Peter Huber Kältemaschinenbau CC-205B — Thermostatic baths.

The CC-205B is a 4.8-litre stainless-steel bath circulator with a variable-speed pressure/suction pump (M16x1 male) and Pilot ONE controller. Its operating range is configuration-dependent: 45–200 °C uncooled, 20–200 °C with water cooling, or -30–200 °C with a refrigerator. It serves as a temperature-control source for external circuits and immersed samples.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
CC-205B Current Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Filling volume 3 l (minimum)
Maximum delivery flow rate 27 l/min (pressure pump; also 0, 7 bar)
suction pump; also 0, 4 bar: 22 l/min
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 45...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 178 x 337 x 355 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 105 x 90 / 150 mm
Bath volume mfr 4, 8 l
Dimensions, Weight & Mechanical Design
Weight 9 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-205B is the 4.8-litre bath opening (105 x 90 mm, 150 mm deep) member of the CC-20xB stainless-steel series, positioned between the smaller CC-202C and the larger CC-208B/CC-212B/CC-215B siblings. Heating capacity is voltage-dependent, from 1.5 kW at 200 V to 2.1 kW at 240 V, and temperature stability is stated as ±0.02 K. The enclosure is 178 x 337 x 355 mm and weighs 9 kg.

Thermal behaviour and media

Control, connection and operation

The Pilot ONE controller provides Ethernet, USB host/device and RS232 interfaces, plus a Pt100 external sensor input. Pump speed is adjustable. The pump connection is M16x1 male. Overtemperature protection is adjustable per DIN 12876, safety class III/FL. The unit is rated IP20 and Class A, Group 1 to EN55011, so it is intended for non-residential EMC environments and must be kept from water spray and dust.

Integration & Operating Notes

Electrical
Branch circuit must be fused between 10 A and 16 A

The supply is 200–240 V 1~/2~ 50/60 Hz with a maximum current of 10 A. A minimum 10 A and maximum 16 A fuse is required, and external branch circuit protection to UL 489 is mandatory; undersizing the protection risks nuisance trips, oversizing removes the intended safeguard.

Datasheet, Technical Specifications, page 1–2
Electrical
Voltage and frequency tolerances must not oppose each other

A ±10% voltage tolerance is permitted only if the frequency tolerance does not simultaneously run in the opposite direction. Combined worst-case deviations on an unstable grid can exceed the unit's design margin, so supply quality should be verified before commissioning.

Datasheet, Technical Specifications, page 2
Cooling & Media
Cooling method fixes the attainable low-temperature limit

Without cooling the range starts at 45 °C; with the 3/8" tap-water cooling coil it starts at 20 °C; only with a refrigerator does it reach -30 °C. The cooling utility must therefore be decided before specifying the unit, since it changes the usable process window.

Datasheet, Technical Specifications, page 1
Process Limits
External circuit pressure drop must stay within pump capability

The pressure pump provides up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar. A circuit whose pressure drop exceeds these values will reduce flow below the required rate, so the loop resistance should be calculated against these curves before installation.

Datasheet, Technical Specifications, page 1
Mounting
Country-specific plug must be specified at ordering

Single/two-phase units are delivered with a power cable and a country-specific plug that has to be stated when ordering. Omitting this detail leaves the unit without a usable mains termination and forces a site retrofit.

Datasheet, Technical Specifications, page 2
Mounting
Install only in a dry, dust-free room between 5 and 40 °C

The IP20 enclosure and the 5–40 °C ambient limit exclude wet, sprayed or dusty locations. Because the unit is Class A, Group 1 to EN55011, additional EMC measures may be needed if it is placed near sensitive residential or measurement electronics.

Datasheet, Technical Specifications, page 1–2
Process Limits
Adjustable overtemperature cut-out protects bath and process

Overtemperature protection per DIN 12876 with safety class III/FL must be set to a value above the intended process temperature but below the limits of the bath fluid and immersed samples; an incorrect setting either trips the unit prematurely or removes the protective function.

Datasheet, Technical Specifications, page 1

Selection & Integration Q&A

When should I choose the CC-205B rather than the CC-208B or CC-212B?

Pick the CC-205B when a 4.8 l bath with a 105 x 90 mm opening and 3 l minimum fill suits the samples or immersion cooler. Larger siblings such as the CC-208B or CC-212B are for bigger bath volumes and openings; the CC-202C is the smaller alternative.

What must the external circuit and its fittings satisfy to work with this unit?

The pump connection is M16x1 male, so the loop plumbing must mate with that thread. Flow must stay within 27 l/min at 0.7 bar (pressure) and 22 l/min at 0.4 bar (suction), and a 3/8" tap-water coil connection is needed if water cooling is used.

What installation environment and electrical supply does the unit require?

Install in a dry room at 5–40 °C ambient, IP20, away from water spray and dust. Supply is 200–240 V 1~/2~ 50/60 Hz, max 10 A, with a 10–16 A fuse and UL 489 branch protection; voltage and frequency tolerances must not oppose each other.

What is not included in the base delivery and what must be ordered or planned separately?

The country-specific mains plug must be specified at ordering. A refrigerator is required to reach -30 °C and is not part of the uncooled configuration; the 3/8" tap-water cooling coil line must be provided on site. External Pt100 sensors and loop plumbing are separate items.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-205B Refrigerated/Heating Circulator
Datasheet ·96,3 КБ ·2 pp.
View PDF