Induztr
Engineering Digest Document Viewer Related Models

CC-205B Refrigerated/Heating Circulator

Contents

  1. Overview p. 1
    Introduces the CC-205B circulator and its main features
  2. Spec pp. 1-2
    Performance, temperature, electrical and physical data per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 36 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:394677
1.2/20

Technical data according to DIN 12876

ParameterValue
Operating temperature range45...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pumpyes
max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Bath volume4,8 l
Filling capacity3 l
Width bath opening WxD105x90 mm
Bath depth150 mm
Height of bath opening190 mm
Overall dimensions WxDxH **178x337x355 mm
Net weight9 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method chosen: 45–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the unit can reach the required process temperatures with the available cooling utility.

Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Performance

Heating power scales with supply voltage from 1.5 kW at 200 V up to 2.1 kW at 240 V, so the actual heat-up performance depends on site voltage.

Needed to verify the available electrical supply delivers the expected heating capacity.

Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Installation

The unit requires a 200–240 V single/two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.

Defines the branch circuit and protection device sizing for safe installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

External branch circuit protection according to UL 489 is mandatory.

Dictates the type of upstream protective device that must be provided in the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances of ±10% are permitted only if they do not act in opposite directions simultaneously.

Prevents installation on unstable grids where combined deviations could damage the unit.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug is supplied for the destination country and avoids rework on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

The unit needs a cooling water connection via a 3/8" cooling coil for tap water.

Requires planning a water supply and drain line at the installation location.

Source
“Cooling coil for (tap) water (3/8").”
Installation

The pump connection is M16x1 male, which must match the external circuit plumbing.

Determines the fittings and adapters needed to connect the external loop.

Source
“Pump connection M16x1 male”
Operation

The unit must be operated within an ambient temperature range of 5 °C to 40 °C.

Constrains the room conditions where the circulator can be installed and operated.

Source
“min. ambient temperature 5 °C”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Confirms the built-in safety cut-out that protects the bath and process from overheating.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.

Used to check whether the pump can overcome the pressure drop of the external circuit.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Selection

The bath has a 4.8 l volume with 3 l filling capacity and a 105x90 mm opening, 150 mm deep.

Determines whether the bath opening and volume suit the samples or immersion coolers to be used.

Source
“Bath volume 4,8 l Filling capacity 3 l Width bath opening WxD 105x90 mm Bath depth 150 mm”
Compliance

The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit has IP20 degree of protection, so it must be kept away from water spray and dust ingress.

Restricts placement in wet or dusty environments and affects enclosure planning.

Source
“Degree of Protection IP20”