CC-225B Immersion Circulator — Technical Data Sheet
Contents
Overviewp. 1
Product identification and general description of the CC-225B unit
Specpp. 1-2
Technical data table per DIN 12876 including temperature and performance ratings
p. 1table, 34 rows
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connenction (optional)
M16x1 male
Bath volume
23,5 l
min. filling capacity
17 l
Width bath opening WxD/ bath depth
290x329/ 200 mm
Overall dimensions WxDxH **
350x555x425 mm
Net weight
16 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The achievable operating temperature range depends on the cooling method chosen: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether an external cooler or refrigerator is required for the intended process temperature.
Source
“Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C”
Installation
The unit accepts 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so branch protection must be sized between 10 A and 16 A.
Needed to plan the electrical supply, cable, and fuse rating at the installation site.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A”
Installation
Voltage and frequency tolerances of ±10% are only acceptable if they do not act in opposite directions, so supply quality must be verified before commissioning.
Prevents installation on a supply whose combined voltage/frequency deviation is outside the permitted envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety
External branch circuit protection according to UL 489 is mandatory, so the site must provide a suitable UL-listed breaker.
Affects the electrical panel design and compliance of the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
The correct plug type must be stated on the order to avoid rework at installation.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A arrive with cable but without plug, and above 63 A without cable or plug, so the site must plan termination hardware.
Determines what electrical termination work and hardware the installer must provide.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The bath requires a minimum filling of 17 l within its 23.5 l volume, so the heat transfer fluid charge must be planned accordingly.
Sets the minimum fluid quantity needed for safe pump operation and the fluid volume to procure.
Source
“Bath volume
23,5 l
min. filling capacity
17 l”
Operation
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar, which bounds the external circuit resistance that can be served.
Used to check whether the external loop and connected equipment can be supplied by the pump.