voltage and frequency can invalidate the output data.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Compliance
External branch circuit protection according to UL 489 is required.
The installation must include a UL 489-listed branch circuit protector to meet the manufacturer's requirement.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified when ordering.
The correct plug type must be stated on the order to avoid a mismatch with the site socket.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The bath has a 6 l volume with a minimum filling capacity of 4 l.
The minimum fill level must be respected for safe pump operation and temperature control.
Source
“Bath volume
6 l
min. filling capacity
4 l”
Operation
The pressure pump delivers up to 27 l/min at 0.7 bar, and the suction pump delivers up to 22 l/min at 0.4 bar.
Pump performance determines whether the external circuit can achieve the required flow and pressure drop.